Study sheet

Chemistry Methods and Instrumentation

Beer's law and photometers, optical methods, immunoassay formats and their interferences, electrodes, osmometry, electrophoresis, chromatography and mass spectrometry, automation, point of care, and specimen interferences

Checked against current sources November 2026

Page 1 of the Chemistry Methods and Instrumentation PDF, a printed page in columns with sections on Spectrophotometry and Beer's law, Other optical and atomic methods, Immunoassay formats and their interferences, Electrochemistry, Osmometry, electrophoresis and separation methods, Hemolysis, icterus and lipemia, Tubes, additives and order of draw and Automation, autoverification and point of careEnlarge

Printable PDFLetter landscape, ready to print or keep with your notes

View PDFDownload

Chemistry Methods and Instrumentation: PDF preview

Page 1 of the Chemistry Methods and Instrumentation PDF, a printed page in columns with sections on Spectrophotometry and Beer's law, Other optical and atomic methods, Immunoassay formats and their interferences, Electrochemistry, Osmometry, electrophoresis and separation methods, Hemolysis, icterus and lipemia, Tubes, additives and order of draw and Automation, autoverification and point of care

View PDFDownload

Spectrophotometry and Beer's law

ComponentWhat it doesExam points
Light sourceDeuterium lamp for ultraviolet, tungsten-halogen lamp for visible lightWrong or aging lamp: low energy, noisy readings in its range
MonochromatorPrism or diffraction grating with slits isolates a narrow band of wavelengthsNarrow slit: better resolution but less light and more noise
CuvetteHolds the sample in a fixed light path (usually 1 cm)Quartz for ultraviolet (glass absorbs it); scratches and fingerprints add absorbance
DetectorPhotodiode or photomultiplier tube turns light into an electrical signalPhotomultiplier: most sensitive, for low light
Double-beam designMeasures reference and sample paths at the same timeCancels lamp drift and fluctuations
CheckMaterialA failure means
Wavelength accuracyHolmium oxide glass or solution: sharp bands at certified wavelengthsEvery band shifted by the same amount: recalibrate the wavelength scale
Photometric accuracy and linearityNeutral-density (nearly neutral absorbing) glass filters or certified absorbance solutionsAbsorbance values wrong even at the right wavelength
Stray lightCutoff solutions or filters that transmit essentially no light at the test wavelength (light still detected there is stray light)Light outside the band reaches the detector; high absorbances read low and the curve bends

Other optical and atomic methods

MethodWhat is measuredTypical usesWatch for
Flame emission photometryLight emitted at the element's own wavelength by atoms excited in a flameSodium, potassium, lithium (largely replaced by ISE)Sample is diluted, so lipemia and high protein lower the result (volume displacement)
Atomic absorption (flame or graphite furnace)Light from a hollow cathode lamp of the element, absorbed by ground-state atomsLithium, trace metals; graphite furnace for very low levels such as leadOne element per lamp; being replaced by ICP-MS
FluorometryLight emitted when a molecule excited at one wavelength returns to ground state; detector at 90 degreesFluorescent immunoassays, time-resolved fluorescenceSelf-quenching and inner-filter effects at high concentration; background fluorescence (read a blank)
ChemiluminescenceLight from a chemical reaction; no excitation lamp neededMost automated immunoassays (acridinium or ruthenium labels, or enzyme labels with luminol or dioxetane substrates)Very wide dynamic range and high sensitivity
TurbidimetryLoss of transmitted light (absorbance) through a suspension of immune complexesCRP, immunoglobulins, latex-enhanced assays on general chemistry analyzersLipemia; antigen excess (prozone) reads falsely low
NephelometryLight scattered by immune complexes at a fixed angle (for example 13 to 24 degrees)Immunoglobulins, complement, free light chainsLipemia; antigen excess; kinetic readings help detect excess
ICP-MSElements ionized in an argon plasma and counted by massTrace and toxic elements, many in one runCostly; spectral interferences must be controlled

