# Chemistry Methods and Instrumentation

Beer's law and photometers, optical methods, immunoassay formats and their interferences, electrodes, osmometry, electrophoresis, chromatography and mass spectrometry, automation, point of care, and specimen interferences

Checked against current sources November 2026

[![Page 1 of the Chemistry Methods and Instrumentation PDF, a printed page in columns with sections on Spectrophotometry and Beer's law, Other optical and atomic methods, Immunoassay formats and their interferences, Electrochemistry, Osmometry, electrophoresis and separation methods, Hemolysis, icterus and lipemia, Tubes, additives and order of draw and Automation, autoverification and point of care](https://medlabbasics.com/sheets/thumbs/lab-chemistry-methods.webp) Enlarge](https://medlabbasics.com/sheets/thumbs/lab-chemistry-methods-lg.webp)

**Printable PDF** Letter landscape, ready to print or keep with your notes

[View PDF](https://medlabbasics.com/sheets/lab-chemistry-methods.pdf) [Download](https://medlabbasics.com/sheets/lab-chemistry-methods.pdf)

## Spectrophotometry and Beer's law

- Beer's law: A = εbc. Absorbance (A) is proportional to concentration (c) when the path length (b, cm) and molar absorptivity (ε) are fixed.
- Absorbance = log (I0 / I), where I0 is light reaching the detector through the blank and I through the sample. From percent transmittance: A = 2 − log %T.
- Unknown concentration = (A unknown ÷ A standard) × concentration of the standard, valid only inside the method's linear range. Above it, dilute, re-read and multiply by the dilution factor (see the Beer's Law and Dilution Factor calculators).
- Zero the photometer with a reagent blank so that color from the reagent itself is not counted as analyte.

| Component | What it does | Exam points |
| --- | --- | --- |
| Light source | Deuterium lamp for ultraviolet, tungsten-halogen lamp for visible light | Wrong or aging lamp: low energy, noisy readings in its range |
| Monochromator | Prism or diffraction grating with slits isolates a narrow band of wavelengths | Narrow slit: better resolution but less light and more noise |
| Cuvette | Holds the sample in a fixed light path (usually 1 cm) | Quartz for ultraviolet (glass absorbs it); scratches and fingerprints add absorbance |
| Detector | Photodiode or photomultiplier tube turns light into an electrical signal | Photomultiplier: most sensitive, for low light |
| Double-beam design | Measures reference and sample paths at the same time | Cancels lamp drift and fluctuations |

| Check | Material | A failure means |
| --- | --- | --- |
| Wavelength accuracy | Holmium oxide glass or solution: sharp bands at certified wavelengths | Every band shifted by the same amount: recalibrate the wavelength scale |
| Photometric accuracy and linearity | Neutral-density (nearly neutral absorbing) glass filters or certified absorbance solutions | Absorbance values wrong even at the right wavelength |
| Stray light | Cutoff solutions or filters that transmit essentially no light at the test wavelength (light still detected there is stray light) | Light outside the band reaches the detector; high absorbances read low and the curve bends |

**Exam tip** Absorbance above the linear range underestimates the concentration. Never report it: dilute until the reading falls in range, then multiply by the total parts (1 part sample + 4 parts diluent = a factor of 5).

## Other optical and atomic methods

| Method | What is measured | Typical uses | Watch for |
| --- | --- | --- | --- |
| Flame emission photometry | Light emitted at the element's own wavelength by atoms excited in a flame | Sodium, potassium, lithium (largely replaced by ISE) | Sample is diluted, so lipemia and high protein lower the result (volume displacement) |
| Atomic absorption (flame or graphite furnace) | Light from a hollow cathode lamp of the element, absorbed by ground-state atoms | Lithium, trace metals; graphite furnace for very low levels such as lead | One element per lamp; being replaced by ICP-MS |
| Fluorometry | Light emitted when a molecule excited at one wavelength returns to ground state; detector at 90 degrees | Fluorescent immunoassays, time-resolved fluorescence | Self-quenching and inner-filter effects at high concentration; background fluorescence (read a blank) |
| Chemiluminescence | Light from a chemical reaction; no excitation lamp needed | Most automated immunoassays (acridinium or ruthenium labels, or enzyme labels with luminol or dioxetane substrates) | Very wide dynamic range and high sensitivity |
| Turbidimetry | Loss of transmitted light (absorbance) through a suspension of immune complexes | CRP, immunoglobulins, latex-enhanced assays on general chemistry analyzers | Lipemia; antigen excess (prozone) reads falsely low |
| Nephelometry | Light scattered by immune complexes at a fixed angle (for example 13 to 24 degrees) | Immunoglobulins, complement, free light chains | Lipemia; antigen excess; kinetic readings help detect excess |
| ICP-MS | Elements ionized in an argon plasma and counted by mass | Trace and toxic elements, many in one run | Costly; spectral interferences must be controlled |

