Study sheet
CBC, Analyzers and Spurious Results
CBC parameters and indices with the ASCP ranges, how analyzers count cells, flags and histograms, spurious results and their fixes, reticulocytes, ESR, manual counts and body fluids
Practice 120 CBC, Analyzers and Spurious Results questions free
CBC parameters and the ASCP ranges
| Parameter | Measured or calculated | ASCP range, conventional (SI) |
|---|---|---|
| WBC | Measured (counted) | 3.6–10.6 ×10³/µL (×10⁹/L) |
| RBC | Measured (counted) | 4.00–6.00 ×10⁶/µL (×10¹²/L) |
| Hemoglobin | Measured (photometry after lysis) | 12.0–18.0 g/dL (120–180 g/L) |
| Hematocrit | Calculated on most analyzers: RBC × MCV ÷ 10 | 35–50% (0.35–0.50 L/L) |
| MCV | Measured (mean cell size); by hand: Hct × 10 ÷ RBC | 76–100 fL |
| MCH | Calculated: Hgb × 10 ÷ RBC | 26–34 pg |
| MCHC | Calculated: Hgb × 100 ÷ Hct | 32–36 g/dL (320–360 g/L) |
| RDW | Calculated from the RBC size curve | 11.5–14.5% (0.115–0.145) |
| Platelets | Measured (counted) | 150–450 ×10³/µL (×10⁹/L) |
| Reticulocytes | Measured (fluorescence or supravital stain) | 0.5–2.5%; absolute 20–115 ×10³/µL |
| nRBCs | Counted per 100 WBCs | 0 per 100 WBCs |
| Cell | Relative | Absolute, ×10³/µL (×10⁹/L) |
|---|---|---|
| Neutrophils | 50–70% | 1.7–7.5 |
| Lymphocytes | 18–42% | 1.0–3.2 |
| Monocytes | 2–11% | 0.1–1.3 |
| Eosinophils | 1–3% | 0–0.3 |
| Basophils | 0–2% | 0–0.2 |
Indices, RDW and the rule of three
- MCV (fL) = Hct (%) × 10 ÷ RBC (×10⁶/µL). Under 76 fL is microcytic and over 100 fL macrocytic on the ASCP range (many laboratories use 80 fL as the lower limit).
- MCH (pg) = Hgb (g/dL) × 10 ÷ RBC (×10⁶/µL).
- MCHC (g/dL) = Hgb (g/dL) × 100 ÷ Hct (%).
- RDW is the spread of red cell sizes. RDW-CV = SD of the size curve ÷ MCV × 100, so it moves with MCV; RDW-SD (in fL) does not.
- A high RDW with a normal MCV can be the only sign of two populations (one small, one large) whose average looks normal.
- Rule of three: RBC × 3 ≈ Hgb, and Hgb × 3 ≈ Hct ± 3. It holds only for normocytic, normochromic cells; a failure in such a patient means a result needs checking.
How analyzers count
| Principle | What it measures | Weak point |
|---|---|---|
| Electrical impedance (Coulter principle) | Each cell in an electrolyte passes through a small aperture between two electrodes; the pulse is proportional to cell volume. Particles in a set size window are counted as that cell. | Anything of the right size is counted: microcytes, fragments, lipid particles and bacteria in the platelet window; giant platelets and clumps fall outside it. |
| Light scatter (flow cell) | Low-angle (forward) scatter tracks size; high-angle (side) scatter tracks internal complexity and granularity. Two-dimensional plots separate cell types. | Cells with unusual size or granules can land in the wrong cluster; the analyzer flags rather than identifies. |
| Fluorescence flow cytometry | A nucleic acid dye stains RNA and DNA, so reticulocytes and platelets (separated from red cell fragments) are counted by fluorescence plus scatter; some analyzers run the whole WBC differential this way. | Fewer interferences, but still needs smear review for flagged samples. |
| VCS (volume, conductivity, scatter) | Impedance gives volume, radiofrequency conductivity gives internal composition (nucleus, N:C ratio), laser scatter gives granularity. | Same limits as other flow methods: abnormal cells are flagged, not named. |
| Immunological platelet count (CD41/CD61 flow) | Labeled antibodies to GPIIb-IIIa; the ICSH/ISLH reference method for platelets. | A reference method, not a routine one. |
Spurious results and the fix
