# Quality Control, Validation and Regulation | MedLab Basics

Web page: https://medlabbasics.com/sheets/lab-qc-quality.html

Study sheet

# Quality Control, Validation and Regulation

Levey-Jennings charts and Westgard rules, sigma, method verification, calibration, reference intervals, proficiency testing, competency, CLIA, POCT, error phases and reporting

Checked against current sources November 2026

[Practice 42 Quality Control, Validation and Regulation questions free](https://medlabbasics.com/app/?from=/sheets/lab-qc-quality.html#/topic/lab-qc-quality)

[![Page 1 of the Quality Control, Validation and Regulation PDF, a printed page in columns with sections on QC statistics, Westgard multirules, Validation, calibration and reference intervals, Proficiency testing and competency, CLIA, accreditation and POCT, Errors, reporting and turnaround time and Management and education (MLS only)](https://medlabbasics.com/sheets/thumbs/lab-qc-quality.webp) Enlarge](https://medlabbasics.com/sheets/thumbs/lab-qc-quality-lg.webp)

**Printable PDF** Letter landscape, ready to print or keep with your notes

[View PDF](https://medlabbasics.com/sheets/lab-qc-quality.pdf) [Download](https://medlabbasics.com/sheets/lab-qc-quality.pdf)

## QC statistics

| Measure | Formula | What it tells you |
| --- | --- | --- |
| Mean | Sum of the control results ÷ number of results | Center of the control's distribution; the laboratory's own target |
| Standard deviation (SD) | √\[Σ(x − mean)² ÷ (n − 1)\] | Spread of results: imprecision (random error), in the analyte's units |
| Coefficient of variation (CV) | SD ÷ mean × 100 | Imprecision as a percent, so methods and levels can be compared |
| Bias | (Laboratory mean − target or peer mean) ÷ target × 100 | Inaccuracy (systematic error) |
| z-score (SD index) | (Result − mean) ÷ SD | How many SDs a result lies from the mean; what the Westgard rules read |
| Total error (TE) | \|Bias\| + 1.96 × CV | Worst likely error of a single result: bias plus about 2 SD of imprecision |
| Sigma metric | (TEa − \|bias\|) ÷ CV, all in % | 6 or more is world class; below 3 is poor and needs strict QC |

| Term | Meaning | Error it reflects |
| --- | --- | --- |
| Accuracy | Closeness of results to the true value | Systematic error (bias); can be found and corrected, for example by calibration |
| Precision | Closeness of repeated results to each other | Random error; can be reduced but not removed |
| Shift | Control results jump to a new level and stay there | Systematic: new reagent or control lot, calibration |
| Trend (drift) | Results move steadily up or down over several runs | Systematic: progressive change such as reagent deterioration |

**Exam tip** Worked: control results 98, 100, 102, 99 and 101 mg/dL. Mean = 500 ÷ 5 = 100. Deviations −2, 0, +2, −1, +1; squares sum to 10; 10 ÷ 4 = 2.5; SD = √2.5 = 1.58 mg/dL. CV = 1.58 ÷ 100 × 100 = 1.6%. A result of 104 gives z = (104 − 100) ÷ 1.58 = +2.5.

## Westgard multirules

*Applied as in the Westgard Rules Check calculator: every rejection rule checked on every run, 1-2s reported as a warning only when nothing else fires. The calculator checks 4-1s within one level and does not check 10x; Westgard allows 4-1s and 10x to count across levels.*

| Rule | Violated when | Error type | Action |
| --- | --- | --- | --- |
| 1-2s | One result beyond ±2 SD | Warning only | Inspect with the other rules; do not reject on 1-2s alone |
| 1-3s | One result beyond ±3 SD | Random (or the start of a large systematic error) | Reject the run |
| 2-2s | Two results beyond the same 2 SD limit (both above +2 or both below −2): the same level in two consecutive runs, or both levels in one run | Systematic | Reject the run |
| R-4s | Within one run, one level beyond +2 SD and the other beyond −2 SD. Never across runs | Random | Reject the run |
| 4-1s | Four consecutive results beyond the same 1 SD limit (all above +1 or all below −1) | Systematic | Reject the run |
| 10x | Ten consecutive results on the same side of the mean | Systematic | Reject the run |