Immunoassay formats and their interferences

FeatureCompetitive (reagent-limited)Sandwich, two-site (reagent excess)
DesignLabeled analyte competes with the patient's analyte for a limited amount of antibodyCapture antibody and labeled detection antibody bind two sites on the analyte
Signal vs concentrationInverse: more analyte, less signalDirect: more analyte, more signal
Typical analytesSmall molecules: drugs, steroids, free T4 and free T3Large molecules: TSH, troponin, hCG, ferritin, prolactin
High-dose biotin (streptavidin-biotin capture)Falsely HIGHFalsely LOW
Heterophile antibodies, HAMA, rheumatoid factorCan bind the conjugate or label: error in either directionBridge capture to detection antibody: usually falsely high
High-dose hook effectNot the classic problemAntigen excess, especially in one-step assays: falsely low; dilution reveals the true value

Electrochemistry

MethodPrincipleClinical examples
Potentiometry (ion-selective electrodes)Voltage between an indicator and a reference electrode at almost zero current; voltage changes with the log of ion activity (Nernst equation)Na, K, Cl, Li, ionized Ca; pH (glass electrode); pCO2 (Severinghaus electrode)
Severinghaus pCO2 electrodeCO2 diffuses through a membrane into a bicarbonate solution and changes its pH, read by a glass pH electrodeBlood gas pCO2
AmperometryCurrent at a fixed applied voltage, proportional to the amount of substance oxidized or reducedpO2 (Clark electrode), glucose biosensors and meters
Clark pO2 electrodeOxygen crosses a permeable membrane and is reduced at a platinum cathode held at about −0.8 V; silver/silver chloride anodeBlood gas pO2
CoulometryA silver anode releases silver ions that precipitate chloride as AgCl; when chloride is used up, free silver ions raise the current and stop the titration; charge (time) is proportional to chlorideChloridometer: sweat chloride, serum or urine chloride
ConductometryElectrical conductivity of the whole solution, not one ionSweat conductivity: screening only, never diagnostic for cystic fibrosis
FeatureDirect ISEIndirect ISE
SampleUndilutedDiluted (about 1:5 to 1:46) before measurement
MeasuresIon activity in plasma waterIon in the diluted whole sample, assuming plasma is about 93% water
WhereBlood gas and point-of-care analyzers, some dry-slide systemsHigh-throughput automated chemistry analyzers
Very high lipids or proteinUnaffectedFalsely LOW Na, K and Cl (electrolyte exclusion): pseudohyponatremia

Osmometry, electrophoresis and separation methods

MethodPrincipleExam points
Freezing point osmometerDissolved particles lower the freezing point of water in proportion to osmolality (colligative property)The usual clinical method; detects ethanol, methanol and ethylene glycol, so it is the one used for the osmolal gap
Vapor pressure osmometerDissolved particles lower the vapor pressure (dew point)Misses volatile alcohols, so an osmolal gap from it is unreliable for toxic alcohols
Electrophoresis (agarose gel)Proteins move in an electric field by charge and size (usually in barbital buffer near pH 8.6), are stained and scannedFractions: albumin, alpha-1, alpha-2, beta (beta-1, beta-2), gamma; abnormal band typed by immunofixation
Capillary zone electrophoresisSeparation in a narrow capillary filled with buffer; proteins are detected, not stainedAutomated, fast; abnormal peak typed by immunosubtraction
HPLCLiquid mobile phase pumped through a packed column; compounds separate by how strongly they partition into the stationary phase (normal or reverse phase)Reference method for drugs such as cyclosporine; detectors include UV and mass spectrometry
Gas chromatographyGas mobile phase carries volatile, heat-stable compounds through a column; identified by retention timeVolatile alcohols; GC-MS confirms drugs of abuse with library-matched spectra
LC-MS/MS (tandem mass spectrometry)LC separation, electrospray ionization, then two mass filters select a precursor ion and its product ionMost specific routine method: steroids, vitamin D, drugs, newborn screening; stable isotope internal standards correct ion suppression
MALDI-TOF MSSample mixed with matrix, hit by a laser; ions fly down a tube and lighter ions arrive first; ribosomal protein spectrum is matched to a databaseIdentifies bacteria and yeast from colonies in minutes; E. coli and Shigella are not separated