## Immunoassay formats and their interferences

| Feature | Competitive (reagent-limited) | Sandwich, two-site (reagent excess) |
| --- | --- | --- |
| Design | Labeled analyte competes with the patient's analyte for a limited amount of antibody | Capture antibody and labeled detection antibody bind two sites on the analyte |
| Signal vs concentration | Inverse: more analyte, less signal | Direct: more analyte, more signal |
| Typical analytes | Small molecules: drugs, steroids, free T4 and free T3 | Large molecules: TSH, troponin, hCG, ferritin, prolactin |
| High-dose biotin (streptavidin-biotin capture) | Falsely HIGH | Falsely LOW |
| Heterophile antibodies, HAMA, rheumatoid factor | Can bind the conjugate or label: error in either direction | Bridge capture to detection antibody: usually falsely high |
| High-dose hook effect | Not the classic problem | Antigen excess, especially in one-step assays: falsely low; dilution reveals the true value |

- Heterogeneous assays separate bound from free label (a wash step) before reading; homogeneous assays read without separation because binding itself changes the signal.
- Cross-reactivity: structurally similar compounds or metabolites are read as the analyte (cortisol assays and fludrocortisone, cyclosporine metabolites, digoxin-like factors). Drug screens detect a class by cross-reactivity, so positives are confirmed by GC-MS or LC-MS/MS.
- Immunoassays are generally unaffected by hemolysis and icterus; lipemia matters most for nephelometric and turbidimetric methods.

**Safety** Biotin above 1 mg a day (hair and nail supplements up to 10 mg, much more for multiple sclerosis) can give a falsely LOW troponin on affected sandwich assays and a Graves-like thyroid pattern (low TSH, high free T4). An undetectable troponin in a patient with a proven infarction must be questioned before release.

**Exam tip** Clues to interference: results that do not fit the patient, disagreement between platforms, and a corrected result that changes with dilution. Rising corrected results on dilution = hook effect; falling corrected results on dilution or a large drop after blocking reagent = heterophile antibody.

## Electrochemistry

| Method | Principle | Clinical examples |
| --- | --- | --- |
| Potentiometry (ion-selective electrodes) | Voltage between an indicator and a reference electrode at almost zero current; voltage changes with the log of ion activity (Nernst equation) | Na, K, Cl, Li, ionized Ca; pH (glass electrode); pCO2 (Severinghaus electrode) |
| Severinghaus pCO2 electrode | CO2 diffuses through a membrane into a bicarbonate solution and changes its pH, read by a glass pH electrode | Blood gas pCO2 |
| Amperometry | Current at a fixed applied voltage, proportional to the amount of substance oxidized or reduced | pO2 (Clark electrode), glucose biosensors and meters |
| Clark pO2 electrode | Oxygen crosses a permeable membrane and is reduced at a platinum cathode held at about −0.8 V; silver/silver chloride anode | Blood gas pO2 |
| Coulometry | A silver anode releases silver ions that precipitate chloride as AgCl; when chloride is used up, free silver ions raise the current and stop the titration; charge (time) is proportional to chloride | Chloridometer: sweat chloride, serum or urine chloride |
| Conductometry | Electrical conductivity of the whole solution, not one ion | Sweat conductivity: screening only, never diagnostic for cystic fibrosis |

| Feature | Direct ISE | Indirect ISE |
| --- | --- | --- |
| Sample | Undiluted | Diluted (about 1:5 to 1:46) before measurement |
| Measures | Ion activity in plasma water | Ion in the diluted whole sample, assuming plasma is about 93% water |
| Where | Blood gas and point-of-care analyzers, some dry-slide systems | High-throughput automated chemistry analyzers |
| Very high lipids or protein | Unaffected | Falsely LOW Na, K and Cl (electrolyte exclusion): pseudohyponatremia |

**Exam tip** Low sodium on the chemistry analyzer with severe lipemia or a paraprotein, a normal measured osmolality, and a normal sodium on the blood gas analyzer = pseudohyponatremia from indirect ISE. Bromide is read as chloride: very high chloride and a negative anion gap.