- Use the laboratory's own mean and SD for each control lot, not the manufacturer's printed range.
- 1-3s and R-4s point to random error: for example clots or debris in the pipette, inadequate cleaning, electronic noise or an operator error.
- 2-2s, 4-1s and 10x point to systematic error: calibration, a new reagent or control lot, deteriorating reagent, wrong fluid levels or temperature.
- After a rejection: stop reporting, test a fresh vial of control (rules out a spoiled vial), find and fix the cause, rerun controls, then review every patient result since the last acceptable run.

**Safety** Repeating the same control until it falls inside the limits is not troubleshooting. CLIA requires the laboratory to evaluate all patient results from the failed run and since the last acceptable run (42 CFR 493.1282).

Try one

Quality Control, Validation and Regulation practice question 1 of 3

An outpatient clinic laboratory performs a CLIA-waived visual urine hCG cassette test. Each cassette has an internal control line. A new shipment arrives with a new lot number, and a medical assistant who has completed training starts on the testing bench the same day. Which practice BEST follows good laboratory practice for waived testing before patient results are reported?

A. Run positive and negative external controls on the new lot, performed by the newly trained operator
B. Test two patient urines on both the old and new lots and compare the lines
C. Rely on each cassette's internal control line, because it checks the device with each test
D. Run external controls at the next monthly QC date already on the clinic schedule

**Show the answer and rationale**

**Answer: A**

- **A.** Correct. CDC recommends external controls at least with each new shipment, each new lot number and each new operator before patient testing, in addition to the internal controls the insert requires.
- **B.** Tempting as a lot-to-lot check, but patient urines of unknown hCG status cannot show that the new lot detects a positive. External controls of known reactivity are the recommended check.
- **C.** Tempting because the internal control is built in, but it does not check the antibody reagents against known positive and negative samples. External controls are recommended with each new lot and shipment and by each new operator.
- **D.** Tempting because a schedule exists, but waiting a month would release results on an unchecked lot and from an operator whose performance has not been checked with known material.

The only CLIA requirement for a certificate-of-waiver site is to follow the current manufacturer's instructions, but surveys found many waived sites skipping QC, using expired kits and keeping poor records. CDC's good laboratory practices recommend running internal controls as often as the insert specifies and external controls at least with each new shipment, each new lot number and each new operator, before patient results are reported. Results, lot numbers, expiration dates and corrective actions are documented, and patient testing stops when controls fail until the problem is resolved.

Sources: [CDC MMWR: Good Laboratory Practices for Waived Testing Sites (2005)](https://www.cdc.gov/mmwr/preview/mmwrhtml/rr5413a1.htm)

[Practice 41 more questions free](https://medlabbasics.com/app/?from=/sheets/lab-qc-quality.html#/topic/lab-qc-quality) Rationales for every answer. No download needed.

## Validation, calibration and reference intervals

| Situation | What CLIA requires before patient results | Notes |
| --- | --- | --- |
| Unmodified FDA-cleared or approved nonwaived test (verification) | Accuracy, precision, reportable range; verify the manufacturer's reference interval fits the patient population | Each instrument, even the same make and model |
| Modified FDA test, laboratory-developed test, or no manufacturer specifications (establishment) | Accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range, reference intervals | 42 CFR 493.1253(b)(2) |
| Calibration verification | Follow the manufacturer's instructions, or the laboratory's own criteria using at least 3 levels (minimal or zero, mid-point, near the upper limit of the reportable range); at least every 6 months and after a trigger | Triggers: complete reagent change (unless the lot change is shown not to matter), major maintenance or critical part replaced, QC shift or trend that other fixes do not correct, or the laboratory's own schedule |
| Reference interval establishment | At least 120 reference subjects per group (CLSI C28-A3) | Central 95% of healthy subjects |
| Reference interval transference | Test 20 local healthy subjects: accept if no more than 2 fall outside | If 3 or more fall outside, test 20 more; if 3 or more fail again, re-examine the method and population |

**Exam tip** Method comparison: run at least 40 patient specimens across the range on both methods. With regression y = a + bx, systematic error at decision level Xc is (a + b·Xc) − Xc. The y-intercept points to constant error and a slope away from 1.00 to proportional error. r tells you whether the range was wide enough (0.99 or more), not whether the method is acceptable.

