Center of the control's distribution; the laboratory's own target
Standard deviation (SD)
√[Σ(x − mean)² ÷ (n − 1)]
Spread of results: imprecision (random error), in the analyte's units
Coefficient of variation (CV)
SD ÷ mean × 100
Imprecision as a percent, so methods and levels can be compared
Bias
(Laboratory mean − target or peer mean) ÷ target × 100
Inaccuracy (systematic error)
z-score (SD index)
(Result − mean) ÷ SD
How many SDs a result lies from the mean; what the Westgard rules read
Total error (TE)
|Bias| + 1.96 × CV
Worst likely error of a single result: bias plus about 2 SD of imprecision
Sigma metric
(TEa − |bias|) ÷ CV, all in %
6 or more is world class; below 3 is poor and needs strict QC
Term
Meaning
Error it reflects
Accuracy
Closeness of results to the true value
Systematic error (bias); can be found and corrected, for example by calibration
Precision
Closeness of repeated results to each other
Random error; can be reduced but not removed
Shift
Control results jump to a new level and stay there
Systematic: new reagent or control lot, calibration
Trend (drift)
Results move steadily up or down over several runs
Systematic: progressive change such as reagent deterioration
Westgard multirules
Applied as in the Westgard Rules Check calculator: every rejection rule checked on every run, 1-2s reported as a warning only when nothing else fires. The calculator checks 4-1s within one level and does not check 10x; Westgard allows 4-1s and 10x to count across levels.
Rule
Violated when
Error type
Action
1-2s
One result beyond ±2 SD
Warning only
Inspect with the other rules; do not reject on 1-2s alone
1-3s
One result beyond ±3 SD
Random (or the start of a large systematic error)
Reject the run
2-2s
Two results beyond the same 2 SD limit (both above +2 or both below −2): the same level in two consecutive runs, or both levels in one run
Systematic
Reject the run
R-4s
Within one run, one level beyond +2 SD and the other beyond −2 SD. Never across runs
Random
Reject the run
4-1s
Four consecutive results beyond the same 1 SD limit (all above +1 or all below −1)
Systematic
Reject the run
10x
Ten consecutive results on the same side of the mean
Systematic
Reject the run
Use the laboratory's own mean and SD for each control lot, not the manufacturer's printed range.
1-3s and R-4s point to random error: for example clots or debris in the pipette, inadequate cleaning, electronic noise or an operator error.
2-2s, 4-1s and 10x point to systematic error: calibration, a new reagent or control lot, deteriorating reagent, wrong fluid levels or temperature.
After a rejection: stop reporting, test a fresh vial of control (rules out a spoiled vial), find and fix the cause, rerun controls, then review every patient result since the last acceptable run.
Validation, calibration and reference intervals
Situation
What CLIA requires before patient results
Notes
Unmodified FDA-cleared or approved nonwaived test (verification)
Accuracy, precision, reportable range; verify the manufacturer's reference interval fits the patient population
Each instrument, even the same make and model
Modified FDA test, laboratory-developed test, or no manufacturer specifications (establishment)
Follow the manufacturer's instructions, or the laboratory's own criteria using at least 3 levels (minimal or zero, mid-point, near the upper limit of the reportable range); at least every 6 months and after a trigger
Triggers: complete reagent change (unless the lot change is shown not to matter), major maintenance or critical part replaced, QC shift or trend that other fixes do not correct, or the laboratory's own schedule
Reference interval establishment
At least 120 reference subjects per group (CLSI C28-A3)
Central 95% of healthy subjects
Reference interval transference
Test 20 local healthy subjects: accept if no more than 2 fall outside
If 3 or more fall outside, test 20 more; if 3 or more fail again, re-examine the method and population
Proficiency testing and competency
Topic
Rule (42 CFR 493)
Handling PT samples
Test with the routine workload, by the staff who usually do the test, by the routine method, the same number of times as patient specimens
Forbidden
Discussing PT results with another laboratory before the deadline; sending PT samples (or any part) to another laboratory for a test you perform. Intentional referral can cost the certificate for at least a year
Passing score
At least 80% for most analytes and events; 100% for ABO group and D typing, unexpected antibody detection and compatibility testing
Unsatisfactory
Below the passing score in one event; not participating or returning late scores 0
Unsuccessful
Unsatisfactory for the same analyte in 2 consecutive events or 2 of 3 consecutive events: triggers CMS action
No PT available (alternative assessment)
Verify accuracy at least twice a year (split samples, blind specimens, other means)
Competency, nonwaived testing