## Osmometry, electrophoresis and separation methods

| Method | Principle | Exam points |
| --- | --- | --- |
| Freezing point osmometer | Dissolved particles lower the freezing point of water in proportion to osmolality (colligative property) | The usual clinical method; detects ethanol, methanol and ethylene glycol, so it is the one used for the osmolal gap |
| Vapor pressure osmometer | Dissolved particles lower the vapor pressure (dew point) | Misses volatile alcohols, so an osmolal gap from it is unreliable for toxic alcohols |
| Electrophoresis (agarose gel) | Proteins move in an electric field by charge and size (usually in barbital buffer near pH 8.6), are stained and scanned | Fractions: albumin, alpha-1, alpha-2, beta (beta-1, beta-2), gamma; abnormal band typed by immunofixation |
| Capillary zone electrophoresis | Separation in a narrow capillary filled with buffer; proteins are detected, not stained | Automated, fast; abnormal peak typed by immunosubtraction |
| HPLC | Liquid mobile phase pumped through a packed column; compounds separate by how strongly they partition into the stationary phase (normal or reverse phase) | Reference method for drugs such as cyclosporine; detectors include UV and mass spectrometry |
| Gas chromatography | Gas mobile phase carries volatile, heat-stable compounds through a column; identified by retention time | Volatile alcohols; GC-MS confirms drugs of abuse with library-matched spectra |
| LC-MS/MS (tandem mass spectrometry) | LC separation, electrospray ionization, then two mass filters select a precursor ion and its product ion | Most specific routine method: steroids, vitamin D, drugs, newborn screening; stable isotope internal standards correct ion suppression |
| MALDI-TOF MS | Sample mixed with matrix, hit by a laser; ions fly down a tube and lighter ions arrive first; ribosomal protein spectrum is matched to a database | Identifies bacteria and yeast from colonies in minutes; E. coli and Shigella are not separated |

**Exam tip** SPEP is run on serum. Plasma, or serum from a tube that did not clot fully, keeps fibrinogen, which forms a narrow band at the border of beta and gamma that can be mistaken for a monoclonal protein; immunofixation settles it.

## Hemolysis, icterus and lipemia

| Interferent | How it interferes | Tests most affected | What to do |
| --- | --- | --- | --- |
| Hemolysis | Release of cell contents; hemoglobin color; chemical reaction with reagents | Potassium, LD, AST raised; bilirubin can be lowered | Check the H-index; recollect for affected tests; report with a comment if policy allows |
| Icterus | Bilirubin color (spectral) and reaction with reagents | Peroxidase-coupled methods (cholesterol, triglyceride, uric acid) and creatinine read low; co-oximetry | Check the I-index; use assay limits; dilution or another method |
| Lipemia | Light scatter and absorption (worst at short wavelengths); volume displacement; uneven layering | NADH methods read at 340 nm (ALT, AST, glucose); Na, K, Cl by indirect ISE; nephelometry | Check the L-index (visual grading is unreliable); high-speed or ultracentrifugation, then test the infranatant |

**Safety** Lipid removal is not acceptable for triglycerides or for hydrophobic analytes (some drugs and hormones) that move into the lipid layer: run those on the original specimen first, or recollect after the patient has fasted.

## Tubes, additives and order of draw

| Tube (usual stopper) | Additive | Effects to know |
| --- | --- | --- |
| Light blue | 3.2% (0.109 M) sodium citrate, 1 part to 9 parts blood | Coagulation; chelation can inhibit AST and ALP |
| Red, gold (SST) | Clot activator, with or without separator gel | Gel can absorb hydrophobic drugs (phenytoin, phenobarbital, carbamazepine, lidocaine): 20 to 50% lower after 24 h at 4 °C; very dense serum may not float above gel |
| Green | Sodium or lithium heparin | Heparin binds ions, so balanced heparin is used for ionized calcium; no lithium heparin for lithium levels |
| Lavender | K2 or K3 EDTA | Hematology; chelates Ca, Mg and Zn: carried into other tubes it gives falsely high K and low Ca and Mg |
| Gray | Sodium fluoride with potassium oxalate (or EDTA) | Fluoride blocks enolase, but glycolysis goes on for 1 to 4 hours; oxalate shrinks cells and inhibits several enzymes |

1. Blood culture bottles (sterile first)
2. Non-additive tube (or discard tube)
3. Sodium citrate (coagulation)
4. Serum tubes: clot activator, then serum separator
5. Heparin, including plasma separator tubes
6. EDTA
7. Fluoride/oxalate

**Exam tip** A very high potassium with a very low calcium in a patient who looks well points to EDTA contamination, not hyperkalemia. Ask for a recollection; do not report the potassium as a critical value without checking.

## Automation, autoverification and point of care

| Area | What it covers | Exam points |
| --- | --- | --- |
| Total laboratory automation | Preanalytical: tracks, barcode reception, sorting, centrifugation, aliquoting. Analytical: analyzers linked by middleware. Postanalytical: autoverification, result transfer, storage | Fewer manual steps, shorter and more predictable turnaround |
| Random access analyzer | Any test on any specimen in any order | Unlike batch analyzers that run one test on a group of specimens |
| Autoverification | Rules in middleware or the LIS release results without a person | Rules use instrument flags, QC status, serum indices, critical values, delta checks, measuring range and moving averages; each laboratory sets and validates its own limits (CLSI AUTO10 and AUTO15) |
| Point-of-care testing | Handheld strips and lateral flow devices, cartridges, small bench analyzers near the patient | Waived tests must follow the manufacturer's instructions (42 CFR 493.15); operator training and QC records matter most |
| Glucose meters | Amperometric strips with glucose oxidase or glucose dehydrogenase on whole blood | High hematocrit reads low, anemia reads high; oxygen therapy lowers glucose oxidase results; maltose (icodextrin dialysate) falsely raises some GDH strips |

**Safety** A meter reading that does not fit the patient (for example a high reading in a sweaty, confused patient on icodextrin dialysate) is confirmed at once with a laboratory plasma glucose by a glucose-specific method, and the laboratory value is acted on.

