Beer's law and photometers, optical methods, immunoassay formats and their interferences, electrodes, osmometry, electrophoresis, chromatography and mass spectrometry, automation, point of care, and specimen interferences
Beer's law: A = εbc. Absorbance (A) is proportional to concentration (c) when the path length (b, cm) and molar absorptivity (ε) are fixed.
Absorbance = log (I0 / I), where I0 is light reaching the detector through the blank and I through the sample. From percent transmittance: A = 2 − log %T.
Unknown concentration = (A unknown ÷ A standard) × concentration of the standard, valid only inside the method's linear range. Above it, dilute, re-read and multiply by the dilution factor (see the Beer's Law and Dilution Factor calculators).
Zero the photometer with a reagent blank so that color from the reagent itself is not counted as analyte.
Component
What it does
Exam points
Light source
Deuterium lamp for ultraviolet, tungsten-halogen lamp for visible light
Wrong or aging lamp: low energy, noisy readings in its range
Monochromator
Prism or diffraction grating with slits isolates a narrow band of wavelengths
Narrow slit: better resolution but less light and more noise
Cuvette
Holds the sample in a fixed light path (usually 1 cm)
Quartz for ultraviolet (glass absorbs it); scratches and fingerprints add absorbance
Detector
Photodiode or photomultiplier tube turns light into an electrical signal
Photomultiplier: most sensitive, for low light
Double-beam design
Measures reference and sample paths at the same time
Cancels lamp drift and fluctuations
Check
Material
A failure means
Wavelength accuracy
Holmium oxide glass or solution: sharp bands at certified wavelengths
Every band shifted by the same amount: recalibrate the wavelength scale
Photometric accuracy and linearity
Neutral-density (nearly neutral absorbing) glass filters or certified absorbance solutions
Absorbance values wrong even at the right wavelength
Stray light
Cutoff solutions or filters that transmit essentially no light at the test wavelength (light still detected there is stray light)
Light outside the band reaches the detector; high absorbances read low and the curve bends
Other optical and atomic methods
Method
What is measured
Typical uses
Watch for
Flame emission photometry
Light emitted at the element's own wavelength by atoms excited in a flame
Sodium, potassium, lithium (largely replaced by ISE)
Sample is diluted, so lipemia and high protein lower the result (volume displacement)
Atomic absorption (flame or graphite furnace)
Light from a hollow cathode lamp of the element, absorbed by ground-state atoms
Lithium, trace metals; graphite furnace for very low levels such as lead
One element per lamp; being replaced by ICP-MS
Fluorometry
Light emitted when a molecule excited at one wavelength returns to ground state; detector at 90 degrees
Light scattered by immune complexes at a fixed angle (for example 13 to 24 degrees)
Immunoglobulins, complement, free light chains
Lipemia; antigen excess; kinetic readings help detect excess
ICP-MS
Elements ionized in an argon plasma and counted by mass
Trace and toxic elements, many in one run
Costly; spectral interferences must be controlled
Immunoassay formats and their interferences
Feature
Competitive (reagent-limited)
Sandwich, two-site (reagent excess)
Design
Labeled analyte competes with the patient's analyte for a limited amount of antibody
Capture antibody and labeled detection antibody bind two sites on the analyte
Signal vs concentration
Inverse: more analyte, less signal
Direct: more analyte, more signal
Typical analytes
Small molecules: drugs, steroids, free T4 and free T3
Large molecules: TSH, troponin, hCG, ferritin, prolactin
High-dose biotin (streptavidin-biotin capture)
Falsely HIGH
Falsely LOW
Heterophile antibodies, HAMA, rheumatoid factor
Can bind the conjugate or label: error in either direction
Bridge capture to detection antibody: usually falsely high
High-dose hook effect
Not the classic problem
Antigen excess, especially in one-step assays: falsely low; dilution reveals the true value
Heterogeneous assays separate bound from free label (a wash step) before reading; homogeneous assays read without separation because binding itself changes the signal.
Cross-reactivity: structurally similar compounds or metabolites are read as the analyte (cortisol assays and fludrocortisone, cyclosporine metabolites, digoxin-like factors). Drug screens detect a class by cross-reactivity, so positives are confirmed by GC-MS or LC-MS/MS.
Immunoassays are generally unaffected by hemolysis and icterus; lipemia matters most for nephelometric and turbidimetric methods.