Six elements for every test: (1) direct observation of routine testing, (2) monitoring recording and reporting of results, (3) review of worksheets, QC, PT and maintenance records, (4) direct observation of instrument maintenance and function checks, (5) testing previously analyzed, blind or PT samples, (6) problem-solving
Competency timing
Twice in the first year of testing, then once a year; retrain and assess before reporting on a new method or instrument
CLIA, accreditation and POCT
Category
How a test gets there
Main requirements
Waived
Cleared for home use, or so simple and accurate that error is negligible, or no reasonable risk of harm if done wrong (examples: dipstick urinalysis, urine hCG by visual color, fecal occult blood, home-use glucose meters, spun microhematocrit)
Certificate of waiver; follow the manufacturer's instructions exactly
Moderate complexity (includes PPM)
FDA scores 7 criteria from 1 to 3; total 12 or less
QC, PT, verification, competency; testing personnel need at least a high school diploma with documented training
High complexity
Total score above 12; any test not yet categorized
All of the above, stricter personnel rules (for example an associate degree in laboratory science or equivalent)
Agencies: FDA categorizes tests; CMS runs CLIA and certifies laboratories; CDC provides technical support, studies and the advisory committee.
Certificates: waiver, PPM, registration, compliance (state inspection) and accreditation (deemed status through a CMS-approved organization: CAP, The Joint Commission, COLA, AABB, A2LA, ACHC, ASHI).
Default QC for quantitative nonwaived tests is two levels each day of testing; qualitative tests need a positive and a negative control. IQCP is the voluntary risk-based alternative.
IQCP = risk assessment (specimen, test system, reagent, environment, testing personnel) + quality control plan + quality assessment. Every specialty except pathology is eligible.
POCT oversight: a POCT coordinator and director, operator records with certification dates, devices that lock out when QC is overdue or fails, connectivity to the LIS, and the same CLIA rules as any test of that complexity.
Errors, reporting and turnaround time
Phase
Examples
Controls
Preanalytical (most errors)
Wrong patient or mislabeled tube, wrong tube or additive, hemolysis, prolonged tourniquet or fist clenching (false high potassium), delayed transport, IV contamination
Failed QC, expired reagent, calibration error, result above the reportable range reported without dilution, interference
QC and Westgard rules, calibration verification, maintenance, dilution protocols
Postanalytical
Transcription error, result sent to the wrong chart, critical value not called, delayed report
Interfaces and autoverification, critical value policy, corrected reports
Critical (panic) value: the laboratory must alert the person who ordered the test, or who uses the result, immediately. Laboratory policy usually adds read-back and a record of who was told and when.
Corrected report: promptly tell the ordering provider of the error, send a corrected report, and keep both the original and the corrected report.
Delta check: compares a result with the same patient's previous result; a change too large for biology suggests misidentification or another error.
Autoverification: rules in the LIS or middleware release results without manual review when they pass QC, flags, limit and delta checks (CLSI AUTO-10A and AUTO-15).
Turnaround time is right-skewed: report the median and the time by which 90% of tests are done, not the mean alone. Most delays are pre- or postanalytical.
Records: QC, PT, patient test records and reports at least 2 years; immunohematology and transfusion records follow the FDA blood rules in 21 CFR 606.160.
Management and education (MLS only)
Bloom's level (revised)
Learner can
Objective verbs
1 Remember
Recall facts
List, name, define
2 Understand
Explain, classify, compare
Explain, describe, compare
3 Apply
Use knowledge in a new situation
Calculate, perform, use
4 Analyze
Break information into parts
Differentiate, troubleshoot, interpret
5 Evaluate
Judge value or quality
Judge, appraise, decide
6 Create
Produce something new
Design, develop, write
Write objectives in behavioral terms: what the learner will be able to do, with a measurable verb ('calculate the CV', not 'understand QC').
Domains: cognitive (knowledge), psychomotor (bench skills, such as making a smear), affective (attitudes, such as professionalism).
Staffing: enough staff to cover the workload, a current competency-based job description for each position, orientation, training, competency assessment, continuing education and an annual performance appraisal.
Budget: part of every plan, with staff, training, equipment, reagents, supplies and QC material; supply quantification each year feeds the budget.
Quality management: the 12 quality system essentials and the plan-do-check-act cycle (see In depth).
