Units, solutions, molarity and normality, dilutions and titers, Beer's law, centrifuge force, QC statistics, sensitivity and predictive value, cell counts and clearance, each worked step by step
0.85% NaCl: 0.85 g/100 mL, so 8.5 g made up to 1000 mL
% v/v
mL of solute per 100 mL of solution
500 mL of 70% ethanol from 95%: 70 × 500 ÷ 95 = 368 mL of 95%, made up to 500 mL
% w/w
g of solute per 100 g of solution
Concentrated H₂SO₄ labeled 96% (w/w): 96 g acid per 100 g of liquid
Molarity (M, mol/L)
Moles of solute per liter of solution; g = M × L × molar mass
0.9% NaCl = 9 g/L ÷ 58.44 g/mol = 0.154 mol/L
Molality (mol/kg)
Moles of solute per kg of solvent (not solution)
5.844 g NaCl (0.100 mol) in 500 g water = 0.200 mol/kg
Normality (N, Eq/L)
Equivalents per liter; N = M × equivalents per mole
1 M H₂SO₄ = 2 N; 1 M NaOH = 1 N; 1 M CaCl₂ = 2 N
Equivalent weight
Molar mass ÷ equivalents per mole
H₂SO₄ 98.08 ÷ 2 = 49.04 g/Eq; 500 mL of 0.1 N needs 0.1 × 0.5 × 49.04 = 2.45 g
Equivalents per mole: replaceable H⁺ for an acid, OH⁻ for a base, total positive charge for a salt.
Hydrate swap: g of new form = g of old form × (molar mass of new ÷ molar mass of old). 17.8 g Na₂HPO₄·2H₂O (177.99) = 17.8 × 268.07 ÷ 177.99 = 26.8 g of the heptahydrate.
Concentrated acid: M = specific gravity × 1000 × purity (as a fraction) ÷ molar mass. 96% H₂SO₄, SG 1.84: 1.84 × 1000 × 0.96 ÷ 98.08 = 18.0 M (36.0 N).
Dilutions, serial dilutions and titers
Task
Formula
Worked example
Dilution of one tube
Sample ÷ (sample + diluent); written 1:total
100 µL serum + 900 µL diluent = 1:10
Make a set volume (C1V1 = C2V2)
V1 = C2 × V2 ÷ C1; diluent = V2 − V1
50 mL of a 2 unit solution from 10 unit stock: V1 = 2 × 50 ÷ 10 = 10 mL stock + 40 mL diluent (1:5)
Serial dilution
Multiply the tube factors
Three 1:10 tubes = 1:1000. Tubes of 1:5, 1:2 and 1:10 = 1:100
Twofold series
Equal volumes each step: 1:2, 1:4, 1:8, 1:16 ...
Tube 4 of a twofold series = 1:16
Patient result
Result read × total dilution factor
1:10 dilution reads 45: report 450
Titer
Reciprocal of the highest dilution still positive
Positive to 1:32, negative at 1:64: titer 32
Significant rise
Fourfold (two tubes) between acute and convalescent sera
Titer 8 then 32
Beer's law, standard curves and centrifuge force
Calculation
Formula
Worked example
Beer's law
A = a × b × c (absorptivity × path length × concentration)
Absorbance doubles when concentration doubles (inside the linear range)
Unknown from one standard
C unknown = (A unknown ÷ A standard) × C standard
0.25 ÷ 0.50 × 100 mg/dL = 50 mg/dL
With a reagent blank
Subtract the blank from both readings first
Blank 0.050, standard 0.450, unknown 0.250: (0.200 ÷ 0.400) × C standard
%T to absorbance
A = 2 − log %T
50 %T: A = 0.301; 10 %T: A = 1.0
Standard curve
Plot A (y) against standard concentration (x); read unknowns off the line
Dilutions of a 150 g/L hemoglobin reference: 1:2 = 75, 1:3 = 50, 1:4 = 37.5 g/L
RCF (× g)
1.118 × 10⁻⁵ × r (cm) × RPM²
15 cm rotor at 3000 RPM: 1509 × g
RPM for a set RCF
√(RCF ÷ (1.118 × 10⁻⁵ × r))
1509 × g at 15 cm: 3000 RPM
Statistics and QC numbers
Statistic
Formula
Worked on 10 control values: 90, 91, 89, 84, 88, 93, 80, 90, 85, 87
Mean
Sum ÷ n
877 ÷ 10 = 87.7
Median
Middle value in rank order; with an even n, the average of the middle two
Units needed ÷ (product of the fractions negative), rounded up
2 units lacking antigens found on 70% and 30% of donors: 0.30 × 0.70 = 0.21; 2 ÷ 0.21 = 9.5, screen 10
Creatinine clearance
(U × V) ÷ P, V in mL/min; × 1.73 ÷ BSA to correct
U 100 mg/dL, 1440 mL in 1440 min, P 1.0 mg/dL: 100 mL/min
Timed urine total
Concentration × volume (in matching units)
Protein 25 mg/dL × 1500 mL (15 dL) = 375 mg/24 h
In depth
The summary above is what the PDF prints. These sections go further.