Electrochemistry
Method
Principle
Clinical examples
Potentiometry (ion-selective electrodes)
Voltage between an indicator and a reference electrode at almost zero current; voltage changes with the log of ion activity (Nernst equation)
CO2 diffuses through a membrane into a bicarbonate solution and changes its pH, read by a glass pH electrode
Blood gas pCO2
Amperometry
Current at a fixed applied voltage, proportional to the amount of substance oxidized or reduced
pO2 (Clark electrode), glucose biosensors and meters
Clark pO2 electrode
Oxygen crosses a permeable membrane and is reduced at a platinum cathode held at about −0.8 V; silver/silver chloride anode
Blood gas pO2
Coulometry
A silver anode releases silver ions that precipitate chloride as AgCl; when chloride is used up, free silver ions raise the current and stop the titration; charge (time) is proportional to chloride
Chloridometer: sweat chloride, serum or urine chloride
Conductometry
Electrical conductivity of the whole solution, not one ion
Sweat conductivity: screening only, never diagnostic for cystic fibrosis
Feature
Direct ISE
Indirect ISE
Sample
Undiluted
Diluted (about 1:5 to 1:46) before measurement
Measures
Ion activity in plasma water
Ion in the diluted whole sample, assuming plasma is about 93% water
Where
Blood gas and point-of-care analyzers, some dry-slide systems
High-throughput automated chemistry analyzers
Very high lipids or protein
Unaffected
Falsely LOW Na, K and Cl (electrolyte exclusion): pseudohyponatremia
Osmometry, electrophoresis and separation methods
Method
Principle
Exam points
Freezing point osmometer
Dissolved particles lower the freezing point of water in proportion to osmolality (colligative property)
The usual clinical method; detects ethanol, methanol and ethylene glycol, so it is the one used for the osmolal gap
Vapor pressure osmometer
Dissolved particles lower the vapor pressure (dew point)
Misses volatile alcohols, so an osmolal gap from it is unreliable for toxic alcohols
Electrophoresis (agarose gel)
Proteins move in an electric field by charge and size (usually in barbital buffer near pH 8.6), are stained and scanned
Fractions: albumin, alpha-1, alpha-2, beta (beta-1, beta-2), gamma; abnormal band typed by immunofixation
Capillary zone electrophoresis
Separation in a narrow capillary filled with buffer; proteins are detected, not stained
Automated, fast; abnormal peak typed by immunosubtraction
HPLC
Liquid mobile phase pumped through a packed column; compounds separate by how strongly they partition into the stationary phase (normal or reverse phase)
Reference method for drugs such as cyclosporine; detectors include UV and mass spectrometry
Gas chromatography
Gas mobile phase carries volatile, heat-stable compounds through a column; identified by retention time
Volatile alcohols; GC-MS confirms drugs of abuse with library-matched spectra
LC-MS/MS (tandem mass spectrometry)
LC separation, electrospray ionization, then two mass filters select a precursor ion and its product ion
Most specific routine method: steroids, vitamin D, drugs, newborn screening; stable isotope internal standards correct ion suppression
MALDI-TOF MS
Sample mixed with matrix, hit by a laser; ions fly down a tube and lighter ions arrive first; ribosomal protein spectrum is matched to a database
Identifies bacteria and yeast from colonies in minutes; E. coli and Shigella are not separated
Hemolysis, icterus and lipemia
Interferent
How it interferes
Tests most affected
What to do
Hemolysis
Release of cell contents; hemoglobin color; chemical reaction with reagents
Potassium, LD, AST raised; bilirubin can be lowered
Check the H-index; recollect for affected tests; report with a comment if policy allows
Icterus
Bilirubin color (spectral) and reaction with reagents
Check the I-index; use assay limits; dilution or another method
Lipemia
Light scatter and absorption (worst at short wavelengths); volume displacement; uneven layering
NADH methods read at 340 nm (ALT, AST, glucose); Na, K, Cl by indirect ISE; nephelometry
Check the L-index (visual grading is unreliable); high-speed or ultracentrifugation, then test the infranatant
Tubes, additives and order of draw
Tube (usual stopper)
Additive
Effects to know
Light blue
3.2% (0.109 M) sodium citrate, 1 part to 9 parts blood
Coagulation; chelation can inhibit AST and ALP
Red, gold (SST)
Clot activator, with or without separator gel
Gel can absorb hydrophobic drugs (phenytoin, phenobarbital, carbamazepine, lidocaine): 20 to 50% lower after 24 h at 4 °C; very dense serum may not float above gel
Green
Sodium or lithium heparin
Heparin binds ions, so balanced heparin is used for ionized calcium; no lithium heparin for lithium levels