| Cause | What happens | How you spot it | Fix |
|---|---|---|---|
| Cold agglutinin (IgM autoagglutination) | RBC and Hct falsely low; MCV, MCH and MCHC falsely high; WBC can be low if trapped in clumps | MCHC far above 36; Hgb × 3 far above Hct; grainy tube wall; red cell clumps on the smear | Warm the tube at 37 °C for 10–15 min and rerun at once; check a fresh smear. If it fails, recollect and keep at 37 °C |
| Lipemia | Hgb, MCH and MCHC falsely high (turbidity); impedance platelets can rise from lipid particles | Milky plasma; high MCHC; Hgb and Hct disagree | Replace plasma with an equal volume of diluent and rerun; report Hgb, MCH and MCHC from the rerun (and counts, Hct, MCV and RDW from the first run if the two runs disagree) |
| In vitro hemolysis | RBC and Hct falsely low; MCH and MCHC high; ghosts may raise the platelet count | Red plasma; mildly high MCHC; K, LD and AST up with a normal retic | Report with a 'specimen hemolyzed' comment, or report only WBC, Hgb, MCV, RDW and a smear-verified platelet count |
| Icterus (bilirubin above about 25 mg/dL) | Hgb, MCH and MCHC falsely high | Bright yellow plasma; Hgb interference flag | Dilute 1:2 to 1:5 with diluent, rerun and correct for the dilution |
| Platelet clumping (EDTA-dependent) | Platelets falsely low; large clumps counted as WBCs | Clump flag; clumps at the feather edge; no bleeding history | Vortex 1–2 min and rerun (works in about half), or recollect in citrate and multiply by 1.1 |
| Platelet satellitism | Platelets can be falsely low, usually only slightly (platelets ringed around neutrophils) | Smear shows rosettes of platelets on neutrophils | Citrate tube (× 1.1) or vortex and rerun |
| Giant platelets | Platelets falsely low; WBC sometimes falsely high | Platelets as big as red cells on the smear; MPN, MDS, inherited macrothrombocytopenias | Smear estimate or phase-microscopy manual count |
| nRBCs | WBC falsely high (counted as WBCs) unless the analyzer corrects | nRBC flag; nRBCs on the differential | Corrected WBC = WBC × 100 ÷ (100 + nRBCs per 100 WBCs) |
| Cryoglobulins or cryofibrinogen | WBC and/or platelets falsely high (precipitate counted) | Turbid cold plasma; pale amorphous material on the smear; hepatitis C, myeloma | Warm to 37 °C for 10–20 min and rerun |
| Very high WBC (100 ×10³/µL or more) | Hgb, MCH, MCHC falsely high (turbidity); RBC, MCV and Hct can rise | Leukemia sample; high MCHC | Dilute, rerun and correct for the dilution |
| Microcytes (MCV under 60 fL) and schistocytes | Platelets falsely high on impedance; MPV may be high | Platelet histogram does not return to baseline; fragments on the smear | Optical or fluorescent platelet count, or smear estimate |
| Glucose 600 mg/dL (33 mmol/L) or more | MCV and Hct falsely high; MCHC falsely low | High MCV with low MCHC; often IV glucose contamination | Dilute, let stand 10 min, rerun and correct for the dilution; or spun hematocrit and recalculate MCV and MCHC |
| Clots or fibrin | Small clots: all counts and Hgb low; fibrin clumps can raise the WBC; either can block the aperture (no result) | Fibrin strands on the smear; delta failure | Recollect; do not report |
| IV fluid dilution | All counts, Hgb and Hct low | Delta failure; chemistry also diluted | Recollect away from the line |
Corrections and estimates to know by hand
- Corrected WBC = uncorrected WBC × 100 ÷ (100 + nRBCs per 100 WBCs). The ASCP lists it for more than 10 nRBCs. Example: 15.0 ×10³/µL with 25 nRBCs → 15.0 × 100 ÷ 125 = 12.0 ×10³/µL.
- Citrate (blue top) tube results: multiply WBC, RBC, Hgb, Hct and platelets by 1.1 (9 parts blood to 1 part citrate). Indices need no correction.
