Study sheet

Quality Control, Validation and Regulation

Levey-Jennings charts and Westgard rules, sigma, method verification, calibration, reference intervals, proficiency testing, competency, CLIA, POCT, error phases and reporting

Checked against current sources November 2026

Practice 42 Quality Control, Validation and Regulation questions free

Page 1 of the Quality Control, Validation and Regulation PDF, a printed page in columns with sections on QC statistics, Westgard multirules, Validation, calibration and reference intervals, Proficiency testing and competency, CLIA, accreditation and POCT, Errors, reporting and turnaround time and Management and education (MLS only)Enlarge

Printable PDFLetter landscape, ready to print or keep with your notes

View PDFDownload

Quality Control, Validation and Regulation: PDF preview

Page 1 of the Quality Control, Validation and Regulation PDF, a printed page in columns with sections on QC statistics, Westgard multirules, Validation, calibration and reference intervals, Proficiency testing and competency, CLIA, accreditation and POCT, Errors, reporting and turnaround time and Management and education (MLS only)

View PDFDownload

QC statistics

MeasureFormulaWhat it tells you
MeanSum of the control results ÷ number of resultsCenter of the control's distribution; the laboratory's own target
Standard deviation (SD)√[Σ(x − mean)² ÷ (n − 1)]Spread of results: imprecision (random error), in the analyte's units
Coefficient of variation (CV)SD ÷ mean × 100Imprecision as a percent, so methods and levels can be compared
Bias(Laboratory mean − target or peer mean) ÷ target × 100Inaccuracy (systematic error)
z-score (SD index)(Result − mean) ÷ SDHow many SDs a result lies from the mean; what the Westgard rules read
Total error (TE)|Bias| + 1.96 × CVWorst likely error of a single result: bias plus about 2 SD of imprecision
Sigma metric(TEa − |bias|) ÷ CV, all in %6 or more is world class; below 3 is poor and needs strict QC
TermMeaningError it reflects
AccuracyCloseness of results to the true valueSystematic error (bias); can be found and corrected, for example by calibration
PrecisionCloseness of repeated results to each otherRandom error; can be reduced but not removed
ShiftControl results jump to a new level and stay thereSystematic: new reagent or control lot, calibration
Trend (drift)Results move steadily up or down over several runsSystematic: progressive change such as reagent deterioration

Westgard multirules

Applied as in the Westgard Rules Check calculator: every rejection rule checked on every run, 1-2s reported as a warning only when nothing else fires. The calculator checks 4-1s within one level and does not check 10x; Westgard allows 4-1s and 10x to count across levels.
RuleViolated whenError typeAction
1-2sOne result beyond ±2 SDWarning onlyInspect with the other rules; do not reject on 1-2s alone
1-3sOne result beyond ±3 SDRandom (or the start of a large systematic error)Reject the run
2-2sTwo results beyond the same 2 SD limit (both above +2 or both below −2): the same level in two consecutive runs, or both levels in one runSystematicReject the run
R-4sWithin one run, one level beyond +2 SD and the other beyond −2 SD. Never across runsRandomReject the run
4-1sFour consecutive results beyond the same 1 SD limit (all above +1 or all below −1)SystematicReject the run
10xTen consecutive results on the same side of the meanSystematicReject the run

Validation, calibration and reference intervals

SituationWhat CLIA requires before patient resultsNotes
Unmodified FDA-cleared or approved nonwaived test (verification)Accuracy, precision, reportable range; verify the manufacturer's reference interval fits the patient populationEach instrument, even the same make and model
Modified FDA test, laboratory-developed test, or no manufacturer specifications (establishment)Accuracy, precision, analytical sensitivity, analytical specificity including interfering substances, reportable range, reference intervals42 CFR 493.1253(b)(2)
Calibration verificationFollow the manufacturer's instructions, or the laboratory's own criteria using at least 3 levels (minimal or zero, mid-point, near the upper limit of the reportable range); at least every 6 months and after a triggerTriggers: complete reagent change (unless the lot change is shown not to matter), major maintenance or critical part replaced, QC shift or trend that other fixes do not correct, or the laboratory's own schedule
Reference interval establishmentAt least 120 reference subjects per group (CLSI C28-A3)Central 95% of healthy subjects
Reference interval transferenceTest 20 local healthy subjects: accept if no more than 2 fall outsideIf 3 or more fall outside, test 20 more; if 3 or more fail again, re-examine the method and population