Lavender
K2 or K3 EDTA
Hematology; chelates Ca, Mg and Zn: carried into other tubes it gives falsely high K and low Ca and Mg
Gray
Sodium fluoride with potassium oxalate (or EDTA)
Fluoride blocks enolase, but glycolysis goes on for 1 to 4 hours; oxalate shrinks cells and inhibits several enzymes
Blood culture bottles (sterile first)
Non-additive tube (or discard tube)
Sodium citrate (coagulation)
Serum tubes: clot activator, then serum separator
Heparin, including plasma separator tubes
EDTA
Fluoride/oxalate
Automation, autoverification and point of care
Area
What it covers
Exam points
Total laboratory automation
Preanalytical: tracks, barcode reception, sorting, centrifugation, aliquoting. Analytical: analyzers linked by middleware. Postanalytical: autoverification, result transfer, storage
Fewer manual steps, shorter and more predictable turnaround
Random access analyzer
Any test on any specimen in any order
Unlike batch analyzers that run one test on a group of specimens
Autoverification
Rules in middleware or the LIS release results without a person
Rules use instrument flags, QC status, serum indices, critical values, delta checks, measuring range and moving averages; each laboratory sets and validates its own limits (CLSI AUTO10 and AUTO15)
Point-of-care testing
Handheld strips and lateral flow devices, cartridges, small bench analyzers near the patient
Waived tests must follow the manufacturer's instructions (42 CFR 493.15); operator training and QC records matter most
Glucose meters
Amperometric strips with glucose oxidase or glucose dehydrogenase on whole blood
The summary above is what the PDF prints. These sections go further.
Beer's law worked examples
Within range. A 200 mg/dL standard reads A = 0.400 and a patient reads A = 0.300 against the same reagent blank. Concentration = 0.300 ÷ 0.400 × 200 = 150 mg/dL.
Above range. The method is linear to A = 0.800 and a patient reads 1.250. The sample is diluted 1 part plus 4 parts saline (factor 5) and reads 0.250. 0.250 ÷ 0.400 × 200 = 125 mg/dL; × 5 = 625 mg/dL. The first reading (1.250) would have given 625 only if the curve stayed straight, which it does not.
Transmittance. A = 2 − log %T, so 100%T gives A = 0, 10%T gives A = 1.0 and 1%T gives A = 2.0. Half the light through (50%T) is A = 0.301. Absorbance, not %T, is proportional to concentration.
Molar absorptivity. If ε is known, c = A ÷ (ε × b). A compound with ε = 6,220 L/mol/cm (the usual NADH value at 340 nm) giving A = 0.311 in a 1 cm cuvette is at 0.311 ÷ 6,220 = 5.0 × 10⁻⁵ mol/L (50 µmol/L).
Why curves bend at the top: at high concentration molecules interact and scatter light, and stray light (light outside the selected band) is a larger share of the little light that still reaches the detector, so absorbance rises less than concentration. Lipemia, turbidity and bubbles add apparent absorbance that is not analyte. Wavelength error moves the reading off the absorbance peak, lowering sensitivity and linearity.
Immunoassay interference: recognizing and resolving it
Clue
Likely cause
Next step
Result far higher than the clinical picture; different on another platform
Heterophile antibody or human anti-animal antibody (HAMA)
Serial dilution (nonlinear recovery), heterophile blocking reagent, PEG precipitation, another platform with different antibodies
Result low or only mildly raised while the patient has a large tumor or very high level expected
High-dose hook effect
Repeat at a dilution (for example 1:10 to 1:100); a rising corrected value confirms it
Sandwich results low and competitive results high in the same patient
Biotin
Ask about supplements; retest after the manufacturer's washout, on a biotin-free platform, or after streptavidin bead depletion
High prolactin without symptoms
Macroprolactin (prolactin bound to IgG)
PEG precipitation or gel filtration to estimate monomeric prolactin
Drug level higher than the parent drug can explain
Cross-reacting metabolites (cyclosporine immunoassays up to 174% above HPLC)
Chromatographic or LC-MS/MS method
Digoxin result after anti-digoxin Fab treatment
Therapeutic antibody fragments bind digoxin in the sample
Report that the result is affected: the fragments compete with the assay antibody
Hook effect examples from the literature: a prolactin of 1,500 mU/L rose to 950,000 mU/L when the sample was diluted, and a urine myoglobin reading 100 nmol/L undiluted read 1,496 nmol/L at 1:50. Thyroglobulin is recommended to be run neat and at 1:10. The hook occurs in two-site assays where capture and detection antibodies are added together: excess free analyte saturates both antibodies separately, so fewer sandwiches form.