- Diluted sample: multiply WBC, RBC, Hgb, Hct and platelets by the dilution factor; MCV, MCH, MCHC and RDW are unchanged.
- Platelet estimate: average platelets per 100× oil field (10 fields) × 15 = ×10³/µL (laboratories use factors of 10 to 20).
- Absolute count = WBC × percentage, or for retics, RBC × retic %. Example: 2.0% × 3.50 ×10⁶/µL × 10 = 70 ×10³/µL.
In depth
The summary above is what the PDF prints. These sections go further.
Histograms and scatterplots in words
A histogram plots cell size (x-axis) against the number of cells (y-axis). A scatterplot (scattergram) plots two measurements against each other, such as side scatter against fluorescence, and each cell type forms a cluster. The analyzer sets discriminators between populations and raises a flag when cells fall between or outside them.
| Pattern | What it means |
|---|---|
| RBC curve shifted left (smaller) | Microcytic population (low MCV) |
| RBC curve shifted right (larger) | Macrocytic population; large numbers of reticulocytes also shift it right |
| RBC curve broad, or with two peaks | High RDW: mixed populations (for example a combined microcytic and macrocytic picture with a normal MCV) |
| Platelet curve does not fall to baseline at the upper end and runs into the RBC curve | Microcytes or fragments counted as platelets, or giant platelets crossing the discriminator |
| Platelet curve with a long right tail and a low count | Large or giant platelets or clumps: the count is falsely low (one example in Cattaneo 2025 had platelets past a 25 fL discriminator) |
| WBC histogram or scatterplot with extra small particles at the lower WBC threshold | Particles counted as WBCs: nRBCs, platelet clumps or unlysed red cells |
| Cells between or outside clusters on the differential plot | Suspect flags (blasts, immature granulocytes, atypical lymphocytes): do a smear |
Recognizing an unreliable CBC
- Read the flags, histograms and scatterplots for every sample.
- Check the delta: a large change from the patient's last result needs a reason (transfusion, dilution, wrong patient, clot).
- Check internal consistency: the rule of three, MCHC above 36 or under 28 g/dL, a result that does not fit the patient.
- Look at the tube: settled or spun, for lipemia, hemolysis, icterus, agglutinates, cryoprecipitate and clots. Chemistry HIL indices help.
- Look at the smear: clumps of platelets, WBCs or RBCs, fibrin strands, organisms, fragments, giant platelets, nRBCs, fat globules.
| Level | What it is | Done when |
|---|---|---|
| Smear scan | Check of the whole smear at 10× (including the feather edge) for clumps, then higher power for fragments, giant platelets and organisms; platelet estimate | Analyzer flags; first platelet count under 100 ×10³/µL; a 50% fall; clump, giant platelet, fragment or suspect WBC flags |
| Smear examination (manual differential) | A 100-cell differential (25 cells in severe leukopenia, up to 200–300 in leukocytosis) with red cell, white cell and platelet morphology | The automated differential is not reportable, by request, or by laboratory criteria |
| Smear review | Review by a hematopathologist or experienced technologist, with interpretation | Preset criteria (for example first blasts or organisms) or a clinician's request |
Worked example: a cold agglutinin
| Result | First run (room temperature) | After 15 min at 37 °C |
|---|---|---|
| RBC (×10⁶/µL) | 1.80 | 4.00 |
| Hgb (g/dL) | 12.0 | 12.0 |
| Hct (%) | 17.0 | 36.0 |
| MCV (fL) | 94.4 | 90.0 |
| MCH (pg) | 66.7 | 30.0 |
| MCHC (g/dL) | 70.6 | 33.3 |
The first run fails both rules of three (1.80 × 3 = 5.4, not 12.0; 12.0 × 3 = 36, not 17) and the MCHC is impossible. Hemoglobin is unchanged because the cells are lysed for that measurement; the cell-counting results are wrong because doublets and clumps are counted as single large cells or not at all. After warming, the indices are normal and the rules hold. Report the warmed results, and confirm on a smear made right after warming that the agglutination is gone.
Pseudothrombocytopenia in depth
- About 0.1% of all blood counts, and 15–17% of low platelet counts in outpatients.