Concentration step by step
The SI name for molarity is amount (substance) concentration in mol/L. IUPAC and NIST call the words 'molarity' and 'normality' obsolete, but the exam and many procedures still use M and N, so learn both. Percent with no qualifier is ambiguous: always say w/v, v/v or w/w.
Percent w/v to molarity (0.9% NaCl): 0.9 g per 100 mL = 9 g per L. 9 g/L ÷ 58.44 g/mol = 0.154 mol/L (154 mmol/L). Reading 0.9% as 0.9 g/L gives a tenfold error.
Grams for a molar solution (250 mL of 0.20 M NaCl): 0.20 mol/L × 0.250 L × 58.44 g/mol = 2.92 g, dissolved and made up to 250 mL in a volumetric flask.
Normality (500 mL of 0.1 N H₂SO₄): H₂SO₄ gives 2 H⁺, so 0.1 N = 0.05 M. 0.05 mol/L × 0.500 L × 98.08 g/mol = 2.45 g. Using 0.1 M by mistake doubles the acid.
Equivalents of ions: mEq/L = mmol/L × charge. Sodium, potassium and chloride (charge 1) have the same number in both; calcium and magnesium (charge 2) have twice as many mEq as mmol.
Molality: moles per kilogram of solvent. 0.100 mol NaCl in 500 g of water is 0.200 mol/kg. Osmolality (mOsm/kg) uses the same per-kilogram basis, counted in particles.
Concentrated stock acid: the label gives purity (w/w) and specific gravity (density relative to water). 96% H₂SO₄, SG 1.84: each liter weighs 1840 g, of which 96% is acid = 1766 g; 1766 ÷ 98.08 = 18.0 mol/L, or 36.0 N.
Diluting that stock (500 mL of 0.5 M, which is 1.0 N): V1 = 0.5 × 500 ÷ 18.0 = 13.9 mL of concentrated acid, added slowly to about 400 mL of water, cooled, then made up to 500 mL.
Each added water of crystallization adds 18.015 g to the molar mass: 141.96 + 2 × 18.015 = 177.99. The moles of salt must stay the same, so the mass rises with the water.
Hydrate problem
Working
Answer
A procedure calls for 17.8 g Na₂HPO₄·2H₂O (177.99 g/mol); only the heptahydrate (268.07 g/mol) is on the shelf
17.8 × 268.07 ÷ 177.99
26.8 g
Same procedure with the dodecahydrate (358.14 g/mol)
17.8 × 358.14 ÷ 177.99
35.8 g
How much anhydrous Na₂HPO₄ (141.96 g/mol) is in 17.8 g of the dihydrate?
17.8 × 141.96 ÷ 177.99
14.2 g
Dilutions: worked series and traps
A mixed serial dilution. If tube 3 reads 2,300 units, the patient value is 2,300 × 100 = 230,000. Adding the factors (5 + 2 + 10) instead of multiplying them is the usual wrong answer.
Tube
Made from
Tube dilution
Total dilution
1
0.1 mL serum + 0.4 mL diluent
1:5
1:5
2
0.5 mL of tube 1 + 0.5 mL diluent
1:2
1:10
3
0.1 mL of tube 2 + 0.9 mL diluent
1:10
1:100
Reading a diluted sample: dilute until the result falls inside the reportable (measuring) range, then multiply back. A result that is still above the range after dilution needs a further dilution, not extrapolation.
Several dilutions from one tube: a 1:4 dilution of a 1:10 dilution is 1:40.
Making a set volume of a stated dilution: 1:20 in 1 mL total is 0.05 mL sample + 0.95 mL diluent (the WHO manual's CSF and leukocyte dilution).
Titers: in a twofold series, each tube is half the one before. A fourfold rise (two tubes, for example 1:8 to 1:32) between acute and convalescent sera is the usual mark of a significant change.
