Study sheet

CBC, Analyzers and Spurious Results

CBC parameters and indices with the ASCP ranges, how analyzers count cells, flags and histograms, spurious results and their fixes, reticulocytes, ESR, manual counts and body fluids

Checked against current sources November 2026

Practice 120 CBC, Analyzers and Spurious Results questions free

Page 1 of the CBC, Analyzers and Spurious Results PDF, a printed page in columns with sections on CBC parameters and the ASCP ranges, Indices, RDW and the rule of three, How analyzers count, Spurious results and the fix and Corrections and estimates to know by handEnlarge

Printable PDFLetter landscape, ready to print or keep with your notes

View PDFDownload

CBC, Analyzers and Spurious Results: PDF preview

Page 1 of the CBC, Analyzers and Spurious Results PDF, a printed page in columns with sections on CBC parameters and the ASCP ranges, Indices, RDW and the rule of three, How analyzers count, Spurious results and the fix and Corrections and estimates to know by hand

View PDFDownload

CBC parameters and the ASCP ranges

Ranges are the ASCP BOC composite ranges (June 2026), for exam use only. Units ×10³/µL equal ×10⁹/L.
ParameterMeasured or calculatedASCP range, conventional (SI)
WBCMeasured (counted)3.6–10.6 ×10³/µL (×10⁹/L)
RBCMeasured (counted)4.00–6.00 ×10⁶/µL (×10¹²/L)
HemoglobinMeasured (photometry after lysis)12.0–18.0 g/dL (120–180 g/L)
HematocritCalculated on most analyzers: RBC × MCV ÷ 1035–50% (0.35–0.50 L/L)
MCVMeasured (mean cell size); by hand: Hct × 10 ÷ RBC76–100 fL
MCHCalculated: Hgb × 10 ÷ RBC26–34 pg
MCHCCalculated: Hgb × 100 ÷ Hct32–36 g/dL (320–360 g/L)
RDWCalculated from the RBC size curve11.5–14.5% (0.115–0.145)
PlateletsMeasured (counted)150–450 ×10³/µL (×10⁹/L)
ReticulocytesMeasured (fluorescence or supravital stain)0.5–2.5%; absolute 20–115 ×10³/µL
nRBCsCounted per 100 WBCs0 per 100 WBCs
Absolute count = WBC × percentage. Judge a differential on absolute numbers: a 'high' percentage can be a normal absolute count.
CellRelativeAbsolute, ×10³/µL (×10⁹/L)
Neutrophils50–70%1.7–7.5
Lymphocytes18–42%1.0–3.2
Monocytes2–11%0.1–1.3
Eosinophils1–3%0–0.3
Basophils0–2%0–0.2

Indices, RDW and the rule of three

How analyzers count

PrincipleWhat it measuresWeak point
Electrical impedance (Coulter principle)Each cell in an electrolyte passes through a small aperture between two electrodes; the pulse is proportional to cell volume. Particles in a set size window are counted as that cell.Anything of the right size is counted: microcytes, fragments, lipid particles and bacteria in the platelet window; giant platelets and clumps fall outside it.
Light scatter (flow cell)Low-angle (forward) scatter tracks size; high-angle (side) scatter tracks internal complexity and granularity. Two-dimensional plots separate cell types.Cells with unusual size or granules can land in the wrong cluster; the analyzer flags rather than identifies.
Fluorescence flow cytometryA nucleic acid dye stains RNA and DNA, so reticulocytes and platelets (separated from red cell fragments) are counted by fluorescence plus scatter; some analyzers run the whole WBC differential this way.Fewer interferences, but still needs smear review for flagged samples.
VCS (volume, conductivity, scatter)Impedance gives volume, radiofrequency conductivity gives internal composition (nucleus, N:C ratio), laser scatter gives granularity.Same limits as other flow methods: abnormal cells are flagged, not named.
Immunological platelet count (CD41/CD61 flow)Labeled antibodies to GPIIb-IIIa; the ICSH/ISLH reference method for platelets.A reference method, not a routine one.