## In depth

The summary above is what the PDF prints. These sections go further.

### Beer's law worked examples

1. Within range. A 200 mg/dL standard reads A = 0.400 and a patient reads A = 0.300 against the same reagent blank. Concentration = 0.300 ÷ 0.400 × 200 = 150 mg/dL.
2. Above range. The method is linear to A = 0.800 and a patient reads 1.250. The sample is diluted 1 part plus 4 parts saline (factor 5) and reads 0.250. 0.250 ÷ 0.400 × 200 = 125 mg/dL; × 5 = 625 mg/dL. The first reading (1.250) would have given 625 only if the curve stayed straight, which it does not.
3. Transmittance. A = 2 − log %T, so 100%T gives A = 0, 10%T gives A = 1.0 and 1%T gives A = 2.0. Half the light through (50%T) is A = 0.301. Absorbance, not %T, is proportional to concentration.
4. Molar absorptivity. If ε is known, c = A ÷ (ε × b). A compound with ε = 6,220 L/mol/cm (the usual NADH value at 340 nm) giving A = 0.311 in a 1 cm cuvette is at 0.311 ÷ 6,220 = 5.0 × 10⁻⁵ mol/L (50 µmol/L).

Why curves bend at the top: at high concentration molecules interact and scatter light, and stray light (light outside the selected band) is a larger share of the little light that still reaches the detector, so absorbance rises less than concentration. Lipemia, turbidity and bubbles add apparent absorbance that is not analyte. Wavelength error moves the reading off the absorbance peak, lowering sensitivity and linearity.

**Exam tip** Endpoint vs kinetic: an endpoint method reads once after the reaction is complete; a kinetic (rate) method reads the change in absorbance per minute, which is how enzyme activities and many NADH-linked assays are measured. A falling absorbance at 340 nm means NADH is being used up.

### Immunoassay interference: recognizing and resolving it

| Clue | Likely cause | Next step |
| --- | --- | --- |
| Result far higher than the clinical picture; different on another platform | Heterophile antibody or human anti-animal antibody (HAMA) | Serial dilution (nonlinear recovery), heterophile blocking reagent, PEG precipitation, another platform with different antibodies |
| Result low or only mildly raised while the patient has a large tumor or very high level expected | High-dose hook effect | Repeat at a dilution (for example 1:10 to 1:100); a rising corrected value confirms it |
| Sandwich results low and competitive results high in the same patient | Biotin | Ask about supplements; retest after the manufacturer's washout, on a biotin-free platform, or after streptavidin bead depletion |
| High prolactin without symptoms | Macroprolactin (prolactin bound to IgG) | PEG precipitation or gel filtration to estimate monomeric prolactin |
| Drug level higher than the parent drug can explain | Cross-reacting metabolites (cyclosporine immunoassays up to 174% above HPLC) | Chromatographic or LC-MS/MS method |
| Digoxin result after anti-digoxin Fab treatment | Therapeutic antibody fragments bind digoxin in the sample | Report that the result is affected: the fragments compete with the assay antibody |

Hook effect examples from the literature: a prolactin of 1,500 mU/L rose to 950,000 mU/L when the sample was diluted, and a urine myoglobin reading 100 nmol/L undiluted read 1,496 nmol/L at 1:50. Thyroglobulin is recommended to be run neat and at 1:10. The hook occurs in two-site assays where capture and detection antibodies are added together: excess free analyte saturates both antibodies separately, so fewer sandwiches form.

Heterophile antibodies are polyreactive natural antibodies or autoantibodies, including rheumatoid factor; HAMA follow exposure to mouse antibodies. In a two-site assay they link the animal capture and detection antibodies with no analyte present. Manufacturers add nonimmune animal immunoglobulin as a blocker, but not every interference is blocked. The laboratory reports suspected interference to the provider, comments the result, and keeps the patient's history on file.

Biotin: most multivitamins contain about 30 µg, which does not interfere. Supplements for hair and nails contain up to 10 mg, and multiple sclerosis studies have used far higher doses. Many laboratories ask patients to stop biotin for at least 48 hours before collection; the assay's package insert gives the interval and the biotin level that interferes, and some troponin assays listed by the FDA have not addressed the risk.