Proficiency testing and competency

TopicRule (42 CFR 493)
Handling PT samplesTest with the routine workload, by the staff who usually do the test, by the routine method, the same number of times as patient specimens
ForbiddenDiscussing PT results with another laboratory before the deadline; sending PT samples (or any part) to another laboratory for a test you perform. Intentional referral can cost the certificate for at least a year
Passing scoreAt least 80% for most analytes and events; 100% for ABO group and D typing, unexpected antibody detection and compatibility testing
UnsatisfactoryBelow the passing score in one event; not participating or returning late scores 0
UnsuccessfulUnsatisfactory for the same analyte in 2 consecutive events or 2 of 3 consecutive events: triggers CMS action
No PT available (alternative assessment)Verify accuracy at least twice a year (split samples, blind specimens, other means)
Competency, nonwaived testingSix elements for every test: (1) direct observation of routine testing, (2) monitoring recording and reporting of results, (3) review of worksheets, QC, PT and maintenance records, (4) direct observation of instrument maintenance and function checks, (5) testing previously analyzed, blind or PT samples, (6) problem-solving
Competency timingTwice in the first year of testing, then once a year; retrain and assess before reporting on a new method or instrument

CLIA, accreditation and POCT

CategoryHow a test gets thereMain requirements
WaivedCleared for home use, or so simple and accurate that error is negligible, or no reasonable risk of harm if done wrong (examples: dipstick urinalysis, urine hCG by visual color, fecal occult blood, home-use glucose meters, spun microhematocrit)Certificate of waiver; follow the manufacturer's instructions exactly
Moderate complexity (includes PPM)FDA scores 7 criteria from 1 to 3; total 12 or lessQC, PT, verification, competency; testing personnel need at least a high school diploma with documented training
High complexityTotal score above 12; any test not yet categorizedAll of the above, stricter personnel rules (for example an associate degree in laboratory science or equivalent)

Errors, reporting and turnaround time

PhaseExamplesControls
Preanalytical (most errors)Wrong patient or mislabeled tube, wrong tube or additive, hemolysis, prolonged tourniquet or fist clenching (false high potassium), delayed transport, IV contaminationPositive patient identification, specimen rejection criteria, written collection procedures, delta checks
AnalyticalFailed QC, expired reagent, calibration error, result above the reportable range reported without dilution, interferenceQC and Westgard rules, calibration verification, maintenance, dilution protocols
PostanalyticalTranscription error, result sent to the wrong chart, critical value not called, delayed reportInterfaces and autoverification, critical value policy, corrected reports

Management and education (MLS only)

Bloom's level (revised)Learner canObjective verbs
1 RememberRecall factsList, name, define
2 UnderstandExplain, classify, compareExplain, describe, compare
3 ApplyUse knowledge in a new situationCalculate, perform, use
4 AnalyzeBreak information into partsDifferentiate, troubleshoot, interpret
5 EvaluateJudge value or qualityJudge, appraise, decide
6 CreateProduce something newDesign, develop, write
Practice Quality Control, Validation and Regulation questions free in your browser42 exam-style questions with rationales. No account or download needed.

In depth

The summary above is what the PDF prints. These sections go further.

A worked Levey-Jennings chart

A Levey-Jennings chart plots each control result against the day or run. A center line marks the mean, with lines at ±1, ±2 and ±3 SD. In control, points scatter evenly above and below the mean and about 95% fall within ±2 SD, which is why roughly 1 good run in 20 gives a 1-2s warning on a single level. A shift shows as points jumping to one side and staying there. A trend shows as points climbing or falling steadily. Random error shows as one point far out, or wider scatter. Below is a glucose level 1 control with an established mean of 100 mg/dL (5.6 mmol/L) and an SD of 2.0 mg/dL (0.11 mmol/L), so CV = 2.0%. Each call is the one the Westgard Rules Check calculator makes with this level alone.