Spurious results and the fix

From Gulati 2022 and Cattaneo 2025. Check the analyzer manufacturer's manual for model-specific behavior.
CauseWhat happensHow you spot itFix
Cold agglutinin (IgM autoagglutination)RBC and Hct falsely low; MCV, MCH and MCHC falsely high; WBC can be low if trapped in clumpsMCHC far above 36; Hgb × 3 far above Hct; grainy tube wall; red cell clumps on the smearWarm the tube at 37 °C for 10–15 min and rerun at once; check a fresh smear. If it fails, recollect and keep at 37 °C
LipemiaHgb, MCH and MCHC falsely high (turbidity); impedance platelets can rise from lipid particlesMilky plasma; high MCHC; Hgb and Hct disagreeReplace plasma with an equal volume of diluent and rerun; report Hgb, MCH and MCHC from the rerun (and counts, Hct, MCV and RDW from the first run if the two runs disagree)
In vitro hemolysisRBC and Hct falsely low; MCH and MCHC high; ghosts may raise the platelet countRed plasma; mildly high MCHC; K, LD and AST up with a normal reticReport with a 'specimen hemolyzed' comment, or report only WBC, Hgb, MCV, RDW and a smear-verified platelet count
Icterus (bilirubin above about 25 mg/dL)Hgb, MCH and MCHC falsely highBright yellow plasma; Hgb interference flagDilute 1:2 to 1:5 with diluent, rerun and correct for the dilution
Platelet clumping (EDTA-dependent)Platelets falsely low; large clumps counted as WBCsClump flag; clumps at the feather edge; no bleeding historyVortex 1–2 min and rerun (works in about half), or recollect in citrate and multiply by 1.1
Platelet satellitismPlatelets can be falsely low, usually only slightly (platelets ringed around neutrophils)Smear shows rosettes of platelets on neutrophilsCitrate tube (× 1.1) or vortex and rerun
Giant plateletsPlatelets falsely low; WBC sometimes falsely highPlatelets as big as red cells on the smear; MPN, MDS, inherited macrothrombocytopeniasSmear estimate or phase-microscopy manual count
nRBCsWBC falsely high (counted as WBCs) unless the analyzer correctsnRBC flag; nRBCs on the differentialCorrected WBC = WBC × 100 ÷ (100 + nRBCs per 100 WBCs)
Cryoglobulins or cryofibrinogenWBC and/or platelets falsely high (precipitate counted)Turbid cold plasma; pale amorphous material on the smear; hepatitis C, myelomaWarm to 37 °C for 10–20 min and rerun
Very high WBC (100 ×10³/µL or more)Hgb, MCH, MCHC falsely high (turbidity); RBC, MCV and Hct can riseLeukemia sample; high MCHCDilute, rerun and correct for the dilution
Microcytes (MCV under 60 fL) and schistocytesPlatelets falsely high on impedance; MPV may be highPlatelet histogram does not return to baseline; fragments on the smearOptical or fluorescent platelet count, or smear estimate
Glucose 600 mg/dL (33 mmol/L) or moreMCV and Hct falsely high; MCHC falsely lowHigh MCV with low MCHC; often IV glucose contaminationDilute, let stand 10 min, rerun and correct for the dilution; or spun hematocrit and recalculate MCV and MCHC
Clots or fibrinSmall clots: all counts and Hgb low; fibrin clumps can raise the WBC; either can block the aperture (no result)Fibrin strands on the smear; delta failureRecollect; do not report
IV fluid dilutionAll counts, Hgb and Hct lowDelta failure; chemistry also dilutedRecollect away from the line

Corrections and estimates to know by hand

Practice CBC, Analyzers and Spurious Results questions free in your browser120 exam-style questions with rationales. No account or download needed.

In depth

The summary above is what the PDF prints. These sections go further.

Histograms and scatterplots in words

A histogram plots cell size (x-axis) against the number of cells (y-axis). A scatterplot (scattergram) plots two measurements against each other, such as side scatter against fluorescence, and each cell type forms a cluster. The analyzer sets discriminators between populations and raises a flag when cells fall between or outside them.

PatternWhat it means
RBC curve shifted left (smaller)Microcytic population (low MCV)
RBC curve shifted right (larger)Macrocytic population; large numbers of reticulocytes also shift it right
RBC curve broad, or with two peaksHigh RDW: mixed populations (for example a combined microcytic and macrocytic picture with a normal MCV)
Platelet curve does not fall to baseline at the upper end and runs into the RBC curveMicrocytes or fragments counted as platelets, or giant platelets crossing the discriminator
Platelet curve with a long right tail and a low countLarge or giant platelets or clumps: the count is falsely low (one example in Cattaneo 2025 had platelets past a 25 fL discriminator)
WBC histogram or scatterplot with extra small particles at the lower WBC thresholdParticles counted as WBCs: nRBCs, platelet clumps or unlysed red cells
Cells between or outside clusters on the differential plotSuspect flags (blasts, immature granulocytes, atypical lymphocytes): do a smear