Hemolysis, icterus and lipemia

InterferentHow it interferesTests most affectedWhat to do
HemolysisRelease of cell contents; hemoglobin color; chemical reaction with reagentsPotassium, LD, AST raised; bilirubin can be loweredCheck the H-index; recollect for affected tests; report with a comment if policy allows
IcterusBilirubin color (spectral) and reaction with reagentsPeroxidase-coupled methods (cholesterol, triglyceride, uric acid) and creatinine read low; co-oximetryCheck the I-index; use assay limits; dilution or another method
LipemiaLight scatter and absorption (worst at short wavelengths); volume displacement; uneven layeringNADH methods read at 340 nm (ALT, AST, glucose); Na, K, Cl by indirect ISE; nephelometryCheck the L-index (visual grading is unreliable); high-speed or ultracentrifugation, then test the infranatant

Tubes, additives and order of draw

Tube (usual stopper)AdditiveEffects to know
Light blue3.2% (0.109 M) sodium citrate, 1 part to 9 parts bloodCoagulation; chelation can inhibit AST and ALP
Red, gold (SST)Clot activator, with or without separator gelGel can absorb hydrophobic drugs (phenytoin, phenobarbital, carbamazepine, lidocaine): 20 to 50% lower after 24 h at 4 °C; very dense serum may not float above gel
GreenSodium or lithium heparinHeparin binds ions, so balanced heparin is used for ionized calcium; no lithium heparin for lithium levels
LavenderK2 or K3 EDTAHematology; chelates Ca, Mg and Zn: carried into other tubes it gives falsely high K and low Ca and Mg
GraySodium fluoride with potassium oxalate (or EDTA)Fluoride blocks enolase, but glycolysis goes on for 1 to 4 hours; oxalate shrinks cells and inhibits several enzymes
  1. Blood culture bottles (sterile first)
  2. Non-additive tube (or discard tube)
  3. Sodium citrate (coagulation)
  4. Serum tubes: clot activator, then serum separator
  5. Heparin, including plasma separator tubes
  6. EDTA
  7. Fluoride/oxalate

Automation, autoverification and point of care

AreaWhat it coversExam points
Total laboratory automationPreanalytical: tracks, barcode reception, sorting, centrifugation, aliquoting. Analytical: analyzers linked by middleware. Postanalytical: autoverification, result transfer, storageFewer manual steps, shorter and more predictable turnaround
Random access analyzerAny test on any specimen in any orderUnlike batch analyzers that run one test on a group of specimens
AutoverificationRules in middleware or the LIS release results without a personRules use instrument flags, QC status, serum indices, critical values, delta checks, measuring range and moving averages; each laboratory sets and validates its own limits (CLSI AUTO10 and AUTO15)
Point-of-care testingHandheld strips and lateral flow devices, cartridges, small bench analyzers near the patientWaived tests must follow the manufacturer's instructions (42 CFR 493.15); operator training and QC records matter most
Glucose metersAmperometric strips with glucose oxidase or glucose dehydrogenase on whole bloodHigh hematocrit reads low, anemia reads high; oxygen therapy lowers glucose oxidase results; maltose (icodextrin dialysate) falsely raises some GDH strips

In depth

The summary above is what the PDF prints. These sections go further.