- Mechanism: naturally occurring antibodies (usually IgG, sometimes IgM) bind hidden GPIIb-IIIa epitopes exposed in EDTA. Most act like cold agglutinins (best at 4–20 °C). The count falls within minutes at room temperature, reaches its low point at about 2 hours and does not recover on rewarming to 37 °C, unlike a red cell cold agglutinin, so warming is not the fix.
- Not a disease: the patient has no bleeding tendency from it. Treating the number (transfusion, cancelled surgery) is the harm.
- Citrate and heparin do not prevent clumping in every case, so a citrate count must still be checked on its own smear. The smear is the diagnostic standard.
- Satellitism (platelets adhering to neutrophils) is also EDTA-dependent, much rarer (about 1 in 12,000 counts) and usually lowers the count only slightly; it is mediated by IgG bridging GPIIb-IIIa and neutrophil Fc receptors (FcγRIII). Analyzer flags for it are inconsistent, so the smear makes the diagnosis.
- Giant platelets (wider than about 7 µm) occur in inherited macrothrombocytopenias such as MYH9-related disease and Bernard-Soulier syndrome, in MPN and MDS, after splenectomy and in immune destruction. Type 2B von Willebrand disease shows large platelets and clumps.
Reticulocytes and the IRF
- Reticulocytes are young red cells that still hold RNA. A supravital stain shows the RNA as a blue network; on a Wright stain the same cells look blue-gray and larger (polychromasia).
- Analyzers count reticulocytes by fluorescence: a nucleic acid dye marks RNA and the analyzer counts cells by fluorescence and scatter. Automated counts replaced the microscope from the mid-1990s.
- Absolute retic = retic % × RBC (×10⁶/µL) × 10, in ×10³/µL. ASCP range 0.5–2.5% and 20–115 ×10³/µL.
- In anemia the percentage overstates production. Corrected retic % = retic % × patient Hct ÷ 45 (the normal hematocrit). Reticulocytes released early in anemia (shift reticulocytes) stay longer in the blood, so the reticulocyte production index (RPI) divides the corrected count by the maturation time for the hematocrit: 1.0 day at Hct 40% or more, 1.5 at 35–39%, 2.0 at 25–34%, 2.5 at 15–24%, 3.0 under 15% (the Reticulocyte Count calculator uses these bands; some texts draw them slightly differently, so use the factor a question gives).
- RPI over 3 means an appropriate marrow response (hemolysis or blood loss); under 2 means an inadequate response to the anemia. Example: retic 6.0%, Hct 27% → 6.0 × 27 ÷ 45 = 3.6 corrected; 3.6 ÷ 2.0 = 1.8, inadequate.
- A high count in anemia means destruction, blood loss or recovery; a low or normal count in anemia means the marrow is not keeping up.
- IRF (immature reticulocyte fraction): the share of reticulocytes that are youngest, with the most RNA. The analyzer sorts reticulocytes into low, medium and high fluorescence; IRF is the medium plus high fractions. It rises early when the marrow speeds up and is an early marker of engraftment after stem cell transplant.
- Reticulocyte hemoglobin (CHr or Ret-He) shows the iron available to the marrow over the last 3–4 days. Values under 27 pg (CHr) point to iron-deficient erythropoiesis.
Erythrocyte sedimentation rate (ESR)
- What it measures: how far red cells fall in a vertical tube in one hour (mm/h). Proteins such as fibrinogen and immunoglobulins coat the cells, which then stack (rouleaux) and fall faster.
- Westergren method (ICSH reference): blood diluted 4 parts to 1 part sodium citrate, read at 60 minutes. Black-top citrate tubes or EDTA tubes; testing.com gives two hours from collection as the processing limit.
- Modified and alternate methods (automated, often on the EDTA CBC tube; some read in 20 minutes and convert to Westergren values) now outnumber Westergren: ICSH found 72% of laboratories using them and differences from Westergren of up to 142%, so each method needs its own reference values.
- Raised by inflammation (fibrinogen, globulins), paraproteins, anemia, pregnancy, older age and female sex.
- Lowered or unreliable with polycythemia, sickle cells, spherocytes and very high plasma viscosity; severe anemia also makes it unreliable.