Where the formulas come from
Beer-Lambert: decadic absorbance A = −log T, where T is transmitted ÷ incident light. With %T, A = −log(%T ÷ 100) = 2 − log %T. The molar absorption coefficient ε = A ÷ (c × l), so A = ε × l × c: at a fixed wavelength and path length, A is proportional to c. Transmittance is not linear with concentration; absorbance is.
One-standard calculation: because A ∝ c for both tubes, A unknown ÷ A standard = C unknown ÷ C standard. This holds only inside the linear range; above it, dilute the sample. A curve that bends at the top means the method is out of its linear range there.
Relative centrifugal force: RCF = r × ω² ÷ g, with ω = 2π × RPM ÷ 60 and g = 9.806 65 m/s². With r in cm converted to m: RCF = r(cm) ÷ 100 × (2π ÷ 60)² × RPM² ÷ 9.806 65 = 1.118 × 10⁻⁵ × r × RPM². Measure r from the center of the rotor to the bottom of the tube.
Hemocytometer: each large Neubauer square is 1 mm × 1 mm and the chamber is 0.1 mm deep, so each square holds 0.1 mm³ = 0.1 µL. Cells per µL = cells counted ÷ volume counted × dilution. The WHO manual's leukocyte count: 188 cells in 4 squares (0.4 µL) at 1:20 = 188 ÷ 0.4 × 20 = 9,400/µL = 9.4 × 10⁹/L.
Temperature: °F = (9/5 × °C) + 32 and °C = (°F − 32) × 5/9. 56 °C (serum inactivation) = 132.8 °F.
Statistics in more depth
Mean, median and mode agree in a symmetric (Gaussian) distribution. In skewed data the mean is pulled toward the long tail and the median is the better summary.
SD uses n − 1 (degrees of freedom) because the mean was already estimated from the same data. In a Gaussian distribution about 68% of results fall within ±1 SD, 95% within ±2 SD and more than 99% within ±3 SD.
CV is SD relative to the mean, so it compares precision across levels and methods. CVs tend to be higher at low concentrations, so judge a CV with its mean.
Standard error (SD ÷ √n) shrinks as more runs are added; the SD does not. With n = 10 the 1.96 multiplier gives 85.3 to 90.1; a strict small-sample t value (2.26 for n = 10) widens it to 85.0 to 90.4.
A z-score is unitless and lets results from different levels share one chart. The Westgard example in the calculator: mean 100, SD 2, result 104.5 gives z = 2.25 (a 1-2s warning alone; with the previous run at +2.1 it is a 2-2s rejection).
Low prevalence lowers PPV, raises NPV and here slightly raises efficiency (because specificity is higher than sensitivity; efficiency moves toward specificity as prevalence falls), while sensitivity and specificity stay the same. That is why reactive screens in low-prevalence populations are confirmed by a more specific test.
Same assay (sensitivity 92%, specificity 95%)
Prevalence 20% (500 tested)
Prevalence 2% (10,000 tested)
With the condition
100
200
True positives (92%)
92
184
False positives (5% of those without)
20
490
PPV
92 ÷ 112 = 82%
184 ÷ 674 = 27%
NPV
380 ÷ 388 = 97.9%
9,310 ÷ 9,326 = 99.8%
Efficiency
472 ÷ 500 = 94.4%
9,494 ÷ 10,000 = 94.9%
Fastest route for prevalence problems: build a population of 10,000, fill in the diseased column from sensitivity and the non-diseased column from specificity, then read across the positive row.
Likelihood ratios: LR+ = sensitivity ÷ (1 − specificity); LR− = (1 − sensitivity) ÷ specificity. For the assay above, LR+ = 0.92 ÷ 0.05 = 18.4. They depend only on sensitivity and specificity, not on prevalence.
Counts, odds and timed urine in more depth
Absolute counts judge the differential: a lymphocyte share of 33% in a WBC of 8.0 is 2.64 × 10³/µL. Relative values in SI are fractions of 1 (0.33); absolute values have the same number in ×10³/µL and ×10⁹/L.
Reticulocytes: anemia raises the retic percentage because there are fewer mature red cells, so the corrected count multiplies by patient Hct ÷ 45. The RPI then divides by the maturation time for that hematocrit (2 days at an Hct of 25%; see the Reticulocyte Count calculator).
nRBC correction applies when the analyzer has counted nRBCs as white cells and has not corrected for them. The exam's calculation list names it for more than 10 nRBCs per 100 WBCs.