The summary above is what the PDF prints. These sections go further.
A worked Levey-Jennings chart
A Levey-Jennings chart plots each control result against the day or run. A center line marks the mean, with lines at ±1, ±2 and ±3 SD. In control, points scatter evenly above and below the mean and about 95% fall within ±2 SD, which is why roughly 1 good run in 20 gives a 1-2s warning on a single level. A shift shows as points jumping to one side and staying there. A trend shows as points climbing or falling steadily. Random error shows as one point far out, or wider scatter. Below is a glucose level 1 control with an established mean of 100 mg/dL (5.6 mmol/L) and an SD of 2.0 mg/dL (0.11 mmol/L), so CV = 2.0%. Each call is the one the Westgard Rules Check calculator makes with this level alone.
z = (result − 100) ÷ 2.0. No single result on days 10 to 13 is even a 1-2s warning: only the multirule catches this small shift.
Day
Result (mg/dL)
z
Call
1
101.0
+0.5
Accept
2
98.0
−1.0
Accept
3
104.6
+2.3
1-2s warning; no rejection rule fires (day 2 was not beyond +2): accept
4
100.8
+0.4
Accept
5
106.4
+3.2
Reject: 1-3s (random error). Fresh vial, then look for a probe clot or bubble
6
99.2
−0.4
Accept after the fix
7
104.4
+2.2
1-2s warning only: accept
8
104.8
+2.4
Reject: 2-2s across runs (days 7 and 8 both above +2 SD; systematic). Recalibrate
9
99.6
−0.2
Accept
10
102.4
+1.2
Accept
11
102.6
+1.3
Accept (only days 10 and 11 above +1 SD)
12
102.2
+1.1
Accept (days 10-12 above +1 SD; day 9 was below the mean)
13
102.8
+1.4
Reject: 4-1s (days 10-13 all above +1 SD; systematic). Check reagent lot and calibration
Run with two levels
Level 1 z
Level 2 z
Call
A
+2.3
−2.2
Reject: R-4s (one above +2, the other below −2 in the same run; random error)
B
+2.1
+2.4
Reject: 2-2s across levels (systematic error)
C
+2.4
+0.6
1-2s warning only; accept if earlier runs break no rule
D
+3.1
−0.5
Reject: 1-3s
Why 1-2s is not a rejection rule: with one control level it rejects about 4.6% of good runs, 8.9% with two levels and 13.0% with three. A laboratory running two levels on 20 analytes will see a false 1-2s somewhere in almost every QC event. Some sources define R-4s as a range of more than 4 SD between the two levels in a run; Westgard and the calculator require one level beyond +2 SD and the other beyond −2 SD, which is the form used here.
Setting up QC and troubleshooting
New control lot: run it alongside the current lot. About 10 results on different days give a first mean; update it as more data come in (a floating mean) and set the SD from longer-term data that include day-to-day sources of variation.
Assayed controls: the manufacturer's value may be used only for your method and instrument and only after you verify it. Unassayed controls: the laboratory sets the mean and SD by running them alongside controls of known value.
Run controls the same way as patient specimens and rotate QC among all operators. If calibrator material is used as a control, it must come from a different lot from the one used to calibrate.
Run controls before resuming patient testing after a complete reagent change, major maintenance or replacement of a critical part.
A new reagent or control lot may shift the mean while the SD stays about the same; CLIA requires a mean and SD for each lot of control material.
Transient problems (inadequate cleaning, electronic noise, a clot or debris in the pipette) affect a few samples and may be gone by the next QC event. Persistent problems (calibration, reagent deterioration, wrong fluid levels, wrong temperature) stay until corrected and change either the bias or the imprecision.
Pattern on the chart
Likely error
First checks
One point beyond 3 SD
Random
Fresh control vial, then clots or debris in the pipette, cleaning, the run's events
Two levels on opposite sides beyond 2 SD
Random (imprecision)
Pipetting and sampling, cleaning, operator technique
Points beyond 2 SD on the same side, or many points on one side of the mean
Systematic (shift)
Calibration, new reagent or control lot
Points creeping steadily up or down
Systematic (trend or drift)
Reagent deterioration, temperature, fluid levels
Total allowable error and sigma
Total allowable error (TEa) is the largest error a result can carry without changing a medical decision. There is no single accepted set of TEa values; the laboratory director chooses one with clinicians, in order of preference from outcome studies, biological variation, then the state of the art (what current methods and peer groups achieve). CLIA's PT acceptance limits are also often used as TEa (a convention, not a CLIA rule; this page uses them for its worked example): glucose ±8% or ±6 mg/dL (whichever is greater), potassium ±0.3 mmol/L, sodium ±4 mmol/L, creatinine ±10% or ±0.2 mg/dL, total calcium ±1.0 mg/dL, total cholesterol ±10%.