Beer's law worked examples

  1. Within range. A 200 mg/dL standard reads A = 0.400 and a patient reads A = 0.300 against the same reagent blank. Concentration = 0.300 ÷ 0.400 × 200 = 150 mg/dL.
  2. Above range. The method is linear to A = 0.800 and a patient reads 1.250. The sample is diluted 1 part plus 4 parts saline (factor 5) and reads 0.250. 0.250 ÷ 0.400 × 200 = 125 mg/dL; × 5 = 625 mg/dL. The first reading (1.250) would have given 625 only if the curve stayed straight, which it does not.
  3. Transmittance. A = 2 − log %T, so 100%T gives A = 0, 10%T gives A = 1.0 and 1%T gives A = 2.0. Half the light through (50%T) is A = 0.301. Absorbance, not %T, is proportional to concentration.
  4. Molar absorptivity. If ε is known, c = A ÷ (ε × b). A compound with ε = 6,220 L/mol/cm (the usual NADH value at 340 nm) giving A = 0.311 in a 1 cm cuvette is at 0.311 ÷ 6,220 = 5.0 × 10⁻⁵ mol/L (50 µmol/L).

Why curves bend at the top: at high concentration molecules interact and scatter light, and stray light (light outside the selected band) is a larger share of the little light that still reaches the detector, so absorbance rises less than concentration. Lipemia, turbidity and bubbles add apparent absorbance that is not analyte. Wavelength error moves the reading off the absorbance peak, lowering sensitivity and linearity.

Immunoassay interference: recognizing and resolving it

ClueLikely causeNext step
Result far higher than the clinical picture; different on another platformHeterophile antibody or human anti-animal antibody (HAMA)Serial dilution (nonlinear recovery), heterophile blocking reagent, PEG precipitation, another platform with different antibodies
Result low or only mildly raised while the patient has a large tumor or very high level expectedHigh-dose hook effectRepeat at a dilution (for example 1:10 to 1:100); a rising corrected value confirms it
Sandwich results low and competitive results high in the same patientBiotinAsk about supplements; retest after the manufacturer's washout, on a biotin-free platform, or after streptavidin bead depletion
High prolactin without symptomsMacroprolactin (prolactin bound to IgG)PEG precipitation or gel filtration to estimate monomeric prolactin
Drug level higher than the parent drug can explainCross-reacting metabolites (cyclosporine immunoassays up to 174% above HPLC)Chromatographic or LC-MS/MS method
Digoxin result after anti-digoxin Fab treatmentTherapeutic antibody fragments bind digoxin in the sampleReport that the result is affected: the fragments compete with the assay antibody

Hook effect examples from the literature: a prolactin of 1,500 mU/L rose to 950,000 mU/L when the sample was diluted, and a urine myoglobin reading 100 nmol/L undiluted read 1,496 nmol/L at 1:50. Thyroglobulin is recommended to be run neat and at 1:10. The hook occurs in two-site assays where capture and detection antibodies are added together: excess free analyte saturates both antibodies separately, so fewer sandwiches form.

Heterophile antibodies are polyreactive natural antibodies or autoantibodies, including rheumatoid factor; HAMA follow exposure to mouse antibodies. In a two-site assay they link the animal capture and detection antibodies with no analyte present. Manufacturers add nonimmune animal immunoglobulin as a blocker, but not every interference is blocked. The laboratory reports suspected interference to the provider, comments the result, and keeps the patient's history on file.

Biotin: most multivitamins contain about 30 µg, which does not interfere. Supplements for hair and nails contain up to 10 mg, and multiple sclerosis studies have used far higher doses. Many laboratories ask patients to stop biotin for at least 48 hours before collection; the assay's package insert gives the interval and the biotin level that interferes, and some troponin assays listed by the FDA have not addressed the risk.

Electrodes in more depth

Indirect ISE assumes a normal plasma water fraction. When lipids or proteins take up more of the plasma volume, the same aliquot holds less water and fewer electrolytes, and the dilution multiplies the error. Direct ISE senses the activity in plasma water and is unaffected. Very high total protein (myeloma, macroglobulinemia) causes most of the sodium bias, and high triglycerides add to it; potassium errors are much smaller. The laboratory confirms a doubtful sodium on a direct ISE analyzer or by measured osmolality.