z = (result − 100) ÷ 2.0. No single result on days 10 to 13 is even a 1-2s warning: only the multirule catches this small shift.
DayResult (mg/dL)zCall
1101.0+0.5Accept
298.0−1.0Accept
3104.6+2.31-2s warning; no rejection rule fires (day 2 was not beyond +2): accept
4100.8+0.4Accept
5106.4+3.2Reject: 1-3s (random error). Fresh vial, then look for a probe clot or bubble
699.2−0.4Accept after the fix
7104.4+2.21-2s warning only: accept
8104.8+2.4Reject: 2-2s across runs (days 7 and 8 both above +2 SD; systematic). Recalibrate
999.6−0.2Accept
10102.4+1.2Accept
11102.6+1.3Accept (only days 10 and 11 above +1 SD)
12102.2+1.1Accept (days 10-12 above +1 SD; day 9 was below the mean)
13102.8+1.4Reject: 4-1s (days 10-13 all above +1 SD; systematic). Check reagent lot and calibration
Run with two levelsLevel 1 zLevel 2 zCall
A+2.3−2.2Reject: R-4s (one above +2, the other below −2 in the same run; random error)
B+2.1+2.4Reject: 2-2s across levels (systematic error)
C+2.4+0.61-2s warning only; accept if earlier runs break no rule
D+3.1−0.5Reject: 1-3s

Why 1-2s is not a rejection rule: with one control level it rejects about 4.6% of good runs, 8.9% with two levels and 13.0% with three. A laboratory running two levels on 20 analytes will see a false 1-2s somewhere in almost every QC event. Some sources define R-4s as a range of more than 4 SD between the two levels in a run; Westgard and the calculator require one level beyond +2 SD and the other beyond −2 SD, which is the form used here.

Setting up QC and troubleshooting

Pattern on the chartLikely errorFirst checks
One point beyond 3 SDRandomFresh control vial, then clots or debris in the pipette, cleaning, the run's events
Two levels on opposite sides beyond 2 SDRandom (imprecision)Pipetting and sampling, cleaning, operator technique
Points beyond 2 SD on the same side, or many points on one side of the meanSystematic (shift)Calibration, new reagent or control lot
Points creeping steadily up or downSystematic (trend or drift)Reagent deterioration, temperature, fluid levels

Total allowable error and sigma

Total allowable error (TEa) is the largest error a result can carry without changing a medical decision. There is no single accepted set of TEa values; the laboratory director chooses one with clinicians, in order of preference from outcome studies, biological variation, then the state of the art (what current methods and peer groups achieve). CLIA's PT acceptance limits are also often used as TEa (a convention, not a CLIA rule; this page uses them for its worked example): glucose ±8% or ±6 mg/dL (whichever is greater), potassium ±0.3 mmol/L, sodium ±4 mmol/L, creatinine ±10% or ±0.2 mg/dL, total calcium ±1.0 mg/dL, total cholesterol ±10%.

Grades from Panda 2023. The 'Meaning for QC' column is a general reading of the published advice that higher sigma allows simpler QC; no source sets fixed QC rules for each band.
SigmaGradeMeaning for QC
6 or moreWorld classLow error rate; a less stringent QC strategy can be used
5 to below 6ExcellentFew errors
4 to below 5GoodAcceptable
3 to below 4MarginalMore errors; needs stricter rules or more frequent QC
Below 3Poor or unacceptableHigh error rate; the method itself needs improving

Verification in practice

CharacteristicHow it is commonly shown
AccuracyTest materials of known value (calibrators, assayed controls, PT samples, previously tested patient specimens) or compare split patient samples with a method already shown to be accurate; results within the manufacturer's limits
PrecisionRepeat the same samples within a run and across days, with several operators (one may be enough on fully automated systems)
Reportable rangeSamples of known value at the lowest and highest levels the manufacturer claims (linearity materials or dilutions); report only results inside the verified range; the director decides how results above or below it are handled (for example a validated dilution)
Reference intervalShow the manufacturer's or a published interval fits the local population (the 20-subject transference check)
Analytical sensitivityLowest concentration the method can reliably detect (establishment only)
Analytical specificityMeasures only the analyte: interference (hemolysis, icterus, lipemia, drugs) and cross-reactivity (establishment only)

Regulation details

Quality management and education

The 12 quality system essentials (CLSI framework, as set out in the WHO Laboratory Quality Management System handbook).
Quality system essentialCovers
OrganizationLeadership, structure, quality manual, planning and budget
PersonnelQualifications, job descriptions, orientation, training, competency, continuing education
EquipmentSelection, installation, calibration, maintenance, service records
Purchasing and inventorySuppliers, quantification of needs, stock control
Process controlSample management, QC, method verification
Information managementReports, the LIS, confidentiality
Documents and recordsControlled procedures and retained records
Occurrence managementFinding, recording and correcting errors and near misses
AssessmentInternal audit, external quality assessment (PT), inspections
Process improvementQuality indicators and the plan-do-check-act cycle
Customer serviceSatisfaction surveys, complaints
Facilities and safetySpace, environment, safety programs
Practice free in your browser600+ exam-style questions with rationales. Free today, no account, no ads. Pro is coming in January 2027. Also free on Google Play.