## Proficiency testing and competency

| Topic | Rule (42 CFR 493) |
| --- | --- |
| Handling PT samples | Test with the routine workload, by the staff who usually do the test, by the routine method, the same number of times as patient specimens |
| Forbidden | Discussing PT results with another laboratory before the deadline; sending PT samples (or any part) to another laboratory for a test you perform. Intentional referral can cost the certificate for at least a year |
| Passing score | At least 80% for most analytes and events; 100% for ABO group and D typing, unexpected antibody detection and compatibility testing |
| Unsatisfactory | Below the passing score in one event; not participating or returning late scores 0 |
| Unsuccessful | Unsatisfactory for the same analyte in 2 consecutive events or 2 of 3 consecutive events: triggers CMS action |
| No PT available (alternative assessment) | Verify accuracy at least twice a year (split samples, blind specimens, other means) |
| Competency, nonwaived testing | Six elements for every test: (1) direct observation of routine testing, (2) monitoring recording and reporting of results, (3) review of worksheets, QC, PT and maintenance records, (4) direct observation of instrument maintenance and function checks, (5) testing previously analyzed, blind or PT samples, (6) problem-solving |
| Competency timing | Twice in the first year of testing, then once a year; retrain and assess before reporting on a new method or instrument |

**Exam tip** Training, PT results and annual performance appraisals do not, on their own, meet the competency requirement. Waived-only testing needs trained staff who follow the package insert, but CLIA does not require competency assessment for it.

## CLIA, accreditation and POCT

| Category | How a test gets there | Main requirements |
| --- | --- | --- |
| Waived | Cleared for home use, or so simple and accurate that error is negligible, or no reasonable risk of harm if done wrong (examples: dipstick urinalysis, urine hCG by visual color, fecal occult blood, home-use glucose meters, spun microhematocrit) | Certificate of waiver; follow the manufacturer's instructions exactly |
| Moderate complexity (includes PPM) | FDA scores 7 criteria from 1 to 3; total 12 or less | QC, PT, verification, competency; testing personnel need at least a high school diploma with documented training |
| High complexity | Total score above 12; any test not yet categorized | All of the above, stricter personnel rules (for example an associate degree in laboratory science or equivalent) |

- Agencies: FDA categorizes tests; CMS runs CLIA and certifies laboratories; CDC provides technical support, studies and the advisory committee.
- Certificates: waiver, PPM, registration, compliance (state inspection) and accreditation (deemed status through a CMS-approved organization: CAP, The Joint Commission, COLA, AABB, A2LA, ACHC, ASHI).
- Default QC for quantitative nonwaived tests is two levels each day of testing; qualitative tests need a positive and a negative control. IQCP is the voluntary risk-based alternative.
- IQCP = risk assessment (specimen, test system, reagent, environment, testing personnel) + quality control plan + quality assessment. Every specialty except pathology is eligible.
- POCT oversight: a POCT coordinator and director, operator records with certification dates, devices that lock out when QC is overdue or fails, connectivity to the LIS, and the same CLIA rules as any test of that complexity.

## Errors, reporting and turnaround time

| Phase | Examples | Controls |
| --- | --- | --- |
| Preanalytical (most errors) | Wrong patient or mislabeled tube, wrong tube or additive, hemolysis, prolonged tourniquet or fist clenching (false high potassium), delayed transport, IV contamination | Positive patient identification, specimen rejection criteria, written collection procedures, delta checks |
| Analytical | Failed QC, expired reagent, calibration error, result above the reportable range reported without dilution, interference | QC and Westgard rules, calibration verification, maintenance, dilution protocols |
| Postanalytical | Transcription error, result sent to the wrong chart, critical value not called, delayed report | Interfaces and autoverification, critical value policy, corrected reports |