Heterophile antibodies are polyreactive natural antibodies or autoantibodies, including rheumatoid factor; HAMA follow exposure to mouse antibodies. In a two-site assay they link the animal capture and detection antibodies with no analyte present. Manufacturers add nonimmune animal immunoglobulin as a blocker, but not every interference is blocked. The laboratory reports suspected interference to the provider, comments the result, and keeps the patient's history on file.
Biotin: most multivitamins contain about 30 µg, which does not interfere. Supplements for hair and nails contain up to 10 mg, and multiple sclerosis studies have used far higher doses. Many laboratories ask patients to stop biotin for at least 48 hours before collection; the assay's package insert gives the interval and the biotin level that interferes, and some troponin assays listed by the FDA have not addressed the risk.
Electrodes in more depth
Nernst equation: E = E0 + (RT ÷ zF) ln a. For a monovalent ion at 25 °C this is about 59 mV per tenfold change in activity; for a divalent ion (Ca²⁺) about half that. ISEs measure activity, which is why ionized calcium is a direct ISE measurement.
Selectivity: an ISE membrane (glass, crystal or polymer with an ionophore) favors one ion but is not perfectly selective; other halides such as bromide are read as chloride by chloride electrodes and colorimetric methods.
pH electrode: glass membrane sensitive to H⁺, read against a reference electrode of constant potential.
pCO2: the Severinghaus electrode is a pH electrode behind a CO2-permeable membrane with a bicarbonate solution; the measured pH change is converted to pCO2. Bicarbonate on a blood gas report is calculated from pH and pCO2.
pO2: the Clark electrode reduces O2 at a platinum cathode (about −0.8 V); the current is proportional to pO2. Very high leukocyte counts keep consuming oxygen in a delayed syringe (see the Electrolytes, Acid-Base and Blood Gases page).
Coulometric chloride: Ag → Ag⁺ + e⁻ at the silver anode; Ag⁺ + Cl⁻ → AgCl. The end point is the sudden rise in free Ag⁺. ICP-MS and ion chromatography are also acceptable for sweat chloride; sweat conductivity and sweat sodium or osmolality are not diagnostic.
Lithium: flame emission and atomic absorption were the original methods; lithium ISEs are precise but other ions can interfere.
Indirect ISE assumes a normal plasma water fraction. When lipids or proteins take up more of the plasma volume, the same aliquot holds less water and fewer electrolytes, and the dilution multiplies the error. Direct ISE senses the activity in plasma water and is unaffected. Very high total protein (myeloma, macroglobulinemia) causes most of the sodium bias, and high triglycerides add to it; potassium errors are much smaller. The laboratory confirms a doubtful sodium on a direct ISE analyzer or by measured osmolality.
Osmolality and the osmolal gap
Calculated osmolality = 2 × Na + glucose ÷ 18 + BUN ÷ 2.8 (glucose and BUN in mg/dL). Osmolal gap = measured − calculated; about 10 mOsm/kg or more suggests unmeasured osmoles (ethanol, methanol, ethylene glycol, mannitol, acetone). See the Osmolality and Osmolar Gap calculators.
Freezing point depression: one osmole of particles per kg of water lowers the freezing point by about 1.86 °C. The instrument supercools the sample, induces freezing and reads the plateau temperature with a thermistor.
Vapor pressure (dew point) osmometers may not detect volatile solutes such as ethanol and methanol: a patient with a high methanol level had no osmolal gap by vapor pressure osmometry, so the gap is sought only with a freezing point osmometer.
Serum and heparin plasma both work; specimens are stable at room temperature for up to 36 hours before separation, with cooling advised for longer delays.
One large study found the gap by freezing point rose with glucose above 110 mg/dL, most strongly above 450 mg/dL, while the vapor pressure gap did not: interpret gaps cautiously in severe hyperglycemia.