Related study sheets

Calculators for this topic

Sources

  1. Westgard JO. 'Westgard Rules' and multirules. Westgard QC (westgard.com). Link
  2. Westgard JO. The comparison of methods experiment. Westgard QC (westgard.com). Link
  3. Westgard JO. QC: the multirule interpretation. Westgard QC (westgard.com). Link
  4. Centers for Medicare & Medicaid Services. QSO-25-10-CLIA (revised June 23, 2025): CLIA 2024 final rule guidance and State Operations Manual Appendix C updates. Link
  5. Electronic Code of Federal Regulations. 42 CFR Part 493, Laboratory Requirements (sections 493.2, 493.5, 493.15, 493.17, 493.551, 493.801-493.959, 493.1105, 493.1236-493.1291, 493.1413, 493.1423, 493.1451, 493.1489). Link
  6. Centers for Medicare & Medicaid Services. CLIA brochure: Assessing Personnel Competency. Link
  7. Centers for Medicare & Medicaid Services. CLIA brochure: Verification of Performance Specifications. Link
  8. Centers for Medicare & Medicaid Services. Specialty/subspecialty information for approved accreditation organizations under CLIA. Link
  9. Centers for Disease Control and Prevention, Centers for Medicare & Medicaid Services. Developing an IQCP: A Step-by-Step Guide. Link
  10. Gruber L, Hausch A, Mueller T. Internal quality controls in the medical laboratory: a narrative review of the basic principles of an appropriate quality control plan. Diagnostics (Basel). 2024;14(19):2223. doi:10.3390/diagnostics14192223 Link
  11. Cristelli L, Occhipinti F, Tumiatti D, Antonia L, Jani E, Daves M. Implementation of new Westgard rules suggested by the Westgard Advisor software for five immunological parameters. Biochem Med (Zagreb). 2025;35(1):010701. doi:10.11613/BM.2025.010701 Link
  12. Giannoli JM, Albarede S, Avellan T, et al. Recommendations for the application and follow-up of quality controls in medical laboratories. Biochem Med (Zagreb). 2021;31(2):020501. doi:10.11613/BM.2021.020501 Link
  13. Hedayati M, Tohidi M, Azimzadeh I, et al. A 25-year experience in quality assurance: a methodological review of laboratory quality control and quality assurance in the Tehran Lipid and Glucose Study (TLGS). Int J Endocrinol Metab. 2027;25(1):e168018. doi:10.5812/ijem-168018 Link
  14. Panda CR, Kumari S, Mangaraj M, Nayak S. The evaluation of the quality performance of biochemical analytes in clinical biochemistry laboratory using Six Sigma matrices. Cureus. 2023;15(12):e51386. doi:10.7759/cureus.51386 Link
  15. Ozarda Y. Reference intervals: current status, recent developments and future considerations. Biochem Med (Zagreb). 2016;26(1):5-16. doi:10.11613/BM.2016.001 Link
  16. Hawkins RC. Laboratory turnaround time. Clin Biochem Rev. 2007;28(4):179-194. Link
  17. Adams NE. Bloom's taxonomy of cognitive learning objectives. J Med Libr Assoc. 2015;103(3):152-153. Link
  18. Gül BÜ, Özcan O, Doğan S, Arpaci A. Designing and validating an autoverification system of biochemical test results in Hatay Mustafa Kemal University, clinical laboratory. Biochem Med (Zagreb). 2022;32(3):030704. doi:10.11613/BM.2022.030704 Link
  19. Shin S, Yu S, Kim S, Yoo SJ, Cho EJ, Chung JW. Proposal for delta check limits of frequently requested hormones using real-world data. Biochem Med (Zagreb). 2025;35(1):010704. doi:10.11613/BM.2025.010704 Link
  20. Romero A, Gómez-Salgado J, Romero-Arana A, et al. Preanalytical errors in emergency department samples: investigating error sources. J Med Biochem. 2020;39(4):474-480. doi:10.5937/jomb0-25263 Link
  21. Fung AWS. Utilizing connectivity and data management system for effective quality management and regulatory compliance in point of care testing. Pract Lab Med. 2020;22:e00187. doi:10.1016/j.plabm.2020.e00187 Link
  22. World Health Organization. Laboratory Quality Management System: Handbook. Geneva: WHO; 2011. Link