### Electrodes in more depth

- Nernst equation: E = E0 + (RT ÷ zF) ln a. For a monovalent ion at 25 °C this is about 59 mV per tenfold change in activity; for a divalent ion (Ca²⁺) about half that. ISEs measure activity, which is why ionized calcium is a direct ISE measurement.
- Selectivity: an ISE membrane (glass, crystal or polymer with an ionophore) favors one ion but is not perfectly selective; other halides such as bromide are read as chloride by chloride electrodes and colorimetric methods.
- pH electrode: glass membrane sensitive to H⁺, read against a reference electrode of constant potential.
- pCO2: the Severinghaus electrode is a pH electrode behind a CO2-permeable membrane with a bicarbonate solution; the measured pH change is converted to pCO2. Bicarbonate on a blood gas report is calculated from pH and pCO2.
- pO2: the Clark electrode reduces O2 at a platinum cathode (about −0.8 V); the current is proportional to pO2. Very high leukocyte counts keep consuming oxygen in a delayed syringe (see the [Electrolytes, Acid-Base and Blood Gases](https://medlabbasics.com/sheets/lab-chemistry-electrolytes-abg.html) page).
- Coulometric chloride: Ag → Ag⁺ + e⁻ at the silver anode; Ag⁺ + Cl⁻ → AgCl. The end point is the sudden rise in free Ag⁺. ICP-MS and ion chromatography are also acceptable for sweat chloride; sweat conductivity and sweat sodium or osmolality are not diagnostic.
- Lithium: flame emission and atomic absorption were the original methods; lithium ISEs are precise but other ions can interfere.

Indirect ISE assumes a normal plasma water fraction. When lipids or proteins take up more of the plasma volume, the same aliquot holds less water and fewer electrolytes, and the dilution multiplies the error. Direct ISE senses the activity in plasma water and is unaffected. Very high total protein (myeloma, macroglobulinemia) causes most of the sodium bias, and high triglycerides add to it; potassium errors are much smaller. The laboratory confirms a doubtful sodium on a direct ISE analyzer or by measured osmolality.

### Osmolality and the osmolal gap

- Calculated osmolality = 2 × Na + glucose ÷ 18 + BUN ÷ 2.8 (glucose and BUN in mg/dL). Osmolal gap = measured − calculated; about 10 mOsm/kg or more suggests unmeasured osmoles (ethanol, methanol, ethylene glycol, mannitol, acetone). See the Osmolality and Osmolar Gap calculators.
- Freezing point depression: one osmole of particles per kg of water lowers the freezing point by about 1.86 °C. The instrument supercools the sample, induces freezing and reads the plateau temperature with a thermistor.
- Vapor pressure (dew point) osmometers may not detect volatile solutes such as ethanol and methanol: a patient with a high methanol level had no osmolal gap by vapor pressure osmometry, so the gap is sought only with a freezing point osmometer.
- Serum and heparin plasma both work; specimens are stable at room temperature for up to 36 hours before separation, with cooling advised for longer delays.
- One large study found the gap by freezing point rose with glucose above 110 mg/dL, most strongly above 450 mg/dL, while the vapor pressure gap did not: interpret gaps cautiously in severe hyperglycemia.

### Mass spectrometry and trace elements

| Part | What it does | Point to know |
| --- | --- | --- |
| Electrospray ionization (ESI) | Soft ionization of compounds leaving the LC column | Little fragmentation; multiply charged ions let large proteins be measured |
| APCI | Chemical ionization at atmospheric pressure | For neutral, low-polarity compounds such as steroids |
| Electron impact (GC-MS) | Hard ionization | Extensive, reproducible fragmentation: spectra match libraries across instruments |
| Quadrupole | Four rods pass a narrow band of mass-to-charge (m/z) values | Triple quadrupole: Q1 selects precursor, Q2 fragments it (collision cell), Q3 selects product |
| Multiple reaction monitoring | Pairs of precursor and product ions for each analyte | Example: 25-hydroxyvitamin D3 401 > 383 m/z |
| Ion suppression | Other sample components compete for ionization | Corrected by a stable isotope-labeled internal standard and sample cleanup |
| Time of flight (MALDI-TOF) | Ions accelerated down a flight tube; flight time gives m/z | 337 nm nitrogen laser; matrix absorbs the energy and protects the sample |

Direct injection tandem MS without chromatography (as in newborn screening) cannot separate isomers such as leucine and isoleucine, and suppression limits it to higher concentrations. LC-MS/MS product-ion spectra vary between instruments, so library searching is easier with GC-MS.

| Trace element method | Strengths | Limits |
| --- | --- | --- |
| Flame atomic emission | Cheap, simple | One element, limited range, high detection limit, flammable gases |
| Flame atomic absorption | Cheap, few interferences | One element per lamp, high detection limit, larger sample |
| Graphite furnace atomic absorption | Low detection limit, small sample | One element, slow, some need acid digestion |
| ICP-MS | Many elements at once, wide range, low detection limit, high throughput | Cost, argon and gases, expert staff, interferences to control |