Mass spectrometry and trace elements
Part
What it does
Point to know
Electrospray ionization (ESI)
Soft ionization of compounds leaving the LC column
Little fragmentation; multiply charged ions let large proteins be measured
APCI
Chemical ionization at atmospheric pressure
For neutral, low-polarity compounds such as steroids
Electron impact (GC-MS)
Hard ionization
Extensive, reproducible fragmentation: spectra match libraries across instruments
Quadrupole
Four rods pass a narrow band of mass-to-charge (m/z) values
Pairs of precursor and product ions for each analyte
Example: 25-hydroxyvitamin D3 401 > 383 m/z
Ion suppression
Other sample components compete for ionization
Corrected by a stable isotope-labeled internal standard and sample cleanup
Time of flight (MALDI-TOF)
Ions accelerated down a flight tube; flight time gives m/z
337 nm nitrogen laser; matrix absorbs the energy and protects the sample
Direct injection tandem MS without chromatography (as in newborn screening) cannot separate isomers such as leucine and isoleucine, and suppression limits it to higher concentrations. LC-MS/MS product-ion spectra vary between instruments, so library searching is easier with GC-MS.
Trace element method
Strengths
Limits
Flame atomic emission
Cheap, simple
One element, limited range, high detection limit, flammable gases
Flame atomic absorption
Cheap, few interferences
One element per lamp, high detection limit, larger sample
Graphite furnace atomic absorption
Low detection limit, small sample
One element, slow, some need acid digestion
ICP-MS
Many elements at once, wide range, low detection limit, high throughput
Cost, argon and gases, expert staff, interferences to control
Serum indices and lipemia management
Analyzers dilute the sample and read absorbance at several wavelengths to report hemolysis (H), icterus (I) and lipemia (L) indices. Bilirubin and hemoglobin absorb in the lower range; only turbidity absorbs near 700 nm, so lipemia is read there.
Visual grading agrees poorly between staff and with the L-index; indices are better and can drive autoverification rules.
A high L-index with little triglyceride can come from a paraprotein; a very high triglyceride with clear serum and a normal L-index points to free glycerol (glycerol kinase deficiency, glycerol in drugs or nutrition) measured by a non-blanked enzymatic triglyceride method.
Lipid layers: chylomicrons cause most turbidity. Ultracentrifuges (100,000 x g and above) remove lipid best; high-speed centrifugation at about 10,000 x g is almost as good. VLDL-rich lipemia may need repeated spins. Extraction reagents (polyethylene glycol, cyclodextrin, commercial clearing agents) are alternatives with their own limits.
Interference thresholds in package inserts are a starting point; laboratories can set their own after testing. A small group of severely icteric patients accounts for about half of icteric index failures.
Point-of-care quality
CLIA waived tests are simple methods cleared for home use, or so simple and accurate that errors are negligible, or with no reasonable risk of harm if done wrong. The list includes non-automated dipstick urinalysis, fecal occult blood, visual urine pregnancy tests, home-use glucose meters, spun microhematocrit and single-analyte hemoglobin instruments.
Cartridge analyzers: run the electronic QC check each day of use before patients; if there is none, run cartridge controls daily. Check each delivery of cartridges or strips and run a QA sample at least monthly to check storage and operators.
Glucose meters read whole blood, which has less water than plasma: plasma glucose is about 11 to 12% higher at a hematocrit of 45%. Strip enzymes are damaged by heat, cold and humidity; strips and meters must be stored as directed.
Glucose oxidase meters: high pO2 gives low results and hypoxia high results; acetaminophen and ascorbic acid lower results on peroxide-based strips. Some glucose dehydrogenase strips (GDH-PQQ) read maltose, formed from icodextrin dialysate, as glucose.
Unfixed samples keep consuming glucose at about 5 to 7% per hour; fluoride takes 1 to 2 hours or more to stop glycolysis, so plasma is separated promptly for comparisons with meters.