Recognizing an unreliable CBC

  1. Read the flags, histograms and scatterplots for every sample.
  2. Check the delta: a large change from the patient's last result needs a reason (transfusion, dilution, wrong patient, clot).
  3. Check internal consistency: the rule of three, MCHC above 36 or under 28 g/dL, a result that does not fit the patient.
  4. Look at the tube: settled or spun, for lipemia, hemolysis, icterus, agglutinates, cryoprecipitate and clots. Chemistry HIL indices help.
  5. Look at the smear: clumps of platelets, WBCs or RBCs, fibrin strands, organisms, fragments, giant platelets, nRBCs, fat globules.
After Gulati 2013. Each laboratory writes its own criteria; the International Consensus Group for Hematology Review criteria are a starting point.
LevelWhat it isDone when
Smear scanCheck of the whole smear at 10× (including the feather edge) for clumps, then higher power for fragments, giant platelets and organisms; platelet estimateAnalyzer flags; first platelet count under 100 ×10³/µL; a 50% fall; clump, giant platelet, fragment or suspect WBC flags
Smear examination (manual differential)A 100-cell differential (25 cells in severe leukopenia, up to 200–300 in leukocytosis) with red cell, white cell and platelet morphologyThe automated differential is not reportable, by request, or by laboratory criteria
Smear reviewReview by a hematopathologist or experienced technologist, with interpretationPreset criteria (for example first blasts or organisms) or a clinician's request

Worked example: a cold agglutinin

ResultFirst run (room temperature)After 15 min at 37 °C
RBC (×10⁶/µL)1.804.00
Hgb (g/dL)12.012.0
Hct (%)17.036.0
MCV (fL)94.490.0
MCH (pg)66.730.0
MCHC (g/dL)70.633.3

The first run fails both rules of three (1.80 × 3 = 5.4, not 12.0; 12.0 × 3 = 36, not 17) and the MCHC is impossible. Hemoglobin is unchanged because the cells are lysed for that measurement; the cell-counting results are wrong because doublets and clumps are counted as single large cells or not at all. After warming, the indices are normal and the rules hold. Report the warmed results, and confirm on a smear made right after warming that the agglutination is gone.

Pseudothrombocytopenia in depth

Reticulocytes and the IRF

Erythrocyte sedimentation rate (ESR)

Manual counts with the hemocytometer

Body fluid counts on analyzers

Less common causes of spurious results

CauseEffectFix
Paraprotein (myeloma, Waldenström)Hgb, MCH, MCHC falsely high; rarely WBC and platelets highPlasma replacement and rerun
WBC clumpingWBC falsely low; no analyzer flags itVortex, warm to 37 °C, or citrate (× 1.1); confirm on a smear
Lysis-resistant red cells (Hb C, Hb S)WBC and Hgb falsely highDilute and rerun, or longer lysis
Bacteria, yeast, malariaPlatelets (single bacteria) or WBC (clusters, large yeasts) falsely highSmear estimate; tell the clinician if organisms are seen
Cytoplasmic fragments of leukemic cellsPlatelets falsely highSmear estimate
Megakaryocytes or carcinoma cells in bloodWBC falsely highCorrect like nRBCs: WBC × 100 ÷ (100 + cells per 100 WBCs)
Fat from a traumatic drawWBC falsely highSmear estimate, or recollect
Sample 1–4 days old at room temperatureMCV and Hct high; MCHC lowReject, or report only unaffected results
Practice free in your browser600+ exam-style questions with rationales. Free today, no account, no ads. Pro is coming in January 2027. Also free on Google Play.