### Serum indices and lipemia management

- Analyzers dilute the sample and read absorbance at several wavelengths to report hemolysis (H), icterus (I) and lipemia (L) indices. Bilirubin and hemoglobin absorb in the lower range; only turbidity absorbs near 700 nm, so lipemia is read there.
- Visual grading agrees poorly between staff and with the L-index; indices are better and can drive autoverification rules.
- A high L-index with little triglyceride can come from a paraprotein; a very high triglyceride with clear serum and a normal L-index points to free glycerol (glycerol kinase deficiency, glycerol in drugs or nutrition) measured by a non-blanked enzymatic triglyceride method.
- Lipid layers: chylomicrons cause most turbidity. Ultracentrifuges (100,000 x g and above) remove lipid best; high-speed centrifugation at about 10,000 x g is almost as good. VLDL-rich lipemia may need repeated spins. Extraction reagents (polyethylene glycol, cyclodextrin, commercial clearing agents) are alternatives with their own limits.
- Interference thresholds in package inserts are a starting point; laboratories can set their own after testing. A small group of severely icteric patients accounts for about half of icteric index failures.

### Point-of-care quality

- CLIA waived tests are simple methods cleared for home use, or so simple and accurate that errors are negligible, or with no reasonable risk of harm if done wrong. The list includes non-automated dipstick [urinalysis](https://medlabbasics.com/sheets/lab-urinalysis.html), fecal occult blood, visual urine pregnancy tests, home-use glucose meters, spun microhematocrit and single-analyte hemoglobin instruments.
- Cartridge analyzers: run the electronic QC check each day of use before patients; if there is none, run cartridge controls daily. Check each delivery of cartridges or strips and run a QA sample at least monthly to check storage and operators.
- Glucose meters read whole blood, which has less water than plasma: plasma glucose is about 11 to 12% higher at a hematocrit of 45%. Strip enzymes are damaged by heat, cold and humidity; strips and meters must be stored as directed.
- Glucose oxidase meters: high pO2 gives low results and hypoxia high results; acetaminophen and ascorbic acid lower results on peroxide-based strips. Some glucose dehydrogenase strips (GDH-PQQ) read maltose, formed from icodextrin dialysate, as glucose.
- Unfixed samples keep consuming glucose at about 5 to 7% per hour; fluoride takes 1 to 2 hours or more to stop glycolysis, so plasma is separated promptly for comparisons with meters.

**Exam tip** Waived does not mean unregulated: the site must hold a certificate of waiver and follow the manufacturer's instructions exactly.

[Practice free in your browser](https://medlabbasics.com/app/?from=/sheets/lab-chemistry-methods.html) 600+ exam-style questions with rationales. Free today, no account, no ads. Pro is coming in January 2027. Also free [on Google Play](https://play.google.com/store/apps/details?id=com.alex.medlabbasics).

## Related study sheets

- [**Clinical Chemistry: Analytes and Disease Patterns** Glucose and HbA1c, lipids, proteins and electrophoresis, nonprotein nitrogen, bilirubin and jaundice, liver and other enzymes, cardiac and tumor markers, porphyrins](https://medlabbasics.com/sheets/lab-chemistry-analytes.html)
- [**Electrolytes, Acid-Base and Blood Gases** Sodium, potassium, chloride and CO2, the anion and osmolal gaps, calcium, magnesium, phosphate and iron studies, the four acid-base disorders with expected compensation, blood gas electrodes, CO-oximetry and specimen errors](https://medlabbasics.com/sheets/lab-chemistry-electrolytes-abg.html)
- [**Endocrinology, Drug Monitoring and Toxicology** Thyroid, adrenal, pituitary and parathyroid patterns, stimulation and suppression tests, the hook effect, peak and trough timing, therapeutic ranges, the osmolal gap, acetaminophen, salicylate, lead, carbon monoxide and drug screens](https://medlabbasics.com/sheets/lab-chemistry-endocrine-tdm-tox.html)
- [**Reference Ranges the Exam Uses** The ASCP BOC composite reference ranges for chemistry, blood gases, hematology, CSF, semen and urine in conventional and SI units, and the calculations the MLS and MLT exams expect](https://medlabbasics.com/sheets/lab-reference-intervals.html)

## Calculators for this topic

- [Anion Gap](https://medlabbasics.com/calculators/anion-gap.html)
- [Osmolar Gap](https://medlabbasics.com/calculators/osmolar-gap.html)
- [Serum Osmolality](https://medlabbasics.com/calculators/serum-osmo.html)