Tate J, Ward G. Interferences in immunoassay. Clin Biochem Rev. 2004;25(2):105-120 Link
Gifford JL, Sadrzadeh SMH, Naugler C. Biotin interference: underrecognized patient safety risk in laboratory testing. Can Fam Physician. 2018;64(5):370 Link
Russwurm M, Wild J, Gobel J. Biotin interference in thyroid function tests. Dtsch Arztebl Int. 2025;122(20):557. doi:10.3238/arztebl.m2025.0114 Link
US Food and Drug Administration. Biotin interference with troponin lab tests: assays subject to biotin interference Link
Hunsaker JJH, La'ulu SL, Zupan E, et al. Detection of a high-dose hook effect and evaluation of dilutions of urine myoglobin specimens using a serum myoglobin assay. Ann Lab Med. 2024;44(4):367-370. doi:10.3343/alm.2023.0427 Link
Nikolac N. Lipemia: causes, interference mechanisms, detection and management. Biochem Med (Zagreb). 2014;24(1):57-67. doi:10.11613/BM.2014.008 Link
Bowen RA, Remaley AT. Interferences from blood collection tube components on clinical chemistry assays. Biochem Med (Zagreb). 2014;24(1):31-44. doi:10.11613/BM.2014.006 Link
Ho CKM, Chen C, Setoh JWS, et al. Optimization of hemolysis, icterus and lipemia interference thresholds for 35 clinical chemistry assays. Pract Lab Med. 2021;25:e00232. doi:10.1016/j.plabm.2021.e00232 Link
Mainali S, Merrill AE, Krasowski MD. Frequency of icteric interference in clinical chemistry laboratory tests and causes of severe icterus. Pract Lab Med. 2021;27:e00259. doi:10.1016/j.plabm.2021.e00259 Link
Bazzano G, Galazzi A, Giusti GD, et al. The order of draw during blood collection: a systematic literature review. Int J Environ Res Public Health. 2021;18(4):1568. doi:10.3390/ijerph18041568 Link
Chopra P, Bharti G, Gupta A, et al. Impact of total protein and triglycerides on the comparability of sodium and potassium measurements by direct and indirect ion-selective electrodes. EJIFCC. 2026;37(4):696-709 Link
Wang Y, Xu H, Zhang J, Li G. Electrochemical sensors for clinic analysis. Sensors (Basel). 2008;8(4):2043-2081. doi:10.3390/s8042043 Link
Bernasconi S, Angelucci A, De Cesari A, et al. Recent technologies for transcutaneous oxygen and carbon dioxide monitoring. Diagnostics (Basel). 2024;14(8):785. doi:10.3390/diagnostics14080785 Link
Taylor PJ, Bouska RA. Automated serum chloride analysis using the Apple computer. J Automat Chem. 1988;10(1):10-14. doi:10.1155/S1463924688000057 Link
Massie J, Greaves R, Metz M, et al. Australasian guideline (2nd edition): an annex to the CLSI and UK guidelines for the performance of the sweat test for the diagnosis of cystic fibrosis. Clin Biochem Rev. 2017;38(3):115-130 Link
Dubot P, Liang J, Dubs J, et al. Sweat chloride quantification using capillary electrophoresis. Pract Lab Med. 2019;13:e00114. doi:10.1016/j.plabm.2018.e00114 Link
Lugassy D, Nelson L. Bromism: forgotten, but not gone. J Med Toxicol. 2009;5(3):151-157 (abstract) Link
Tonyushkina K, Nichols JH. Glucose meters: a review of technical challenges to obtaining accurate results. J Diabetes Sci Technol. 2009;3(4):971-980. doi:10.1177/193229680900300446 Link
St John A, Price CP. Existing and emerging technologies for point-of-care testing. Clin Biochem Rev. 2014;35(3):155-167 Link
Martin CL. Quality control issues in point of care testing. Clin Biochem Rev. 2008;29(Suppl 1):S79-S82 Link
Choy KW, Wijeratne N, Lu ZX, Doery JC. Harmonisation of osmolal gap: can we use a common formula? Clin Biochem Rev. 2016;37(3):113-119 Link
Lin KY, Chen PC, Chen SC, et al. Evaluation of the effects of glucose on osmolal gap using freezing point depression and vapor pressure methods. Kaohsiung J Med Sci. 2018;34(7):409-414. doi:10.1016/j.kjms.2018.03.002 Link
Arif M. UV-visible spectrophotometry in modern research. RSC Adv. 2026;16(43):52112-52147. doi:10.1039/d6ra03975h Link
Pudasaini S, Dangal P, Rathore RS, et al. A portable, low-cost spectrophotometer for absorbance and selected-wavelength fluorescence measurements. J Anal Methods Chem. 2026;2026:9371485. doi:10.1155/jamc/9371485 Link