Related study sheets

Calculators for this topic

Sources

  1. Gulati G, Uppal G, Gong J. Unreliable automated complete blood count results: causes, recognition, and resolution. Ann Lab Med. 2022;42(5):515-530. doi:10.3343/alm.2022.42.5.515 Link
  2. Gulati G, Song J, Florea AD, Gong J. Purpose and criteria for blood smear scan, blood smear examination, and blood smear review. Ann Lab Med. 2013;33(1):1-7. doi:10.3343/alm.2013.33.1.1 Link
  3. Cattaneo M. Pseudothrombocytopenia and other conditions associated with spuriously low platelet counts. Haematologica. 2025;110(8):1677-1692. doi:10.3324/haematol.2024.286234 Link
  4. Krishnamurthy V, Shivamurthy A, Kumar PV. Platelet count in impedance-based hematology analyzer: beware of trap! Asian J Transfus Sci. 2023;17(1):131-132. doi:10.4103/ajts.ajts_65_21 Link
  5. Chhabra G. Automated hematology analyzers: recent trends and applications. J Lab Physicians. 2018;10(1):15-16. doi:10.4103/JLP.JLP_124_17 Link
  6. Yang K, Sun B, Zhang S, Pan Y, Fang J. RDW-SD is superior to RDW-CV in reflecting liver fibrosis stage in patients with chronic hepatitis B. Infect Drug Resist. 2023;16:6881-6891. doi:10.2147/IDR.S427047 Link
  7. Merck Manual Professional Version. Evaluation of anemia. Link
  8. Piva E, Brugnara C, Chiandetti L, Plebani M. Automated reticulocyte counting: state of the art and clinical applications in the evaluation of erythropoiesis. Clin Chem Lab Med. 2010;48(10):1369-1380 (abstract). doi:10.1515/CCLM.2010.292 Link
  9. Piva E, Brugnara C, Spolaore F, Plebani M. Clinical utility of reticulocyte parameters. Clin Lab Med. 2015;35(1):133-163 (abstract). doi:10.1016/j.cll.2014.10.004 Link
  10. Kratz A, Plebani M, Peng M, et al; ICSH. ICSH recommendations for modified and alternate methods measuring the erythrocyte sedimentation rate. Int J Lab Hematol. 2017;39(5):448-457 (abstract). doi:10.1111/ijlh.12693 Link
  11. Lorubbio M, Diamanti D, Pieroni C, et al. Erythrocyte sedimentation rate reference intervals determined via VES-MATIC 5 and CUBE 30 Touch with respect to the Westergren method. Diagnostics (Basel). 2025;15(9):1101. doi:10.3390/diagnostics15091101 Link
  12. Alende-Castro V, Alonso-Sampedro M, Vazquez-Temprano N, et al. Factors influencing erythrocyte sedimentation rate in adults: new evidence for an old test. Medicine (Baltimore). 2019;98(34):e16816. doi:10.1097/MD.0000000000016816 Link
  13. Testing.com (ADLM). Erythrocyte sedimentation rate (ESR). Link
  14. Bourner G, De la Salle B, George T, et al; ICSH. ICSH guidelines for the verification and performance of automated cell counters for body fluids. Int J Lab Hematol. 2014;36(6):598-612 (abstract). doi:10.1111/ijlh.12196 Link
  15. Yoon S, Kim HR. Analytical performance of the digital morphology analyzer Sysmex DI-60 for body fluid cell differential counts. PLoS One. 2023;18(7):e0288551. doi:10.1371/journal.pone.0288551 Link
  16. Chueh HW, Hwang SM, Shim YJ, et al; Korean RBC Disorder Working Party. Korean clinical practice guidelines for the diagnosis of hereditary hemolytic anemia. Blood Res. 2022;57(2):86-94. doi:10.5045/br.2022.2021224 Link
  17. Hillman RS. Characteristics of marrow production and reticulocyte maturation in normal man in response to anemia. J Clin Invest. 1969;48(3):443-453 (abstract). doi:10.1172/JCI106001 Link
  18. Kumar U, Chandra H, Gupta AK, Singh N, Chaturvedi J. Role of reticulocyte parameters in anemia of first trimester pregnancy: a single center observational study. J Lab Physicians. 2020;12(1):15-19. doi:10.1055/s-0040-1713585 Link
  19. Brownstein AJ, Wilkinson JD, Liang LL, et al. Immature reticulocyte fraction: a novel biomarker of hemodynamic severity in pulmonary arterial hypertension. Pulm Circ. 2024;14(3):e12421. doi:10.1002/pul2.12421 Link
  20. Borisova J, Morshchinin IV, Nazarova VI, et al. Low-cost microalgae cell concentration estimation in hydrochemistry applications using computer vision. Sensors (Basel). 2025;25(15):4651 (hemocytometer chamber depth and Neubauer square size). doi:10.3390/s25154651 Link
  21. ASCP Board of Certification. Medical Laboratory Scientist MLS(ASCP) and MLS(ASCPi) Examination Content Guideline, revised June 9, 2026 (hematology outline, calculations and composite reference ranges). Link