Rule of three as a check on indices: Hgb × 3 ≈ Hct (± 3) and RBC × 3 ≈ Hgb, for normocytic, normochromic cells only. In the indices example, 13.5 × 3 = 40.5, which matches the Hct.
Antigen-negative units: for independent antigens, multiply the fractions negative. One antigen on 9% of donors: 91% negative; 4 units ÷ 0.91 = 4.4, screen 5. The result is an average, so more may be needed; use the antigen frequencies of your own donor population.
Creatinine clearance with body size: U 120 mg/dL, 1500 mL in 24 h (1440 min), P 1.25 mg/dL. V = 1500 ÷ 1440 = 1.04 mL/min; 120 × 1.04 ÷ 1.25 = 100 mL/min; × 1.73 ÷ 1.73 m² = 100 mL/min/1.73 m². Urine and serum creatinine must be in the same unit so they cancel.
24-hour collection: the patient empties the bladder on waking and discards that urine, collects everything after it, and ends with the first void the next morning, which is kept. A short or incomplete collection gives a falsely low clearance.
Timed urine totals: convert the volume to the concentration's unit before multiplying (1500 mL = 15 dL for a result in mg/dL). Creatinine 120 mg/dL × 15 dL = 1,800 mg in 24 h.
ASCP Board of Certification. Medical Laboratory Scientist, MLS(ASCP) and MLS(ASCPi), Examination Content Guideline: Laboratory Operations III, Laboratory Mathematics, and Additional calculations. Revised June 9, 2026. Link
ASCP Board of Certification. Medical Laboratory Technician, MLT(ASCP) and MLT(ASCPi), Examination Content Guideline. Revised June 9, 2026. Link
Cohen ER, Cvitaš T, Frey JG, et al. Quantities, Units and Symbols in Physical Chemistry (IUPAC Green Book), 3rd ed. RSC Publishing; 2007 (2012 printing): amount concentration, molality, fractions and %, relative density, absorbance, standard acceleration of gravity, temperature equations. Link
Thompson A, Taylor BN. Guide for the Use of the International System of Units (SI). NIST Special Publication 811, 2008 ed. Chapter 4 (SI prefixes). Link
Thompson A, Taylor BN. NIST Special Publication 811, Chapter 8: Comments on some quantities and their units (amount-of-substance concentration, molality; molarity and normality obsolete). Link
World Health Organization. Manual of basic techniques for a health laboratory, 2nd ed. Geneva: WHO; 2003 (SI units and quantity names, preparing solutions, hemocytometer counts, standard curves, titres, reagent annex). Link
CDC/NIOSH. NIOSH Pocket Guide to Chemical Hazards: Sulfuric acid (specific gravity 1.84 for 96-98% acid). Link
Zady MF. Z-4: Mean, standard deviation, and coefficient of variation. Westgard QC, Basic Statistics lessons; 1999. Link
Zady MF. Z-5: Sum of squares, variance, and the standard error of the mean. Westgard QC; 1999. Link
Zady MF. Z-6: Probability and the standard normal distribution (z-scores). Westgard QC; 1999. Link
Westgard QC. Z-11: Confidence intervals. Basic Statistics lessons; 2000. Link
Centers for Disease Control and Prevention. Principles of Epidemiology in Public Health Practice, 3rd ed. Lesson 2, Section 6: Measures of central location. Link
Mandell BF. Understanding medical tests and test results. Merck Manual Professional Version. Full review April 2026. Link
Šimundić AM. Measures of diagnostic accuracy: basic definitions. EJIFCC. 2009;19(4):203-211. Link
Trevethan R. Sensitivity, specificity, and predictive values: foundations, pliabilities, and pitfalls in research and practice. Front Public Health. 2017;5:307. doi:10.3389/fpubh.2017.00307 Link
Gulati G, Uppal G, Gong J. Unreliable automated complete blood count results: causes, recognition, and resolution. Ann Lab Med. 2022;42(5):515-530 (nRBC-corrected WBC, rules of three). doi:10.3343/alm.2022.42.5.515 Link
Klionsky DJ. It's apples and oranges (or chalk and cheese, depending on where you live): RPM versus RCF. Autophagy. 2026;22(10):2521. doi:10.1080/15548627.2026.2685350 (RCF = 1.118 × 10⁻⁵ × r (cm) × RPM²). Link
Testing.com (ADLM). Creatinine clearance (24-hour collection, mL/min/1.73 m²). Link