- Critical (panic) value: the laboratory must alert the person who ordered the test, or who uses the result, immediately. Laboratory policy usually adds read-back and a record of who was told and when.
- Corrected report: promptly tell the ordering provider of the error, send a corrected report, and keep both the original and the corrected report.
- Delta check: compares a result with the same patient's previous result; a change too large for biology suggests misidentification or another error.
- Autoverification: rules in the LIS or middleware release results without manual review when they pass QC, flags, limit and delta checks (CLSI AUTO-10A and AUTO-15).
- Turnaround time is right-skewed: report the median and the time by which 90% of tests are done, not the mean alone. Most delays are pre- or postanalytical.
- Records: QC, PT, patient test records and reports at least 2 years; immunohematology and transfusion records follow the FDA blood rules in 21 CFR 606.160.

## Management and education (MLS only)

| Bloom's level (revised) | Learner can | Objective verbs |
| --- | --- | --- |
| 1 Remember | Recall facts | List, name, define |
| 2 Understand | Explain, classify, compare | Explain, describe, compare |
| 3 Apply | Use knowledge in a new situation | Calculate, perform, use |
| 4 Analyze | Break information into parts | Differentiate, troubleshoot, interpret |
| 5 Evaluate | Judge value or quality | Judge, appraise, decide |
| 6 Create | Produce something new | Design, develop, write |

- Write objectives in behavioral terms: what the learner will be able to do, with a measurable verb ('calculate the CV', not 'understand QC').
- Domains: cognitive (knowledge), psychomotor (bench skills, such as making a smear), affective (attitudes, such as professionalism).
- Staffing: enough staff to cover the workload, a current competency-based job description for each position, orientation, training, competency assessment, continuing education and an annual performance appraisal.
- Budget: part of every plan, with staff, training, equipment, reagents, supplies and QC material; supply quantification each year feeds the budget.
- Quality management: the 12 quality system essentials and the plan-do-check-act cycle (see In depth).

[Practice Quality Control, Validation and Regulation questions free in your browser](https://medlabbasics.com/app/?from=/sheets/lab-qc-quality.html#/topic/lab-qc-quality) 42 exam-style questions with rationales. No account or download needed.

## In depth

The summary above is what the PDF prints. These sections go further.

### A worked Levey-Jennings chart

A Levey-Jennings chart plots each control result against the day or run. A center line marks the mean, with lines at ±1, ±2 and ±3 SD. In control, points scatter evenly above and below the mean and about 95% fall within ±2 SD, which is why roughly 1 good run in 20 gives a 1-2s warning on a single level. A shift shows as points jumping to one side and staying there. A trend shows as points climbing or falling steadily. Random error shows as one point far out, or wider scatter. Below is a glucose level 1 control with an established mean of 100 mg/dL (5.6 mmol/L) and an SD of 2.0 mg/dL (0.11 mmol/L), so CV = 2.0%. Each call is the one the Westgard Rules Check calculator makes with this level alone.

*z = (result − 100) ÷ 2.0. No single result on days 10 to 13 is even a 1-2s warning: only the multirule catches this small shift.*

| Day | Result (mg/dL) | z | Call |
| --- | --- | --- | --- |
| 1 | 101.0 | +0.5 | Accept |
| 2 | 98.0 | −1.0 | Accept |
| 3 | 104.6 | +2.3 | 1-2s warning; no rejection rule fires (day 2 was not beyond +2): accept |
| 4 | 100.8 | +0.4 | Accept |
| 5 | 106.4 | +3.2 | Reject: 1-3s (random error). Fresh vial, then look for a probe clot or bubble |
| 6 | 99.2 | −0.4 | Accept after the fix |
| 7 | 104.4 | +2.2 | 1-2s warning only: accept |
| 8 | 104.8 | +2.4 | Reject: 2-2s across runs (days 7 and 8 both above +2 SD; systematic). Recalibrate |
| 9 | 99.6 | −0.2 | Accept |
| 10 | 102.4 | +1.2 | Accept |
| 11 | 102.6 | +1.3 | Accept (only days 10 and 11 above +1 SD) |
| 12 | 102.2 | +1.1 | Accept (days 10-12 above +1 SD; day 9 was below the mean) |
| 13 | 102.8 | +1.4 | Reject: 4-1s (days 10-13 all above +1 SD; systematic). Check reagent lot and calibration |