Grades from Panda 2023. The 'Meaning for QC' column is a general reading of the published advice that higher sigma allows simpler QC; no source sets fixed QC rules for each band.
Sigma
Grade
Meaning for QC
6 or more
World class
Low error rate; a less stringent QC strategy can be used
5 to below 6
Excellent
Few errors
4 to below 5
Good
Acceptable
3 to below 4
Marginal
More errors; needs stricter rules or more frequent QC
Below 3
Poor or unacceptable
High error rate; the method itself needs improving
Verification in practice
Characteristic
How it is commonly shown
Accuracy
Test materials of known value (calibrators, assayed controls, PT samples, previously tested patient specimens) or compare split patient samples with a method already shown to be accurate; results within the manufacturer's limits
Precision
Repeat the same samples within a run and across days, with several operators (one may be enough on fully automated systems)
Reportable range
Samples of known value at the lowest and highest levels the manufacturer claims (linearity materials or dilutions); report only results inside the verified range; the director decides how results above or below it are handled (for example a validated dilution)
Reference interval
Show the manufacturer's or a published interval fits the local population (the 20-subject transference check)
Analytical sensitivity
Lowest concentration the method can reliably detect (establishment only)
Analytical specificity
Measures only the analyte: interference (hemolysis, icterus, lipemia, drugs) and cross-reactivity (establishment only)
The same samples can serve accuracy, precision and reportable range at once. The technical consultant (moderate) or technical supervisor (high complexity) is responsible; the laboratory director approves the study before patient testing.
Tests in use before April 24, 2003 do not need verification under 493.1253.
A laboratory that runs one test on more than one instrument or site must compare their results twice a year.
Using a waived device off label (for example an arterial sample on a meter cleared only for capillary and venous blood) modifies the test: CMS treats a modified waived or moderate complexity test as uncategorized, so it becomes high complexity, and the laboratory must establish its own performance specifications and meet every high-complexity requirement, personnel included.
Regulation details
PT: enroll in a CMS-approved program for each regulated analyte; stay with a program at least a year; most programs send 5 samples per event, 3 events a year. The target value is usually the peer-group mean after removing results more than 3 SD out; the acceptance limit is a symmetrical ± around it.
After an unsatisfactory PT result the laboratory must take remedial action (training, technical assistance) and document it; a first unsuccessful performance can be met with required training or technical assistance instead of sanctions, unless patients are in immediate jeopardy.
PT only covers the primary method used for patients during the event; keep PT records, including the signed attestation, at least 2 years.
Competency is assessed by the technical consultant (moderate complexity) or technical supervisor (high complexity; may be delegated in writing to a qualified general supervisor). Supervisors and consultants also need competency assessed against their own regulatory duties.
Accreditation organizations must hold the laboratory to standards equal to or stricter than CLIA. Accredited laboratories still need a CLIA certificate (of accreditation), pay CLIA fees and do PT.
Quality assessment: CLIA requires ongoing review of preanalytic, analytic and postanalytic systems, checking whether corrective actions worked, revising procedures and discussing reviews with staff, all documented.
Specimen rules: written procedures for patient preparation, collection, labeling, storage, transport, processing, acceptance and rejection, and referral only to CLIA-certified laboratories; record the date and time each specimen arrives.
Quality management and education
The 12 quality system essentials (CLSI framework, as set out in the WHO Laboratory Quality Management System handbook).
Quality system essential
Covers
Organization
Leadership, structure, quality manual, planning and budget
Quality indicators and the plan-do-check-act cycle
Customer service
Satisfaction surveys, complaints
Facilities and safety
Space, environment, safety programs
Plan-do-check-act (Deming): plan (find the problem and its causes, plan the improvement), do (put the plan into action), check (review whether it worked and revise the plan), act (take the corrective action, recheck, and start the cycle again).
Quality assessment draws on QC reviews, PT performance reviews, chart reviews, specimen rejection logs, turnaround time reports and complaint reports.