Osmolality and the osmolal gap

Mass spectrometry and trace elements

PartWhat it doesPoint to know
Electrospray ionization (ESI)Soft ionization of compounds leaving the LC columnLittle fragmentation; multiply charged ions let large proteins be measured
APCIChemical ionization at atmospheric pressureFor neutral, low-polarity compounds such as steroids
Electron impact (GC-MS)Hard ionizationExtensive, reproducible fragmentation: spectra match libraries across instruments
QuadrupoleFour rods pass a narrow band of mass-to-charge (m/z) valuesTriple quadrupole: Q1 selects precursor, Q2 fragments it (collision cell), Q3 selects product
Multiple reaction monitoringPairs of precursor and product ions for each analyteExample: 25-hydroxyvitamin D3 401 > 383 m/z
Ion suppressionOther sample components compete for ionizationCorrected by a stable isotope-labeled internal standard and sample cleanup
Time of flight (MALDI-TOF)Ions accelerated down a flight tube; flight time gives m/z337 nm nitrogen laser; matrix absorbs the energy and protects the sample

Direct injection tandem MS without chromatography (as in newborn screening) cannot separate isomers such as leucine and isoleucine, and suppression limits it to higher concentrations. LC-MS/MS product-ion spectra vary between instruments, so library searching is easier with GC-MS.

Trace element methodStrengthsLimits
Flame atomic emissionCheap, simpleOne element, limited range, high detection limit, flammable gases
Flame atomic absorptionCheap, few interferencesOne element per lamp, high detection limit, larger sample
Graphite furnace atomic absorptionLow detection limit, small sampleOne element, slow, some need acid digestion
ICP-MSMany elements at once, wide range, low detection limit, high throughputCost, argon and gases, expert staff, interferences to control

Serum indices and lipemia management

Point-of-care quality

Practice free in your browser600+ exam-style questions with rationales. Free today, no account, no ads. Pro is coming in January 2027. Also free on Google Play.