- Typical Westergren values: men under 50, 0–15 mm/h; men 50 and over, 0–20; women under 50, 0–20; women 50 and over, 0–30. Ranges vary by laboratory and method.
Manual counts with the hemocytometer
- Cells/µL = cells counted × dilution factor ÷ (area counted in mm² × depth in mm). A Neubauer large square is 1 mm × 1 mm and the chamber is 0.1 mm deep, so each large square holds 0.1 µL.
- Example: 150 cells in 8 large squares (4 per side), dilution 1:20 → 150 × 20 ÷ (8 × 0.1) = 3,750 cells/µL (3.75 ×10³/µL).
- Count both sides; they must agree within the laboratory's limit before averaging.
- Count cells touching the boundary lines on two sides only (for example top and left) so none is counted twice.
- Giant platelets and clumps: a phase-microscopy manual platelet count, or a smear estimate, replaces the analyzer count.
Body fluid counts on analyzers
- Many hematology analyzers have a body fluid mode (for example for CSF, pleural fluid and ascites), reporting total nucleated cells, WBC, RBC and a simple differential.
- ICSH asks each laboratory to verify body fluid counting before use, the same way it verifies blood counts. Verification at the low counts typical of CSF (ASCP range 0–5/µL) matters most.
- Macrophages, mesothelial cells and tumor cells fall outside the WBC differential and are counted as high-fluorescence body fluid cells in the total nucleated count; the analyzer cannot identify them.
- The reference differential is manual: at least 100 cells on a cytocentrifuge slide (CLSI H56-A). Suspected malignant cells need slide review.
- Body fluid cells, especially in CSF, deteriorate quickly: count promptly.
Less common causes of spurious results
| Cause | Effect | Fix |
|---|---|---|
| Paraprotein (myeloma, Waldenström) | Hgb, MCH, MCHC falsely high; rarely WBC and platelets high | Plasma replacement and rerun |
| WBC clumping | WBC falsely low; no analyzer flags it | Vortex, warm to 37 °C, or citrate (× 1.1); confirm on a smear |
| Lysis-resistant red cells (Hb C, Hb S) | WBC and Hgb falsely high | Dilute and rerun, or longer lysis |
| Bacteria, yeast, malaria | Platelets (single bacteria) or WBC (clusters, large yeasts) falsely high | Smear estimate; tell the clinician if organisms are seen |
| Cytoplasmic fragments of leukemic cells | Platelets falsely high | Smear estimate |
| Megakaryocytes or carcinoma cells in blood | WBC falsely high | Correct like nRBCs: WBC × 100 ÷ (100 + cells per 100 WBCs) |
| Fat from a traumatic draw | WBC falsely high | Smear estimate, or recollect |
| Sample 1–4 days old at room temperature | MCV and Hct high; MCHC low | Reject, or report only unaffected results |
Related study sheets
- Red Cell Morphology and the AnemiasSize, color, shape and inclusions on the smear, anemia by MCV with the tests that sort it, hemoglobin variants and thalassemias, hemolytic anemias and erythrocytosis
- White Cells, Leukemia and LymphomaMaturation stages, reactive and inherited changes, the WHO 5th edition framework for myeloid and lymphoid neoplasms, cytochemistry, flow cytometry markers and PNH flow
- Hemostasis and Coagulation TestingPlatelets and the cascade, factors and inhibitors, PT/INR, aPTT, TT, fibrinogen, D-dimer, mixing studies, platelet function, anti-Xa, specimen rules and the bleeding and clotting disorders
Calculators for this topic
Sources
- Gulati G, Uppal G, Gong J. Unreliable automated complete blood count results: causes, recognition, and resolution. Ann Lab Med. 2022;42(5):515-530. doi:10.3343/alm.2022.42.5.515 Link
- Gulati G, Song J, Florea AD, Gong J. Purpose and criteria for blood smear scan, blood smear examination, and blood smear review. Ann Lab Med. 2013;33(1):1-7. doi:10.3343/alm.2013.33.1.1 Link
