Laboratory Math
Centrifuge: RPM and Relative Centrifugal Force (× g) Calculator
Converts rotor speed to g-force, or the g-force a procedure asks for to the speed to set, from the rotor radius.
Formula
RCF (× g) = 1.118 × 10⁻⁵ × r (cm) × RPM²
RPM = √(RCF ÷ (1.118 × 10⁻⁵ × r))
How it is used
- Procedures and tube makers state force in × g, not RPM, because the same speed gives more force in a larger rotor.
- Doubling the speed gives four times the force; doubling the radius gives twice the force.
- Check centrifuge speed with a tachometer and timers on the laboratory schedule, and balance tubes opposite each other.
Related study sheets
- Laboratory Mathematicsg/dL to g/L · × 10 · Hgb 13.5 g/dL = 135 g/L
- Clinical Chemistry: Analytes and Disease PatternsGlucose · mg/dL × 0.0555 = mmol/L · 126 mg/dL = 7.0 mmol/L
- Reference Ranges the Exam UsesGlucose (blood, CSF, urine) · mg/dL × 0.0555 = mmol/L · 100 mg/dL → 5.6 mmol/L; CSF 50 → 2.8
- Compatibility Testing and Blood ComponentsPlatelets (whole-blood-derived) · ≥5.5 × 10¹⁰ platelets in 40–70 mL plasma · 20–24 °C, continuous gentle agitation · 5 days (up to 7 with an FDA bacterial risk control strategy)
Related calculators
- Antigen-Negative Units to Screen
- Beer's Law: Unknown Concentration and Absorbance
- Conventional and SI Units
- Dilution Factor, Serial Dilutions and Corrected Result
- eGFR (CKD-EPI 2021, Creatinine)
- Fetal-Maternal Hemorrhage (Kleihauer-Betke) and RhIG Vials
- LDL Cholesterol (Friedewald) and Non-HDL
- Manual Cell Count (Hemocytometer)
Practice free in your browser600+ exam-style questions with rationales. Free today, no account, no ads. Pro is coming in January 2027. Also free on Google Play.
For education and exam preparation only. Follow your institution’s protocols. For education and exam preparation only. Verify results independently and follow your institution’s protocols. Not a substitute for clinical judgment.