Related study sheets

Calculators for this topic

Sources

  1. Tate J, Ward G. Interferences in immunoassay. Clin Biochem Rev. 2004;25(2):105-120 Link
  2. Gifford JL, Sadrzadeh SMH, Naugler C. Biotin interference: underrecognized patient safety risk in laboratory testing. Can Fam Physician. 2018;64(5):370 Link
  3. Russwurm M, Wild J, Gobel J. Biotin interference in thyroid function tests. Dtsch Arztebl Int. 2025;122(20):557. doi:10.3238/arztebl.m2025.0114 Link
  4. US Food and Drug Administration. Biotin interference with troponin lab tests: assays subject to biotin interference Link
  5. Hunsaker JJH, La'ulu SL, Zupan E, et al. Detection of a high-dose hook effect and evaluation of dilutions of urine myoglobin specimens using a serum myoglobin assay. Ann Lab Med. 2024;44(4):367-370. doi:10.3343/alm.2023.0427 Link
  6. Nikolac N. Lipemia: causes, interference mechanisms, detection and management. Biochem Med (Zagreb). 2014;24(1):57-67. doi:10.11613/BM.2014.008 Link
  7. Bowen RA, Remaley AT. Interferences from blood collection tube components on clinical chemistry assays. Biochem Med (Zagreb). 2014;24(1):31-44. doi:10.11613/BM.2014.006 Link
  8. Ho CKM, Chen C, Setoh JWS, et al. Optimization of hemolysis, icterus and lipemia interference thresholds for 35 clinical chemistry assays. Pract Lab Med. 2021;25:e00232. doi:10.1016/j.plabm.2021.e00232 Link
  9. Mainali S, Merrill AE, Krasowski MD. Frequency of icteric interference in clinical chemistry laboratory tests and causes of severe icterus. Pract Lab Med. 2021;27:e00259. doi:10.1016/j.plabm.2021.e00259 Link
  10. Bazzano G, Galazzi A, Giusti GD, et al. The order of draw during blood collection: a systematic literature review. Int J Environ Res Public Health. 2021;18(4):1568. doi:10.3390/ijerph18041568 Link
  11. Chopra P, Bharti G, Gupta A, et al. Impact of total protein and triglycerides on the comparability of sodium and potassium measurements by direct and indirect ion-selective electrodes. EJIFCC. 2026;37(4):696-709 Link
  12. Wang Y, Xu H, Zhang J, Li G. Electrochemical sensors for clinic analysis. Sensors (Basel). 2008;8(4):2043-2081. doi:10.3390/s8042043 Link
  13. Bernasconi S, Angelucci A, De Cesari A, et al. Recent technologies for transcutaneous oxygen and carbon dioxide monitoring. Diagnostics (Basel). 2024;14(8):785. doi:10.3390/diagnostics14080785 Link
  14. Taylor PJ, Bouska RA. Automated serum chloride analysis using the Apple computer. J Automat Chem. 1988;10(1):10-14. doi:10.1155/S1463924688000057 Link
  15. Massie J, Greaves R, Metz M, et al. Australasian guideline (2nd edition): an annex to the CLSI and UK guidelines for the performance of the sweat test for the diagnosis of cystic fibrosis. Clin Biochem Rev. 2017;38(3):115-130 Link
  16. Dubot P, Liang J, Dubs J, et al. Sweat chloride quantification using capillary electrophoresis. Pract Lab Med. 2019;13:e00114. doi:10.1016/j.plabm.2018.e00114 Link
  17. Lugassy D, Nelson L. Bromism: forgotten, but not gone. J Med Toxicol. 2009;5(3):151-157 (abstract) Link
  18. Tonyushkina K, Nichols JH. Glucose meters: a review of technical challenges to obtaining accurate results. J Diabetes Sci Technol. 2009;3(4):971-980. doi:10.1177/193229680900300446 Link
  19. St John A, Price CP. Existing and emerging technologies for point-of-care testing. Clin Biochem Rev. 2014;35(3):155-167 Link
  20. Martin CL. Quality control issues in point of care testing. Clin Biochem Rev. 2008;29(Suppl 1):S79-S82 Link
  21. 42 CFR 493.15, Laboratories performing waived tests (eCFR, read 2026-11-15) Link
  22. Choy KW, Wijeratne N, Lu ZX, Doery JC. Harmonisation of osmolal gap: can we use a common formula? Clin Biochem Rev. 2016;37(3):113-119 Link
  23. Lin KY, Chen PC, Chen SC, et al. Evaluation of the effects of glucose on osmolal gap using freezing point depression and vapor pressure methods. Kaohsiung J Med Sci. 2018;34(7):409-414. doi:10.1016/j.kjms.2018.03.002 Link
  24. Arif M. UV-visible spectrophotometry in modern research. RSC Adv. 2026;16(43):52112-52147. doi:10.1039/d6ra03975h Link