- Cattaneo M. Pseudothrombocytopenia and other conditions associated with spuriously low platelet counts. Haematologica. 2025;110(8):1677-1692. doi:10.3324/haematol.2024.286234 Link
- Krishnamurthy V, Shivamurthy A, Kumar PV. Platelet count in impedance-based hematology analyzer: beware of trap! Asian J Transfus Sci. 2023;17(1):131-132. doi:10.4103/ajts.ajts_65_21 Link
- Chhabra G. Automated hematology analyzers: recent trends and applications. J Lab Physicians. 2018;10(1):15-16. doi:10.4103/JLP.JLP_124_17 Link
- Yang K, Sun B, Zhang S, Pan Y, Fang J. RDW-SD is superior to RDW-CV in reflecting liver fibrosis stage in patients with chronic hepatitis B. Infect Drug Resist. 2023;16:6881-6891. doi:10.2147/IDR.S427047 Link
- Merck Manual Professional Version. Evaluation of anemia. Link
- Piva E, Brugnara C, Chiandetti L, Plebani M. Automated reticulocyte counting: state of the art and clinical applications in the evaluation of erythropoiesis. Clin Chem Lab Med. 2010;48(10):1369-1380 (abstract). doi:10.1515/CCLM.2010.292 Link
- Piva E, Brugnara C, Spolaore F, Plebani M. Clinical utility of reticulocyte parameters. Clin Lab Med. 2015;35(1):133-163 (abstract). doi:10.1016/j.cll.2014.10.004 Link
- Kratz A, Plebani M, Peng M, et al; ICSH. ICSH recommendations for modified and alternate methods measuring the erythrocyte sedimentation rate. Int J Lab Hematol. 2017;39(5):448-457 (abstract). doi:10.1111/ijlh.12693 Link
- Lorubbio M, Diamanti D, Pieroni C, et al. Erythrocyte sedimentation rate reference intervals determined via VES-MATIC 5 and CUBE 30 Touch with respect to the Westergren method. Diagnostics (Basel). 2025;15(9):1101. doi:10.3390/diagnostics15091101 Link
- Alende-Castro V, Alonso-Sampedro M, Vazquez-Temprano N, et al. Factors influencing erythrocyte sedimentation rate in adults: new evidence for an old test. Medicine (Baltimore). 2019;98(34):e16816. doi:10.1097/MD.0000000000016816 Link
- Testing.com (ADLM). Erythrocyte sedimentation rate (ESR). Link
- Bourner G, De la Salle B, George T, et al; ICSH. ICSH guidelines for the verification and performance of automated cell counters for body fluids. Int J Lab Hematol. 2014;36(6):598-612 (abstract). doi:10.1111/ijlh.12196 Link
- Yoon S, Kim HR. Analytical performance of the digital morphology analyzer Sysmex DI-60 for body fluid cell differential counts. PLoS One. 2023;18(7):e0288551. doi:10.1371/journal.pone.0288551 Link
- Chueh HW, Hwang SM, Shim YJ, et al; Korean RBC Disorder Working Party. Korean clinical practice guidelines for the diagnosis of hereditary hemolytic anemia. Blood Res. 2022;57(2):86-94. doi:10.5045/br.2022.2021224 Link
- Hillman RS. Characteristics of marrow production and reticulocyte maturation in normal man in response to anemia. J Clin Invest. 1969;48(3):443-453 (abstract). doi:10.1172/JCI106001 Link
- Kumar U, Chandra H, Gupta AK, Singh N, Chaturvedi J. Role of reticulocyte parameters in anemia of first trimester pregnancy: a single center observational study. J Lab Physicians. 2020;12(1):15-19. doi:10.1055/s-0040-1713585 Link
- Brownstein AJ, Wilkinson JD, Liang LL, et al. Immature reticulocyte fraction: a novel biomarker of hemodynamic severity in pulmonary arterial hypertension. Pulm Circ. 2024;14(3):e12421. doi:10.1002/pul2.12421 Link
- Borisova J, Morshchinin IV, Nazarova VI, et al. Low-cost microalgae cell concentration estimation in hydrochemistry applications using computer vision. Sensors (Basel). 2025;25(15):4651 (hemocytometer chamber depth and Neubauer square size). doi:10.3390/s25154651 Link
- ASCP Board of Certification. Medical Laboratory Scientist MLS(ASCP) and MLS(ASCPi) Examination Content Guideline, revised June 9, 2026 (hematology outline, calculations and composite reference ranges). Link