Allen DW. Holmium oxide glass wavelength standards. J Res Natl Inst Stand Technol. 2007;112(6):303-306. doi:10.6028/jres.112.024 Link
Natali P, Bedin R, Bernardi G, et al. Inter-laboratory concordance of cerebrospinal fluid and serum kappa free light chain measurements. Biomolecules. 2022;12(5):677. doi:10.3390/biom12050677 Link
Mouliou DS. C-reactive protein: pathophysiology, diagnosis, false test results and a novel diagnostic algorithm for clinicians. Diseases. 2023;11(4):132. doi:10.3390/diseases11040132 Link
Cinquanta L, Fontana DE, Bizzaro N. Chemiluminescent immunoassay technology: what does it change in autoantibody detection? Auto Immun Highlights. 2017;8(1):9. doi:10.1007/s13317-017-0097-2 Link
Pitt JJ. Principles and applications of liquid chromatography-mass spectrometry in clinical biochemistry. Clin Biochem Rev. 2009;30(1):19-34 Link
Clark AE, Kaleta EJ, Arora A, Wolk DM. Matrix-assisted laser desorption ionization-time of flight mass spectrometry: a fundamental shift in the routine practice of clinical microbiology. Clin Microbiol Rev. 2013;26(3):547-603. doi:10.1128/CMR.00072-12 Link
Wilschefski SC, Baxter MR. Inductively coupled plasma mass spectrometry: introduction to analytical aspects. Clin Biochem Rev. 2019;40(3):115-133. doi:10.33176/AACB-19-00024 Link
Aliasgharpour M, Hagani H. Evaluation of lithium determination in three analyzers: flame emission, flame atomic absorption spectroscopy and ion selective electrode. N Am J Med Sci. 2009;1(5):244-246. doi:10.4297/najms.2009.5244 Link
Nam Y, Park HD. Revolutionizing laboratory practices: pioneering trends in total laboratory automation. Ann Lab Med. 2025;45(5):472-483. doi:10.3343/alm.2024.0581 Link
Gul BU, Ozcan O, Dogan S, Arpaci A. Designing and validating an autoverification system of biochemical test results in Hatay Mustafa Kemal University, clinical laboratory. Biochem Med (Zagreb). 2022;32(3):030704. doi:10.11613/BM.2022.030704 Link
Jamal I, Shuchismita, Raman RB, Choudhary V. Peaks and distortions: evaluation of irregularities in capillary serum protein electrophoresis. Indian J Hematol Blood Transfus. 2025;41(2):349-356. doi:10.1007/s12288-024-01853-4 Link
Ren Q, Liu T, Li J, et al. Bisalbuminemia: a rare finding identified via serum protein agarose gel electrophoresis and capillary electrophoresis techniques. EJIFCC. 2026;37(3):425-431 Link
Testing.com (ADLM). Protein electrophoresis and immunofixation electrophoresis Link
Reule AG. Errors in spectrophotometry and calibration procedures to avoid them. J Res Natl Bur Stand A Phys Chem. 1976;80A(4):609-624. doi:10.6028/jres.080a.060 Link
Balzer AHA, Whitehurst CB. An analysis of the biotin-(strept)avidin system in immunoassays: interference and mitigation strategies. Curr Issues Mol Biol. 2023;45(11):8733-8754. doi:10.3390/cimb45110549 Link
Walker JA, Schwartzbard A, Krauss EA, et al. The missing gap: a pitfall in the diagnosis of alcohol intoxication by osmometry. Arch Intern Med. 1986;146(9):1843-1844 (abstract). doi:10.1001/archinte.146.9.1843 Link
Brizzee L, Stone A, Palmer MC. False lithium toxicity secondary to lithium heparin test tube: a case report and review. Ment Health Clin. 2020;10(3):90-94. doi:10.9740/mhc.2020.05.090 Link
Upadhyay S. An optimized enzyme-coupled spectrophotometric method for measuring pyruvate kinase kinetics. Bio Protoc. 2025;15(16):e5415. doi:10.21769/bioprotoc.5415 Link
Adedeji AL, Adenikinju RO, Ajele JO, Olawoye TL. Serum protein electrophoresis under effective control of HIV-1 disease progression. EXCLI J. 2014;13:761-771 Link
Sulaiman A, Caturegli P. Falsely abnormal serum protein electrophoresis after administration of intravenous immunoglobulins (IVIG): a retrospective cohort study. Pract Lab Med. 2024;42:e00434. doi:10.1016/j.plabm.2024.e00434 Link