| Run with two levels | Level 1 z | Level 2 z | Call |
| --- | --- | --- | --- |
| A | +2.3 | −2.2 | Reject: R-4s (one above +2, the other below −2 in the same run; random error) |
| B | +2.1 | +2.4 | Reject: 2-2s across levels (systematic error) |
| C | +2.4 | +0.6 | 1-2s warning only; accept if earlier runs break no rule |
| D | +3.1 | −0.5 | Reject: 1-3s |

**Exam tip** Classic vs modern multirule: in the original scheme a 1-2s warning starts the check and the other rules are looked at only then. In current use (and in this app) every rejection rule is checked on every run, so a 4-1s pattern like days 10-13 is rejected even though no point passed 2 SD. If a question states its laboratory's approach, follow the question.

Why 1-2s is not a rejection rule: with one control level it rejects about 4.6% of good runs, 8.9% with two levels and 13.0% with three. A laboratory running two levels on 20 analytes will see a false 1-2s somewhere in almost every QC event. Some sources define R-4s as a range of more than 4 SD between the two levels in a run; Westgard and the calculator require one level beyond +2 SD and the other beyond −2 SD, which is the form used here.

### Setting up QC and troubleshooting

- New control lot: run it alongside the current lot. About 10 results on different days give a first mean; update it as more data come in (a floating mean) and set the SD from longer-term data that include day-to-day sources of variation.
- Assayed controls: the manufacturer's value may be used only for your method and instrument and only after you verify it. Unassayed controls: the laboratory sets the mean and SD by running them alongside controls of known value.
- Run controls the same way as patient specimens and rotate QC among all operators. If calibrator material is used as a control, it must come from a different lot from the one used to calibrate.
- Run controls before resuming patient testing after a complete reagent change, major maintenance or replacement of a critical part.
- A new reagent or control lot may shift the mean while the SD stays about the same; CLIA requires a mean and SD for each lot of control material.
- Transient problems (inadequate cleaning, electronic noise, a clot or debris in the pipette) affect a few samples and may be gone by the next QC event. Persistent problems (calibration, reagent deterioration, wrong fluid levels, wrong temperature) stay until corrected and change either the bias or the imprecision.

| Pattern on the chart | Likely error | First checks |
| --- | --- | --- |
| One point beyond 3 SD | Random | Fresh control vial, then clots or debris in the pipette, cleaning, the run's events |
| Two levels on opposite sides beyond 2 SD | Random (imprecision) | Pipetting and sampling, cleaning, operator technique |
| Points beyond 2 SD on the same side, or many points on one side of the mean | Systematic (shift) | Calibration, new reagent or control lot |
| Points creeping steadily up or down | Systematic (trend or drift) | Reagent deterioration, temperature, fluid levels |

### Total allowable error and sigma

Total allowable error (TEa) is the largest error a result can carry without changing a medical decision. There is no single accepted set of TEa values; the laboratory director chooses one with clinicians, in order of preference from outcome studies, biological variation, then the state of the art (what current methods and peer groups achieve). CLIA's PT acceptance limits are also often used as TEa (a convention, not a CLIA rule; this page uses them for its worked example): glucose ±8% or ±6 mg/dL (whichever is greater), potassium ±0.3 mmol/L, sodium ±4 mmol/L, creatinine ±10% or ±0.2 mg/dL, total calcium ±1.0 mg/dL, total cholesterol ±10%.