## Sources

1. Tate J, Ward G. Interferences in immunoassay. Clin Biochem Rev. 2004;25(2):105-120 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC1904417/)
2. Gifford JL, Sadrzadeh SMH, Naugler C. Biotin interference: underrecognized patient safety risk in laboratory testing. Can Fam Physician. 2018;64(5):370 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC5951654/)
3. Russwurm M, Wild J, Gobel J. Biotin interference in thyroid function tests. Dtsch Arztebl Int. 2025;122(20):557. doi:10.3238/arztebl.m2025.0114 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC12620895/)
4. US Food and Drug Administration. Biotin interference with troponin lab tests: assays subject to biotin interference [Link](https://www.fda.gov/medical-devices/in-vitro-diagnostics/biotin-interference-troponin-lab-tests-assays-subject-biotin-interference)
5. Hunsaker JJH, La'ulu SL, Zupan E, et al. Detection of a high-dose hook effect and evaluation of dilutions of urine myoglobin specimens using a serum myoglobin assay. Ann Lab Med. 2024;44(4):367-370. doi:10.3343/alm.2023.0427 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC10961618/)
6. Nikolac N. Lipemia: causes, interference mechanisms, detection and management. Biochem Med (Zagreb). 2014;24(1):57-67. doi:10.11613/BM.2014.008 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC3936974/)
7. Bowen RA, Remaley AT. Interferences from blood collection tube components on clinical chemistry assays. Biochem Med (Zagreb). 2014;24(1):31-44. doi:10.11613/BM.2014.006 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC3936985/)
8. Ho CKM, Chen C, Setoh JWS, et al. Optimization of hemolysis, icterus and lipemia interference thresholds for 35 clinical chemistry assays. Pract Lab Med. 2021;25:e00232. doi:10.1016/j.plabm.2021.e00232 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC8145753/)
9. Mainali S, Merrill AE, Krasowski MD. Frequency of icteric interference in clinical chemistry laboratory tests and causes of severe icterus. Pract Lab Med. 2021;27:e00259. doi:10.1016/j.plabm.2021.e00259 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC8567001/)
10. Bazzano G, Galazzi A, Giusti GD, et al. The order of draw during blood collection: a systematic literature review. Int J Environ Res Public Health. 2021;18(4):1568. doi:10.3390/ijerph18041568 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC7915193/)
11. Chopra P, Bharti G, Gupta A, et al. Impact of total protein and triglycerides on the comparability of sodium and potassium measurements by direct and indirect ion-selective electrodes. EJIFCC. 2026;37(4):696-709 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC13563522/)
12. Wang Y, Xu H, Zhang J, Li G. Electrochemical sensors for clinic analysis. Sensors (Basel). 2008;8(4):2043-2081. doi:10.3390/s8042043 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC3673406/)
13. Bernasconi S, Angelucci A, De Cesari A, et al. Recent technologies for transcutaneous oxygen and carbon dioxide monitoring. Diagnostics (Basel). 2024;14(8):785. doi:10.3390/diagnostics14080785 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC11049249/)
14. Taylor PJ, Bouska RA. Automated serum chloride analysis using the Apple computer. J Automat Chem. 1988;10(1):10-14. doi:10.1155/S1463924688000057 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC2547731/)
15. Massie J, Greaves R, Metz M, et al. Australasian guideline (2nd edition): an annex to the CLSI and UK guidelines for the performance of the sweat test for the diagnosis of cystic fibrosis. Clin Biochem Rev. 2017;38(3):115-130 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC5759161/)
16. Dubot P, Liang J, Dubs J, et al. Sweat chloride quantification using capillary electrophoresis. Pract Lab Med. 2019;13:e00114. doi:10.1016/j.plabm.2018.e00114 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC6317274/)
17. Lugassy D, Nelson L. Bromism: forgotten, but not gone. J Med Toxicol. 2009;5(3):151-157 (abstract) [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC3550397/)
18. Tonyushkina K, Nichols JH. Glucose meters: a review of technical challenges to obtaining accurate results. J Diabetes Sci Technol. 2009;3(4):971-980. doi:10.1177/193229680900300446 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC2769957/)
19. St John A, Price CP. Existing and emerging technologies for point-of-care testing. Clin Biochem Rev. 2014;35(3):155-167 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC4204237/)
20. Martin CL. Quality control issues in point of care testing. Clin Biochem Rev. 2008;29(Suppl 1):S79-S82 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC2556589/)
21. 42 CFR 493.15, Laboratories performing waived tests (eCFR, read 2026-11-15) [Link](https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493/subpart-A/section-493.15)
22. Choy KW, Wijeratne N, Lu ZX, Doery JC. Harmonisation of osmolal gap: can we use a common formula? Clin Biochem Rev. 2016;37(3):113-119 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC5111243/)
23. Lin KY, Chen PC, Chen SC, et al. Evaluation of the effects of glucose on osmolal gap using freezing point depression and vapor pressure methods. Kaohsiung J Med Sci. 2018;34(7):409-414. doi:10.1016/j.kjms.2018.03.002 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC11915567/)
24. Arif M. UV-visible spectrophotometry in modern research. RSC Adv. 2026;16(43):52112-52147. doi:10.1039/d6ra03975h [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC13573634/)