Turnaround time can be measured order to report, collection to report or receipt to report (in-lab); say which. In-lab TAT underestimates what the clinician experiences.
Teaching: higher Bloom's levels (apply, analyze, evaluate) need case or problem questions, not recall; an objective's verb tells you how to test it. Bench skills (psychomotor) are assessed by observation, as in competency element 1.
Westgard JO. 'Westgard Rules' and multirules. Westgard QC (westgard.com). Link
Westgard JO. The comparison of methods experiment. Westgard QC (westgard.com). Link
Westgard JO. QC: the multirule interpretation. Westgard QC (westgard.com). Link
Centers for Medicare & Medicaid Services. QSO-25-10-CLIA (revised June 23, 2025): CLIA 2024 final rule guidance and State Operations Manual Appendix C updates. Link
Electronic Code of Federal Regulations. 42 CFR Part 493, Laboratory Requirements (sections 493.2, 493.5, 493.15, 493.17, 493.551, 493.801-493.959, 493.1105, 493.1236-493.1291, 493.1413, 493.1423, 493.1451, 493.1489). Link
Centers for Medicare & Medicaid Services. CLIA brochure: Assessing Personnel Competency. Link
Centers for Medicare & Medicaid Services. CLIA brochure: Verification of Performance Specifications. Link
Centers for Medicare & Medicaid Services. Specialty/subspecialty information for approved accreditation organizations under CLIA. Link
Centers for Disease Control and Prevention, Centers for Medicare & Medicaid Services. Developing an IQCP: A Step-by-Step Guide. Link
Gruber L, Hausch A, Mueller T. Internal quality controls in the medical laboratory: a narrative review of the basic principles of an appropriate quality control plan. Diagnostics (Basel). 2024;14(19):2223. doi:10.3390/diagnostics14192223 Link
Cristelli L, Occhipinti F, Tumiatti D, Antonia L, Jani E, Daves M. Implementation of new Westgard rules suggested by the Westgard Advisor software for five immunological parameters. Biochem Med (Zagreb). 2025;35(1):010701. doi:10.11613/BM.2025.010701 Link
Giannoli JM, Albarede S, Avellan T, et al. Recommendations for the application and follow-up of quality controls in medical laboratories. Biochem Med (Zagreb). 2021;31(2):020501. doi:10.11613/BM.2021.020501 Link
Hedayati M, Tohidi M, Azimzadeh I, et al. A 25-year experience in quality assurance: a methodological review of laboratory quality control and quality assurance in the Tehran Lipid and Glucose Study (TLGS). Int J Endocrinol Metab. 2027;25(1):e168018. doi:10.5812/ijem-168018 Link
Panda CR, Kumari S, Mangaraj M, Nayak S. The evaluation of the quality performance of biochemical analytes in clinical biochemistry laboratory using Six Sigma matrices. Cureus. 2023;15(12):e51386. doi:10.7759/cureus.51386 Link
Ozarda Y. Reference intervals: current status, recent developments and future considerations. Biochem Med (Zagreb). 2016;26(1):5-16. doi:10.11613/BM.2016.001 Link
Hawkins RC. Laboratory turnaround time. Clin Biochem Rev. 2007;28(4):179-194. Link
Adams NE. Bloom's taxonomy of cognitive learning objectives. J Med Libr Assoc. 2015;103(3):152-153. Link
Gül BÜ, Özcan O, Doğan S, Arpaci A. Designing and validating an autoverification system of biochemical test results in Hatay Mustafa Kemal University, clinical laboratory. Biochem Med (Zagreb). 2022;32(3):030704. doi:10.11613/BM.2022.030704 Link
Shin S, Yu S, Kim S, Yoo SJ, Cho EJ, Chung JW. Proposal for delta check limits of frequently requested hormones using real-world data. Biochem Med (Zagreb). 2025;35(1):010704. doi:10.11613/BM.2025.010704 Link
Romero A, Gómez-Salgado J, Romero-Arana A, et al. Preanalytical errors in emergency department samples: investigating error sources. J Med Biochem. 2020;39(4):474-480. doi:10.5937/jomb0-25263 Link
Fung AWS. Utilizing connectivity and data management system for effective quality management and regulatory compliance in point of care testing. Pract Lab Med. 2020;22:e00187. doi:10.1016/j.plabm.2020.e00187 Link
World Health Organization. Laboratory Quality Management System: Handbook. Geneva: WHO; 2011. Link