*Grades from Panda 2023. The 'Meaning for QC' column is a general reading of the published advice that higher sigma allows simpler QC; no source sets fixed QC rules for each band.*

| Sigma | Grade | Meaning for QC |
| --- | --- | --- |
| 6 or more | World class | Low error rate; a less stringent QC strategy can be used |
| 5 to below 6 | Excellent | Few errors |
| 4 to below 5 | Good | Acceptable |
| 3 to below 4 | Marginal | More errors; needs stricter rules or more frequent QC |
| Below 3 | Poor or unacceptable | High error rate; the method itself needs improving |

**Exam tip** Worked: glucose at 100 mg/dL, TEa taken from the CLIA PT limit by choice (8% here, because 8% of 100 is 8 mg/dL, more than 6), bias 1.0%, CV 2.0%. Sigma = (8 − 1) ÷ 2 = 3.5: marginal. Total error = 1.0 + 1.96 × 2.0 = 4.9%, inside the 8% allowance, but with little margin.

### Verification in practice

| Characteristic | How it is commonly shown |
| --- | --- |
| Accuracy | Test materials of known value (calibrators, assayed controls, PT samples, previously tested patient specimens) or compare split patient samples with a method already shown to be accurate; results within the manufacturer's limits |
| Precision | Repeat the same samples within a run and across days, with several operators (one may be enough on fully automated systems) |
| Reportable range | Samples of known value at the lowest and highest levels the manufacturer claims (linearity materials or dilutions); report only results inside the verified range; the director decides how results above or below it are handled (for example a validated dilution) |
| Reference interval | Show the manufacturer's or a published interval fits the local population (the 20-subject transference check) |
| Analytical sensitivity | Lowest concentration the method can reliably detect (establishment only) |
| Analytical specificity | Measures only the analyte: interference (hemolysis, icterus, lipemia, drugs) and cross-reactivity (establishment only) |

- The same samples can serve accuracy, precision and reportable range at once. The technical consultant (moderate) or technical supervisor (high complexity) is responsible; the laboratory director approves the study before patient testing.
- Tests in use before April 24, 2003 do not need verification under 493.1253.
- A laboratory that runs one test on more than one instrument or site must compare their results twice a year.
- Using a waived device off label (for example an arterial sample on a meter cleared only for capillary and venous blood) modifies the test: CMS treats a modified waived or moderate complexity test as uncategorized, so it becomes high complexity, and the laboratory must establish its own performance specifications and meet every high-complexity requirement, personnel included.

### Regulation details

- PT: enroll in a CMS-approved program for each regulated analyte; stay with a program at least a year; most programs send 5 samples per event, 3 events a year. The target value is usually the peer-group mean after removing results more than 3 SD out; the acceptance limit is a symmetrical ± around it.
- After an unsatisfactory PT result the laboratory must take remedial action (training, technical assistance) and document it; a first unsuccessful performance can be met with required training or technical assistance instead of sanctions, unless patients are in immediate jeopardy.
- PT only covers the primary method used for patients during the event; keep PT records, including the signed attestation, at least 2 years.
- Competency is assessed by the technical consultant (moderate complexity) or technical supervisor (high complexity; may be delegated in writing to a qualified general supervisor). Supervisors and consultants also need competency assessed against their own regulatory duties.
- Accreditation organizations must hold the laboratory to standards equal to or stricter than CLIA. Accredited laboratories still need a CLIA certificate (of accreditation), pay CLIA fees and do PT.
- Quality assessment: CLIA requires ongoing review of preanalytic, analytic and postanalytic systems, checking whether corrective actions worked, revising procedures and discussing reviews with staff, all documented.
- Specimen rules: written procedures for patient preparation, collection, labeling, storage, transport, processing, acceptance and rejection, and referral only to CLIA-certified laboratories; record the date and time each specimen arrives.