25. Pudasaini S, Dangal P, Rathore RS, et al. A portable, low-cost spectrophotometer for absorbance and selected-wavelength fluorescence measurements. J Anal Methods Chem. 2026;2026:9371485. doi:10.1155/jamc/9371485 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC13717223/)
26. Allen DW. Holmium oxide glass wavelength standards. J Res Natl Inst Stand Technol. 2007;112(6):303-306. doi:10.6028/jres.112.024 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC4655923/)
27. Natali P, Bedin R, Bernardi G, et al. Inter-laboratory concordance of cerebrospinal fluid and serum kappa free light chain measurements. Biomolecules. 2022;12(5):677. doi:10.3390/biom12050677 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC9138559/)
28. Mouliou DS. C-reactive protein: pathophysiology, diagnosis, false test results and a novel diagnostic algorithm for clinicians. Diseases. 2023;11(4):132. doi:10.3390/diseases11040132 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC10594506/)
29. Cinquanta L, Fontana DE, Bizzaro N. Chemiluminescent immunoassay technology: what does it change in autoantibody detection? Auto Immun Highlights. 2017;8(1):9. doi:10.1007/s13317-017-0097-2 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC5483212/)
30. Pitt JJ. Principles and applications of liquid chromatography-mass spectrometry in clinical biochemistry. Clin Biochem Rev. 2009;30(1):19-34 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC2643089/)
31. Clark AE, Kaleta EJ, Arora A, Wolk DM. Matrix-assisted laser desorption ionization-time of flight mass spectrometry: a fundamental shift in the routine practice of clinical microbiology. Clin Microbiol Rev. 2013;26(3):547-603. doi:10.1128/CMR.00072-12 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC3719498/)
32. Wilschefski SC, Baxter MR. Inductively coupled plasma mass spectrometry: introduction to analytical aspects. Clin Biochem Rev. 2019;40(3):115-133. doi:10.33176/AACB-19-00024 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC6719745/)
33. Aliasgharpour M, Hagani H. Evaluation of lithium determination in three analyzers: flame emission, flame atomic absorption spectroscopy and ion selective electrode. N Am J Med Sci. 2009;1(5):244-246. doi:10.4297/najms.2009.5244 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC3364617/)
34. Nam Y, Park HD. Revolutionizing laboratory practices: pioneering trends in total laboratory automation. Ann Lab Med. 2025;45(5):472-483. doi:10.3343/alm.2024.0581 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC12370808/)
35. Gul BU, Ozcan O, Dogan S, Arpaci A. Designing and validating an autoverification system of biochemical test results in Hatay Mustafa Kemal University, clinical laboratory. Biochem Med (Zagreb). 2022;32(3):030704. doi:10.11613/BM.2022.030704 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC9344865/)
36. Jamal I, Shuchismita, Raman RB, Choudhary V. Peaks and distortions: evaluation of irregularities in capillary serum protein electrophoresis. Indian J Hematol Blood Transfus. 2025;41(2):349-356. doi:10.1007/s12288-024-01853-4 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC11992269/)
37. Ren Q, Liu T, Li J, et al. Bisalbuminemia: a rare finding identified via serum protein agarose gel electrophoresis and capillary electrophoresis techniques. EJIFCC. 2026;37(3):425-431 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC13543513/)
38. Testing.com (ADLM). Protein electrophoresis and immunofixation electrophoresis [Link](https://www.testing.com/tests/protein-electrophoresis-immunofixation-electrophoresis/)
39. Reule AG. Errors in spectrophotometry and calibration procedures to avoid them. J Res Natl Bur Stand A Phys Chem. 1976;80A(4):609-624. doi:10.6028/jres.080a.060 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC5293527/)
40. Balzer AHA, Whitehurst CB. An analysis of the biotin-(strept)avidin system in immunoassays: interference and mitigation strategies. Curr Issues Mol Biol. 2023;45(11):8733-8754. doi:10.3390/cimb45110549 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC10670868/)
41. Walker JA, Schwartzbard A, Krauss EA, et al. The missing gap: a pitfall in the diagnosis of alcohol intoxication by osmometry. Arch Intern Med. 1986;146(9):1843-1844 (abstract). doi:10.1001/archinte.146.9.1843 [Link](https://europepmc.org/article/MED/3753127)
42. Brizzee L, Stone A, Palmer MC. False lithium toxicity secondary to lithium heparin test tube: a case report and review. Ment Health Clin. 2020;10(3):90-94. doi:10.9740/mhc.2020.05.090 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC7213950/)
43. Upadhyay S. An optimized enzyme-coupled spectrophotometric method for measuring pyruvate kinase kinetics. Bio Protoc. 2025;15(16):e5415. doi:10.21769/bioprotoc.5415 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC12378421/)
44. Adedeji AL, Adenikinju RO, Ajele JO, Olawoye TL. Serum protein electrophoresis under effective control of HIV-1 disease progression. EXCLI J. 2014;13:761-771 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC4464463/)
45. Sulaiman A, Caturegli P. Falsely abnormal serum protein electrophoresis after administration of intravenous immunoglobulins (IVIG): a retrospective cohort study. Pract Lab Med. 2024;42:e00434. doi:10.1016/j.plabm.2024.e00434 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC11616521/)