  25. Pudasaini S, Dangal P, Rathore RS, et al. A portable, low-cost spectrophotometer for absorbance and selected-wavelength fluorescence measurements. J Anal Methods Chem. 2026;2026:9371485. doi:10.1155/jamc/9371485 Link
  26. Allen DW. Holmium oxide glass wavelength standards. J Res Natl Inst Stand Technol. 2007;112(6):303-306. doi:10.6028/jres.112.024 Link
  27. Natali P, Bedin R, Bernardi G, et al. Inter-laboratory concordance of cerebrospinal fluid and serum kappa free light chain measurements. Biomolecules. 2022;12(5):677. doi:10.3390/biom12050677 Link
  28. Mouliou DS. C-reactive protein: pathophysiology, diagnosis, false test results and a novel diagnostic algorithm for clinicians. Diseases. 2023;11(4):132. doi:10.3390/diseases11040132 Link
  29. Cinquanta L, Fontana DE, Bizzaro N. Chemiluminescent immunoassay technology: what does it change in autoantibody detection? Auto Immun Highlights. 2017;8(1):9. doi:10.1007/s13317-017-0097-2 Link
  30. Pitt JJ. Principles and applications of liquid chromatography-mass spectrometry in clinical biochemistry. Clin Biochem Rev. 2009;30(1):19-34 Link
  31. Clark AE, Kaleta EJ, Arora A, Wolk DM. Matrix-assisted laser desorption ionization-time of flight mass spectrometry: a fundamental shift in the routine practice of clinical microbiology. Clin Microbiol Rev. 2013;26(3):547-603. doi:10.1128/CMR.00072-12 Link
  32. Wilschefski SC, Baxter MR. Inductively coupled plasma mass spectrometry: introduction to analytical aspects. Clin Biochem Rev. 2019;40(3):115-133. doi:10.33176/AACB-19-00024 Link
  33. Aliasgharpour M, Hagani H. Evaluation of lithium determination in three analyzers: flame emission, flame atomic absorption spectroscopy and ion selective electrode. N Am J Med Sci. 2009;1(5):244-246. doi:10.4297/najms.2009.5244 Link
  34. Nam Y, Park HD. Revolutionizing laboratory practices: pioneering trends in total laboratory automation. Ann Lab Med. 2025;45(5):472-483. doi:10.3343/alm.2024.0581 Link
  35. Gul BU, Ozcan O, Dogan S, Arpaci A. Designing and validating an autoverification system of biochemical test results in Hatay Mustafa Kemal University, clinical laboratory. Biochem Med (Zagreb). 2022;32(3):030704. doi:10.11613/BM.2022.030704 Link
  36. Jamal I, Shuchismita, Raman RB, Choudhary V. Peaks and distortions: evaluation of irregularities in capillary serum protein electrophoresis. Indian J Hematol Blood Transfus. 2025;41(2):349-356. doi:10.1007/s12288-024-01853-4 Link
  37. Ren Q, Liu T, Li J, et al. Bisalbuminemia: a rare finding identified via serum protein agarose gel electrophoresis and capillary electrophoresis techniques. EJIFCC. 2026;37(3):425-431 Link
  38. Testing.com (ADLM). Protein electrophoresis and immunofixation electrophoresis Link
  39. Reule AG. Errors in spectrophotometry and calibration procedures to avoid them. J Res Natl Bur Stand A Phys Chem. 1976;80A(4):609-624. doi:10.6028/jres.080a.060 Link
  40. Balzer AHA, Whitehurst CB. An analysis of the biotin-(strept)avidin system in immunoassays: interference and mitigation strategies. Curr Issues Mol Biol. 2023;45(11):8733-8754. doi:10.3390/cimb45110549 Link
  41. Walker JA, Schwartzbard A, Krauss EA, et al. The missing gap: a pitfall in the diagnosis of alcohol intoxication by osmometry. Arch Intern Med. 1986;146(9):1843-1844 (abstract). doi:10.1001/archinte.146.9.1843 Link
  42. Brizzee L, Stone A, Palmer MC. False lithium toxicity secondary to lithium heparin test tube: a case report and review. Ment Health Clin. 2020;10(3):90-94. doi:10.9740/mhc.2020.05.090 Link
  43. Upadhyay S. An optimized enzyme-coupled spectrophotometric method for measuring pyruvate kinase kinetics. Bio Protoc. 2025;15(16):e5415. doi:10.21769/bioprotoc.5415 Link
  44. Adedeji AL, Adenikinju RO, Ajele JO, Olawoye TL. Serum protein electrophoresis under effective control of HIV-1 disease progression. EXCLI J. 2014;13:761-771 Link
  45. Sulaiman A, Caturegli P. Falsely abnormal serum protein electrophoresis after administration of intravenous immunoglobulins (IVIG): a retrospective cohort study. Pract Lab Med. 2024;42:e00434. doi:10.1016/j.plabm.2024.e00434 Link