### Quality management and education

*The 12 quality system essentials (CLSI framework, as set out in the WHO Laboratory Quality Management System handbook).*

| Quality system essential | Covers |
| --- | --- |
| Organization | Leadership, structure, quality manual, planning and budget |
| Personnel | Qualifications, job descriptions, orientation, training, competency, continuing education |
| Equipment | Selection, installation, calibration, maintenance, service records |
| Purchasing and inventory | Suppliers, quantification of needs, stock control |
| Process control | Sample management, QC, method verification |
| Information management | Reports, the LIS, confidentiality |
| Documents and records | Controlled procedures and retained records |
| Occurrence management | Finding, recording and correcting errors and near misses |
| Assessment | Internal audit, external quality assessment (PT), inspections |
| Process improvement | Quality indicators and the plan-do-check-act cycle |
| Customer service | Satisfaction surveys, complaints |
| Facilities and safety | Space, environment, safety programs |

- Plan-do-check-act (Deming): plan (find the problem and its causes, plan the improvement), do (put the plan into action), check (review whether it worked and revise the plan), act (take the corrective action, recheck, and start the cycle again).
- Quality assessment draws on QC reviews, PT performance reviews, chart reviews, specimen rejection logs, turnaround time reports and complaint reports.
- Turnaround time can be measured order to report, collection to report or receipt to report (in-lab); say which. In-lab TAT underestimates what the clinician experiences.
- Teaching: higher Bloom's levels (apply, analyze, evaluate) need case or problem questions, not recall; an objective's verb tells you how to test it. Bench skills (psychomotor) are assessed by observation, as in competency element 1.

**Exam tip** Related pages: the Westgard Rules Check and QC Statistics calculators, [Laboratory Mathematics](https://medlabbasics.com/sheets/lab-math.html) (mean, SD, CV, sensitivity and specificity), [Chemistry Methods and Instrumentation](https://medlabbasics.com/sheets/lab-chemistry-methods.html) (interferences, autoverification) and Laboratory Safety.

[Practice free in your browser](https://medlabbasics.com/app/?from=/sheets/lab-qc-quality.html) 600+ exam-style questions with rationales. Free today, no account, no ads. Pro is coming in January 2027. Also free [on Google Play](https://play.google.com/store/apps/details?id=com.alex.medlabbasics).

## Related study sheets

- [**Laboratory Safety and Biosafety** Bloodborne pathogens and exposures, biosafety levels and safety cabinets, chemical labels and SDS, fire, electrical and radiation safety, waste, shipping specimens, and the sentinel laboratory](https://medlabbasics.com/sheets/lab-safety.html)
- [**Laboratory Mathematics** Units, solutions, molarity and normality, dilutions and titers, Beer's law, centrifuge force, QC statistics, sensitivity and predictive value, cell counts and clearance, each worked step by step](https://medlabbasics.com/sheets/lab-math.html)

## Calculators for this topic

- [Westgard Rules Check](https://medlabbasics.com/calculators/westgard.html)

## Sources

1. Westgard JO. 'Westgard Rules' and multirules. Westgard QC (westgard.com). [Link](https://www.westgard.com/westgard-rules.html)
2. Westgard JO. The comparison of methods experiment. Westgard QC (westgard.com). [Link](https://www.westgard.com/lesson23.htm)
3. Westgard JO. QC: the multirule interpretation. Westgard QC (westgard.com). [Link](https://www.westgard.com/lessons/basic-qc-practices/lesson18.html)
4. Centers for Medicare & Medicaid Services. QSO-25-10-CLIA (revised June 23, 2025): CLIA 2024 final rule guidance and State Operations Manual Appendix C updates. [Link](https://www.cms.gov/files/document/qso-25-10-clia-revised-2025-06-23.pdf)
5. Electronic Code of Federal Regulations. 42 CFR Part 493, Laboratory Requirements (sections 493.2, 493.5, 493.15, 493.17, 493.551, 493.801-493.959, 493.1105, 493.1236-493.1291, 493.1413, 493.1423, 493.1451, 493.1489). [Link](https://www.ecfr.gov/current/title-42/chapter-IV/subchapter-G/part-493)
6. Centers for Medicare & Medicaid Services. CLIA brochure: Assessing Personnel Competency. [Link](https://www.cms.gov/regulations-and-guidance/legislation/clia/downloads/clia_compbrochure_508.pdf)
7. Centers for Medicare & Medicaid Services. CLIA brochure: Verification of Performance Specifications. [Link](https://www.cms.gov/regulations-and-guidance/legislation/clia/downloads/6064bk.pdf)
8. Centers for Medicare & Medicaid Services. Specialty/subspecialty information for approved accreditation organizations under CLIA. [Link](https://www.cms.gov/regulations-and-guidance/legislation/clia/downloads/aolist.pdf)
9. Centers for Disease Control and Prevention, Centers for Medicare & Medicaid Services. Developing an IQCP: A Step-by-Step Guide. [Link](https://www.cdc.gov/lab-quality/docs/developing-iqcp.pdf)
10. Gruber L, Hausch A, Mueller T. Internal quality controls in the medical laboratory: a narrative review of the basic principles of an appropriate quality control plan. Diagnostics (Basel). 2024;14(19):2223. doi:10.3390/diagnostics14192223 [Link](https://europepmc.org/article/PMC/PMC11475633)
11. Cristelli L, Occhipinti F, Tumiatti D, Antonia L, Jani E, Daves M. Implementation of new Westgard rules suggested by the Westgard Advisor software for five immunological parameters. Biochem Med (Zagreb). 2025;35(1):010701. doi:10.11613/BM.2025.010701 [Link](https://europepmc.org/article/PMC/PMC11654237)
12. Giannoli JM, Albarede S, Avellan T, et al. Recommendations for the application and follow-up of quality controls in medical laboratories. Biochem Med (Zagreb). 2021;31(2):020501. doi:10.11613/BM.2021.020501 [Link](https://europepmc.org/article/PMC/PMC8047787)
13. Hedayati M, Tohidi M, Azimzadeh I, et al. A 25-year experience in quality assurance: a methodological review of laboratory quality control and quality assurance in the Tehran Lipid and Glucose Study (TLGS). Int J Endocrinol Metab. 2027;25(1):e168018. doi:10.5812/ijem-168018 [Link](https://europepmc.org/article/PMC/PMC13589465)
14. Panda CR, Kumari S, Mangaraj M, Nayak S. The evaluation of the quality performance of biochemical analytes in clinical biochemistry laboratory using Six Sigma matrices. Cureus. 2023;15(12):e51386. doi:10.7759/cureus.51386 [Link](https://europepmc.org/article/PMC/PMC10826247)
15. Ozarda Y. Reference intervals: current status, recent developments and future considerations. Biochem Med (Zagreb). 2016;26(1):5-16. doi:10.11613/BM.2016.001 [Link](https://europepmc.org/article/PMC/PMC4783089)
16. Hawkins RC. Laboratory turnaround time. Clin Biochem Rev. 2007;28(4):179-194. [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC2282400/)
17. Adams NE. Bloom's taxonomy of cognitive learning objectives. J Med Libr Assoc. 2015;103(3):152-153. [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC4511057/)
18. Gül BÜ, Özcan O, Doğan S, Arpaci A. Designing and validating an autoverification system of biochemical test results in Hatay Mustafa Kemal University, clinical laboratory. Biochem Med (Zagreb). 2022;32(3):030704. doi:10.11613/BM.2022.030704 [Link](https://europepmc.org/article/PMC/PMC9344865)
19. Shin S, Yu S, Kim S, Yoo SJ, Cho EJ, Chung JW. Proposal for delta check limits of frequently requested hormones using real-world data. Biochem Med (Zagreb). 2025;35(1):010704. doi:10.11613/BM.2025.010704 [Link](https://europepmc.org/article/PMC/PMC11838720)
20. Romero A, Gómez-Salgado J, Romero-Arana A, et al. Preanalytical errors in emergency department samples: investigating error sources. J Med Biochem. 2020;39(4):474-480. doi:10.5937/jomb0-25263 [Link](https://europepmc.org/article/PMC/PMC7710373)
21. Fung AWS. Utilizing connectivity and data management system for effective quality management and regulatory compliance in point of care testing. Pract Lab Med. 2020;22:e00187. doi:10.1016/j.plabm.2020.e00187 [Link](https://europepmc.org/article/PMC/PMC7649638)
22. World Health Organization. Laboratory Quality Management System: Handbook. Geneva: WHO; 2011. [Link](https://www.who.int/publications/i/item/9789241548274)